TRIGEN 71634000Recalls & reported problems

Smith & Nephew, Inc. · NAIL CAP 0MM · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03596010495709), received between May 29, 2019 and Aug 12, 2025 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Connection Problem” (2 reports).

6 other recalls cover devices cleared under the same FDA submission (K210980, K981529); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Connection Problem2
  • Fracture2
  • Insufficient Device Problem Information1

Patient problems

  • No Known Impact Or Consequence To Patient2
  • Device Embedded In Tissue or Plaque1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 12, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN ORIF SURGERY, ONE (1) NAIL CAP {} 0MM WAS INSERTED AT THE CORRECT ANGLE, BUT IT DID NOT FULLY ENGAGE AND BACKED OUT. AFTER SEVERAL ATTEMPTS AT DIFFERENT ANGLES AND WITH INCREASED FORCE, THE EDGE OF THE NAIL CAP {} 0MM, WHERE IT ATTACHES TO THE KEY, BROKE MAKING ITS REMOVAL IMPOSSIBLE. THE PROCEDURE WAS RESUMED WITHOUT DELAY USING THE SAME DEVICE, AND NO INJURIES WERE REPORTED AS A RESULT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 22787684 ↗
  • MalfunctionJun 2, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN OSTEOSYNTHESIS PROCEDURE, AT THE END OF FIXING THE NAIL CAP {} 0MM, IT WAS DECIDED TO PLACE A CAP PROXIMAL TO THE NAIL. HOWEVER, WHEN ADJUSTING THE THREAD TO THE NAIL, IT BROKE AND DID NOT ALLOW FOR EASY REMOVAL. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22123444 ↗
  • MalfunctionNov 11, 2020Connection Problem; Connection Problem

    IT WAS REPORTED THAT DURING PROCEDURE WHEN THE TIBIAL INTRAMEDULLARY NAIL WAS FINALLY PUT ON THE NAIL CAP, THERE WAS A PROBLEM WITH THE THREAD OF THE NAIL CAP, WHICH COULD NOT BE SCREWED IN. SURGERY WAS COMPLETED WITH AN S+N BACKUP DEVICE. NO INFORMATION ABOUT INJURIES OR SURGICAL DELAYS WERE REPORTED.

    Patient problems: Insufficient Information; No Known Impact Or Consequence To Patient

    MAUDE report 10826068 ↗
  • InjuryNov 11, 2020Connection Problem; Connection Problem

    IT WAS REPORTED THAT DURING PROCEDURE WHEN THE TIBIAL INTRAMEDULLARY NAIL WAS FINALLY PUT ON THE NAIL CAP, THERE WAS A PROBLEM WITH THE THREAD OF THE NAIL CAP, WHICH COULD NOT BE SCREWED IN. SURGERY WAS COMPLETED WITH AN S+N BACKUP DEVICE. NO INFORMATION ABOUT INJURIES OR SURGICAL DELAYS WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10825980 ↗
  • MalfunctionMay 29, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT DURING PROCEDURE, WHEN THE CAP WAS SCREWED ON THE NAIL IMPLANTED ON THE PATIENT THE CAP BEGAN TO CRACK ON THE HEAD UNTIL IT WAS BROKEN. DELAY WAS NOT SPECIFIED. NO INJURY REPORTED. DOCTOR DECIDED TO LEAVE THE NAIL WITHOUT CAP.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8651734 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K210980, K981529. One submission often covers many models, so check the product and lot details in each record.

  • Z-2456-2026 · Mar 16, 2026 · Open, Classified
    TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
  • Z-1806-2025 · Mar 18, 2025 · Open, Classified
    META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
  • Z-2443-2023 · Jul 24, 2023 · Open, Classified
    TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT, REF 71647336, Femur nail
  • Z-2444-2023 · Jul 24, 2023 · Open, Classified
    TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT, REF 71647340, Femur nail
  • Z-0943-2013 · Feb 7, 2013 · Terminated
    INTERNAL HEX RECON SCREW, TI-6AL-4V, REF 7162380, TRIGEN (R) INTERNAL HEX CAPTURED RECON SCREW, QTY: (1), 6.4 MM, 80 MM, TITANIUM, Smith & Nephew, Inc. Orthopedic. Bone screw for
  • Z-0944-2013 · Feb 7, 2013 · Terminated
    INTERNAL HEX RECON SCREW, TI-6AL-4V, REF 7162375, TRIGEN (R) INTERNAL HEX CAPTURED RECON SCREW, QTY: (1), 6.4 MM, 75 MM, TITANIUM, Smith & Nephew, Inc. Orthopedic. Bone screw for

Other NAIL, FIXATION, BONE devices

All NAIL, FIXATION, BONE devices →