Skip to content

JOURNEY 71461015Recalls & reported problems

Smith & Nephew, Inc. · JOURNEY PFJ IMPLANT MEDIUM RIGHT · Class 2 · Implant

All models of JOURNEY (566)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03596010544223), received between Dec 17, 2018 and Apr 7, 2024 — classified as 4 injuries.

The most frequently reported specific device problem is “Biocompatibility” (1 report).

1 other recall cover devices cleared under the same FDA submission (K051086, K093056, K230653); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Biocompatibility1
  • Break1
  • Crack1
  • Fracture1

Patient problems

  • Arthralgia2
  • Failure of Implant2
  • Fall1
  • High metal ion levels1
  • Hypersensitivity/Allergic reaction1
  • Injury1
  • Joint effusion1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 7, 2024Fracture; Fracture

    IT WAS REPORTED THAT, AFTER A RIGHT PATELLOFEMORAL SURGERY WAS PERFORMED ON (B)(6) 2017, THE PATIENT EXPERIENCED INCREASE IN PAIN. X-RAYS WERE PERFORMED ON (B)(6) 2023 AND THEY WERE UNREMARKABLE. HOWEVER, ON (B)(6) 2023 X-RAYS REVEALED THERE WAS A VISIBLE FRACTURE OF THE TROCHLEA AND A POSTERIOR RUPTURE ON THE SURFACE OF THE JOURNEY PF IMPLANT MED RT. THIS EVENT WAS ADDRESSED BY PERFORMING A REVISION SURGERY ON (B)(6) 2024, DURING WHICH THE FRACTURED FEMORAL COMPONENT WAS REMOVED. PATIENT'S HEALTH STATUS IS GOOD. REPLACEMENT OPERATION WENT WELL.

    Patient problems: Failure of Implant

    MAUDE report 19057131 ↗
  • InjuryOct 10, 2023Biocompatibility; Biocompatibility

    IT WAS REPORTED THAT, AFTER A CONVENTIONAL PATELLOFEMORAL PROSTHESIS AND PATELLA FROM ANOTHER SUPPLIER HAD BEEN IMPLANTED ON (B)(6) 2021, THE PATIENT EXPERIENCED PURIFORM ASEPTIC EFFUSION, SEVERE PAIN AND FUNCTIONAL DIFFICULTIES A FEW MONTHS LATER, THEREFORE, ALLERGY TEST REVEALED ALLERGIES TO CHROMIUM, COBALT, NICKEL AND, TO A LESSER EXTENT, IRON AND MANGANESE (NO ALLERGIES TO CEMENT, TITANIUM, ZIRCONIUM OR NIOBIUM WERE FOUND). AS A CONSEQUENCE, A REVISION SURGERY WAS PERFORMED ON (B)(6) 2021 IN WHICH THE COMPETITOR DEVICE WAS EXPLANTED AND A JOURNEY PF IMPLANT MED RT WAS IMPLANTED. ADDITIONALLY, PROGRESS APPEARED FAVORABLE UNTIL A MECHANICAL FALL FIVE MONTHS LATER RESULTED IN PROFUSE EFFUSION, HOWEVER, THE RADIOLOGICAL EXAMINATION REVEALED NO ABNORMALITIES IN IMPLANT POSITIONING OR FIXATION. AFTER RULING OUT SEPTIC ARTHRITIS, THE DIAGNOSIS OF ALLERGY WAS ADOPTED. FURTHERMORE, IT IS SUSPECTED A RELEASE OF CHROMIUM OR NICKEL METAL IONS. THE PATIENT HAD UNDERGONE CORTICOSTEROID INFILTRATION PRIOR TO IMPLANT REPLACEMENT AFTER INFECTION MANAGEMENT. THE RESULT WAS FAVORABLE FOR THE EFFUSION, BUT NOT VERY LONG-LASTING. THE PATIENT SHOWED DISCRETE IMPROVEMENT DESPITE A STILL ABUNDANT EFF

    Patient problems: Arthralgia; Fall; High metal ion levels; Hypersensitivity/Allergic reaction; Joint effusion

    MAUDE report 17907274 ↗
  • InjuryAug 13, 2021Crack; Crack

    IT WAS REPORTED THAT, AFTER PFJ RECONSTRUCTION SURGERY HAD BEEN PERFORMED ON (B)(6) 2016, THE PATIENT EXPERIENCED PAIN IN THE RIGHT KNEE AFTER A FALL. THIS ADVERSE EVENT WAS SOLVED BY CONVERTING TO TKR ON (B)(6) 2021. IT WAS NOTICED THAT THE JOURNEY PF IMPLANT MED RT WAS CRACKED AND THE GEN II 7.5MM RESUR PAT 32MM WAS ERODED. PATIENT LEFT SURGERY IN GOOD HEALTH.

    Patient problems: Arthralgia; Failure of Implant

    MAUDE report 12320890 ↗
  • InjuryDec 17, 2018Break; Break

    IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED FOR A REPLACEMENT OF THE PARTIAL ENDOPROSTHESIS DUE TO A PROGRESSION OF OSTEOARTHRITIS IN THE RIGHT KNEE. IT WAS DISCOVERED THE COMPONENT WAS BROKEN.

    Patient problems: Injury

    MAUDE report 8172067 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K051086, K093056, K230653. One submission often covers many models, so check the product and lot details in each record.

  • Z-2621-2023 · Aug 31, 2023 · Open, Classified
    smith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint

Other PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates