TRIGEN 71653030Recalls & reported problems
Smith & Nephew, Inc. · TRIGEN META RETROGRADE FEMORAL NAIL 10MM X 30CM · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03596010548160), received between Jun 5, 2020 and Dec 15, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Fracture1
Patient problems
- Injury1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 15, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING AN INTERNAL FIXATION SURGERY, THE META RETRO FEM NAIL 10 X 30 BECAME STUCK IN THE PROXIMAL FEMUR BONE UPON IMPACT. WHILE ATTEMPTING TO REMOVE IT FOR FURTHER MILLING, THE IMPACT ON THE GUIDE CAUSED IT TO BREAK AT THE SCREW THREAD. FINALLY, IT WAS REMOVED BY IMPACTING IT IN THE OPPOSITE DIRECTION WITH PRESSURE PLIERS. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A COMPETITOR'S DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23802785 ↗InjuryJun 5, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A REVISION SURGERY TO REMOVE TRIGEN RETROGRADE FEMORAL NAIL (INITIAL INSERTION DATE (B)(6) 2019, REVISION SURGERY DATE (B)(6) 2020) THE THREADS OF THE 2 PROXIMAL INTERNAL HEX CAPTURE SCREWS DISASSEMBLED DURING REMOVAL AND FURTHERMORE FELL INTO THE MEDULLARY CANAL. THE SURGEON HAD A HARD TIME TO REMOVE THEM AND THE SURGERY WAS DELAYED BY MORE THAN 30 MINS. THERE WAS NO IMPACT ON THE PATIENT.
Patient problems: Injury
MAUDE report 10122902 ↗
Other NAIL, FIXATION, BONE devices
- NA 03.043.024Synthes GmbH · NAIL, FIXATION, BONE234 reports
- NA 03.010.440Synthes GmbH · NAIL, FIXATION, BONE70 reports
- TRIGEN 71676627Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 recall2 reports
- TRIGEN 71676628Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 recall0 reports
- NA 03.043.018Synthes GmbH · NAIL, FIXATION, BONE33 reports
- TRIGEN 71675383Smith & Nephew, Inc. · NAIL, FIXATION, BONE29 reports
- NA 03.010.441Synthes GmbH · NAIL, FIXATION, BONE27 reports
- NA 03.010.435Synthes GmbH · NAIL, FIXATION, BONE24 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗