TRIGEN 71701020RRecalls & reported problems
Smith & Nephew, Inc. · TRIGEN HINDFOOT FUSION NAIL 10MM X 20CM RIGHT · Class 2 · Implant
All models of TRIGEN (381)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03596010542496), received between Nov 13, 2018 and Dec 13, 2024 — classified as 4 injuries.
The most frequently reported specific device problem is “Fracture” (2 reports).
11 other recalls cover devices cleared under the same FDA submission (K043052, K210980); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Adverse Event Without Identified Device or Use Problem1
- Break1
Patient problems
- Failure of Implant2
- Implant Pain2
- Injury1
- Medical device site infection1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryDec 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, FOLLOWING A RIGHT ANKLE ARTHRODESIS PERFORMED ON (B)(6) 2020, THE PATIENT WAS ADMITTED TO THE HOSPITAL FROM 22 THROUGH (B)(6) 2020 FOR OSTEOMYELITIS OF THE RIGHT LOWER EXTREMITY WITH PSEUDOMONAS INFECTION. THIS ADVERSE EVENT WAS TREATED WITH ANTIBIOTICS. THIS INFORMATION WAS COLLECTED RETROSPECTIVELY AS PART OF A POST-MARKET CLINICAL FOLLOW UP (PMCF) ACTIVITY AND IS PROVIDED IN AN ANONYMIZED FORMAT; THEREFORE, NO FURTHER INFORMATION AVAILABLE.
Patient problems: Implant Pain; Medical device site infection
MAUDE report 20925123 ↗InjurySep 21, 2023Fracture; Fracture
IT WAS REPORTED THAT, FOLLOWING A RIGHT HINDFOOT ARTHRODESIS PROCEDURE PERFORMED ON (B)(6) 2023, THE NAIL BROKE AT THE DISTAL LOCKING END. THE PATIENT UNDERWENT A REVISION ON (B)(6) 2023 AT THE SCHULTHESSKLINIK ZÜRICH. THE CURRENT PATIENT HEALTH STATUS IS UNKNOWN.
Patient problems: Failure of Implant
MAUDE report 17789973 ↗InjuryJun 23, 2022Fracture; Fracture
IT WAS REPORTED THAT AFTER A AFTER A HINDFOOT ARTHRODESIS PERFORMED ON AN UNKNOWN DATE, EVEN THOUGH THE SURGICAL FUSION SITE AT TIBIAL TALUS JUNCTION HAD HEALED WELL, THE PATIENT REPORTED PAIN IN THE NAIL AND APPEARS FRACTURED AT NON FUSED JOINT. A REVISION SURGERY WAS PERFORMED TO TREAT THE ADVERSE EVENT, AND DURING SURGERY IT WAS CONFIRMED THAT THE HFN 10MMX20CM RIGHT BROKE IN SITU THROUGH CALCANEUS/ TALUS SCREW HOLE. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Failure of Implant; Implant Pain
MAUDE report 14789123 ↗InjuryNov 13, 2018Break; Break
IT WAS REPORTED THAT A REVISION WAS PERFORMED DUE TO ANKLE PAIN. TRIGEN NAIL IMPLANTED WAS FOUND BROKEN. NO DELAY OR BACK UP REPORTED.
Patient problems: Injury
MAUDE report 8066568 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K043052, K210980. One submission often covers many models, so check the product and lot details in each record.
- Z-2456-2026 · Mar 16, 2026 · Open, ClassifiedTRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
- Z-1806-2025 · Mar 18, 2025 · Open, ClassifiedMETA-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
- Z-2444-2023 · Jul 24, 2023 · Open, ClassifiedTRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT, REF 71647340, Femur nail
- Z-2443-2023 · Jul 24, 2023 · Open, ClassifiedTRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT, REF 71647336, Femur nail
- Z-2087-2008 · Apr 22, 2008 · TerminatedTRIGEN Hind Foot Fusion Nail, 11.5mm X 20mm Right, REF 71701120R. The device is indicated for degeneration, deformity, or trauma of both the tibiotalar and talocalcaneal articulat
- Z-2085-2008 · Apr 22, 2008 · TerminatedTRIGEN Hind Foot Fusion Nail, 10mm X 20mm Left, REF 71701020L. The device is indicated for degeneration, deformity, or trauma of both the tibiotalar and talocalcaneal articulation
- Z-2082-2008 · Apr 22, 2008 · TerminatedTRIGEN Hind Foot Fusion Nail, 10mm X 25mm Left, REF 71701025L. The device is indicated for degeneration, deformity, or trauma of both the tibiotalar and talocalcaneal articulation
- Z-2088-2008 · Apr 22, 2008 · TerminatedTRIGEN Hind Foot Fusion Nail, 11.5mm X 25mm Right, REF 71701125R. The device is indicated for degeneration, deformity, or trauma of both the tibiotalar and talocalcaneal articulat
- Z-2086-2008 · Apr 22, 2008 · TerminatedTRIGEN Hind Foot Fusion Nail, 10mm X 16mm Left, REF 71701016L. The device is indicated for degeneration, deformity, or trauma of both the tibiotalar and talocalcaneal articulation
- Z-2084-2008 · Apr 22, 2008 · TerminatedTRIGEN Hind Foot Fusion Nail, 11.5mm X 20mm Left, REF 71701120L. The device is indicated for degeneration, deformity, or trauma of both the tibiotalar and talocalcaneal articulati
- Z-2083-2008 · Apr 22, 2008 · TerminatedTRIGEN Hind Foot Fusion Nail, 11.5mm X 25mm Left, REF 71701125L. The device is indicated for degeneration, deformity, or trauma of both the tibiotalar and talocalcaneal articulati
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page