TRIGEN 71674062Recalls & reported problems
Smith & Nephew, Inc. · 16MM CHANNEL REAMER · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 03596010560575), received between Apr 28, 2021 and May 21, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Incomplete or Inadequate Connection1
- Loose or Intermittent Connection1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 21, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, ONE (1) 16MM CHANNEL REAMER CAME APART. IT IS UNKNOWN IF BROKEN PIECES FELL INTO THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO PATIENT INJURIES HAVE BEEN REPORTED AS CONSEQUENCE OF THIS PROBLEM
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19367609 ↗MalfunctionMay 6, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT, DURING A TRIGEN INTERTAN NAIL PROCEDURE, THE HOLDING PIN OF ONE (1) 16MM CHANNEL REAMER BECAME LOOSE AND FELL OUT, THEREFORE, THE REAMER CAME APART. ALL PARTS WERE RETRIEVED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16883116 ↗MalfunctionApr 28, 2021Loose or Intermittent Connection; Incomplete or Inadequate Connection; Loose or Intermittent Connection; Incomplete or Inadequate Connection
IT WAS REPORTED THAT DURING SET UP/INSPECTION FOR IM NAIL SURGERY THE 16MM CHANNEL REAMER WAS LOOSE AND DID NOT ATTACH TO THE 12.5 REAMER. THE SURGERY CONTINUED WITHOUT ANY SIGNIFICANT DELAY AND WAS FINISHED WITH A SMITH AND NEPHEW BACK UP DEVICE. NO PATIENT INJURIES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11742458 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.
- Z-0730-2019 · Jan 7, 2019 · TerminatedSL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
- Z-1008-2017 · Nov 16, 2016 · TerminatedREFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116
Other Prosthesis, hip, femoral component, cemented, metal devices
- GENESIS II 71440005Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal599 reports
- NA 71364073Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal436 reports
- Peri-Loc 71173377Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal401 reports
- R3 71368569Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal398 reports
- NA 71364006Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal197 reports
- NA 71364011Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal176 reports
- Accord 71360046Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal171 reports
- REFLECTION 71362279Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal155 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.