Skip to content

TRIGEN 71700004Recalls & reported problems

Smith & Nephew, Inc. · TRIGEN HINDFOOT FUSION NAIL DROP · Class 2

All models of TRIGEN (381)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 03596010543226), received between Mar 30, 2021 and Oct 8, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Failure to Align” (4 reports).

1 other recall cover devices cleared under the same FDA submission (K122170); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Align4
  • Fracture1
  • Loose or Intermittent Connection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 8, 2025Failure to Align; Failure to Align

    IT WAS REPORTED THAT, DURING AN INTRAMEDULLARY NAIL SURGERY, THE SPECIALIST CHECKED THE NAIL LOCKS BEFORE SECURING THEM TO THE NAIL. THEY ARE ALIGNED BY PERFORMING THE FRAME CLICKS ACCORDING TO THE NEEDS OF THE CX. IT WAS NOTICED THAT ONE OF THE BLOCKAGES OF ONE (1) HFN DROP WAS LEFT OUT WHEN THE DRILL BIT WAS PASSED, SO, THE DRILL BIT HIT THE NAIL, WHICH IS WHY THE REPORTED DEVICE WAS DETERMINED TO BE DECALIBRATED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23245704 ↗
  • MalfunctionDec 27, 2022Fracture; Fracture

    IT WAS REPORTED THAT, DURING MALLEATING, A HFN DROP SNAPPED. SURGERY WAS RESUMED, WITH A SIGNIFICANT DELAY, WITH A SMITH AND NEPHEW BACK-UP DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16050311 ↗
  • MalfunctionApr 6, 2021Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT, DURING AN UNKNOWN SURGERY, THE SCREW ON TOP FELL OUT AND NEEDS TO BE REPLACED. NOTHING FELL INTO PATIENT. NO DELAY IN SURGERY. BACKUP WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11624934 ↗
  • MalfunctionApr 2, 2021Failure to Align; Failure to Align

    IT WAS REPORTED THAT DURING SURGERY, THE HFN DRILL GUIDE AND HFN NAIL DROP WERE NOT TARGETING THE DYNAMIC INTERLOCK FOR THE 20CM NAIL. THE PROCEDURE CONTINUED USING THE SAME DEVICES AND NO INJURY TO THE PATIENT WAS REPORTED. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11610606 ↗
  • MalfunctionApr 2, 2021Failure to Align; Failure to Align

    IT WAS REPORTED THAT, DURING USE, OUTSIDE THE PATIENT, THERE WAS TROUBLE WITH THE NAIL DROP AND DRILL GUIDE NOT TARGETING THE DYNAMIC INTERLOCK FOR THE 20CM NAIL. NO OTHER COMPLICATION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11610600 ↗
  • MalfunctionMar 30, 2021Failure to Align; Failure to Align

    IT WAS REPORTED THAT DURING A PROCEDURE THE HFN DRILL GUIDE AND HFN DROP WERE NOT TARGETING CORRECTLY. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE AND WITH NO SIGNIFICANT DELAYS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11593388 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K122170. One submission often covers many models, so check the product and lot details in each record.

  • Z-0351-2020 · Oct 7, 2019 · Terminated
    LAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted sha

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page