TRIGEN 71751133Recalls & reported problems
Smith & Nephew, Inc. · HUMERAL NAIL IMPACTOR · Class 2
All models of TRIGEN (381)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 03596010515445), received between Aug 3, 2020 and Feb 17, 2026 — classified as 2 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K122170); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture3
- Break2
- Crack1
- Detachment of Device or Device Component1
- Mechanical Jam1
- Mechanics Altered1
- Solder Joint Fracture1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 17, 2026Fracture
IT WAS REPORTED THAT DURING AN INTERNAL FIXATION SURGERY, TWO (2) HUMERAL NAIL IMPACTORS BROKE WHILE TRYING TO REMOVE A NAIL FROM A HUMERUS. NO PIECES FELL INSIDE THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24366603 ↗MalfunctionNov 25, 2025Fracture; Fracture
IT WAS REPORTED THAT, DURING SURGERY, THE THREAD OF ONE (1) HUMERAL NAIL IMPACTOR FRACTURED INSIDE ONE (1) STRAIGHT NAIL GUIDE BOLT WHILE IMPACTING. NO PIECES FELL INTO THE PATIENT'S WOUND. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23645570 ↗MalfunctionApr 24, 2025Mechanics Altered; Mechanics Altered
IT WAS REPORTED THAT DURING AN INTERNAL FIXATION SURGERY, WHEN ONE (1) HUMERAL NAIL IMPACTOR WAS ATTACHED TO THE NAIL FIXATION SCREW, IT DID NOT FULLY ENGAGE. THE DOCTOR RETROIMPACTED THE NAIL, RESULTING IN THE THREAD BREAKING, LEAVING PART OF IT INSIDE THE FIXATION SCREW. THE BROKEN PART WAS REMOVED AND THE NAIL WAS IMPLANTED WITH THE SAME IMPACT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21900923 ↗MalfunctionApr 25, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING A OSTEOSYNTHESIS PROCEDURE, UPON IMPACTION ONE (1) HUMERAL NAIL IMPACTOR BROKE AT THE THREADED PART AND REMAINED INSIDE THE COUPLING SCREW OF THE NAIL. NO BROKEN PIECE ENTERED THE PATIENT'S BODY. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, BY PASSING THE NAIL MANUALLY. PATIENT IS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19183693 ↗MalfunctionApr 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING THE OSTEOSYNTHESIS PROCEDURE, THE PROXIMAL PIECE OF THE HUMERAL NAIL IMPACTOR DETACHED IN RETRO IMPACT. ADDITIONALLY, THE DRILL BITS WERE IN POOR CONDITION. SURGERY WAS PERFORMED AFTER A DELAY GREATER THAN 30 MIN, CHANGING SURGICAL TECHNIQUE. PATIENT'S CURRENT HEALTH STATUS IS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19156136 ↗MalfunctionJan 3, 2022Crack; Unintended Movement; Crack; Unintended Movement
IT WAS REPORTED THAT DURING AN INSPECTION, IT WAS NOTICED THAT THE HUMERAL NAIL IMPACTOR WAS SEEN TO BE CRACKED AT THE WELD AND TURNS FREELY. NO CASE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13146243 ↗MalfunctionNov 22, 2021Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT, AFTER A HUMERAL NAILING SURGERY, IT APPEARED THAT THE END OF THE HUMERAL NAIL IMPACTOR WAS STUCK IN THE STRAIGHT NAIL GUIDE BOLT. SURGERY HAD BEEN COMPLETED, WITHOUT ANY DELAY. IT IS UNKNOWN HOW THE PROCEDURE WAS FINISHED. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12855931 ↗InjuryJan 13, 2021Break; Solder Joint Fracture; Break; Solder Joint Fracture
IT WAS REPORTED THAT, DURING A TRAUMA SURGERY, A HUMERAL NAIL IMPACTOR BROKE WHILE INSTALLING THE NAIL. PROCEDURE WAS FINISHED WITH THE SAME DEVICE. NO DELAY REPORTED. NO HARM OR INJURY REPORTED ON PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11165668 ↗InjuryAug 3, 2020Break; Break
IT WAS REPORTED THAT THE HUMERAL NAIL IMPACTOR END BROKE OFF DURING IMPACTION. NO DELAY. A S&N BACK UP WAS AVAILABLE TO COMPLETE THE PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10360100 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K122170. One submission often covers many models, so check the product and lot details in each record.
- Z-0351-2020 · Oct 7, 2019 · TerminatedLAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted sha
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page