TRIGEN 71674070Recalls & reported problems
Smith & Nephew, Inc. · 7.0MM COMPRESSION SCREW STARTER DRILL · Class 2
All models of TRIGEN (381)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Smith & Nephew Trigen compression screw starter drill is a surgical instrument used to start holes for screws during bone fixation surgery. Most of the 71 reports describe the drill tip breaking off or snapping during use, sometimes leaving a fragment inside the patient that has to be retrieved, which in a few cases delayed the operation. There are no deaths, and 28 reports mention injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 71
- Deaths reported
- 0
- Injuries reported
- 28
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 71 adverse event reports that list this exact device (DI 03596010560636), received between Oct 27, 2017 and Apr 8, 2026 — classified as 28 injuries, 43 malfunctions.
The most frequently reported specific device problem is “Fracture” (44 reports).
1 other recall cover devices cleared under the same FDA submission (K122170); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture44
- Break24
- Adverse Event Without Identified Device or Use Problem1
- Detachment of Device or Device Component1
- Material Fragmentation1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions59
- No Known Impact Or Consequence To Patient5
- Foreign Body In Patient3
- No Consequences Or Impact To Patient2
- No Information2
- Bone Fracture(s)1
Recent adverse event reports
The 25 most recent of 71 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 8, 2026Fracture
IT WAS REPORTED THAT, DURING A TRAUMA SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE OFF WHILE IN USE, ALL THE BROKEN PIECES WERE REMOVED AND ACCOUNTED. THE PROCEDURE WAS COMPLETED, AFTER A NON-SIGNIFICANT DELAY, WITH THE SAME S+N PRODUCT. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24820851 ↗MalfunctionMar 20, 2026Fracture
IT WAS REPORTED THAT DURING A TRAUMA SURGERY, THE TIP OF THE 7.0 COMPRESSION SCREW DRILL BROKE OFF WHILE IN USE. THE TIP WAS RETRIEVED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY. IT IS UNKNOWN HOW WAS COMPLETED THE SURGERY. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24649544 ↗MalfunctionMar 18, 2026Fracture
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE OFF WHILE THE DEVICE WAS IN USE, AND THE TIP WAS RETRIEVED. IT IS UNKNOWN HOW THE PROCEDURE WAS PERFORMED; HOWEVER, THERE WAS A 2-HOUR SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24626549 ↗MalfunctionFeb 4, 2026Fracture
IT WAS REPORTED THAT, DURING A FEMUR INTRAMEDULLARY NAILING SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE WHILE IN USE. THE BROKEN TIP WAS RETRIEVED BUT IT IS UNKNOWN HOW. THE PROCEDURE WAS RESUMED, AFTER A 30 MINUTE DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24259323 ↗MalfunctionJan 19, 2026Fracture
IT WAS REPORTED THAT DURING A TRAUMA SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE WHILE IN USE AND FELL INSIDE PATIENT. THE PIECE WAS RETRIEVED AND ACCOUNTED FOR. THE PROCEDURE WAS COMPLETED AFTER A 20 MINUTES DELAY USING A SIMILAR S+N BACK UP. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24108986 ↗MalfunctionNov 3, 2025Fracture; Fracture
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE OFF AFTER DRILLING, JUST BEFORE USING THE NEXT DRILL. THE BROKEN TIP WAS RETRIEVED. THE PROCEDURE WAS RESUMED, AFTER A 10 MINUTES DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23451281 ↗MalfunctionJun 12, 2025Fracture; Fracture
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE. THE PIECE WAS RETRIEVED. IT IS UNKNOWN HOW THE PIECE WAS RETRIEVED. THERE IS NOT SURGICAL REPORTS, OPERATIVE NOTES NOR X-RAYS AVAILABLE. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22199839 ↗MalfunctionMar 14, 2025Fracture; Fracture
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL WAS BROKEN OFF WHILE IN USE. ALL PIECES WERE RETRIEVED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21606700 ↗MalfunctionJan 29, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING AN INTERNAL FIXATION SURGERY, THE TIP OF A 7.0 COMPRESSION SCREW DRILL BROKE OFF WHILE IN USE. PIECES WERE REMOVED FROM THE PATIENT WITH AN UNKNOWN METHOD. THERE WAS A NON-SIGNIFICANT DELAY AND IT IS UNKNOWN HOW THE PROCEDURE WAS FINISHED. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21264851 ↗MalfunctionJan 6, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING A TRAUMA SURGERY, THE TIP ON THE INTERTAN 7.0MM COMP SCREW STARTER DRILL BROKE WHILE IN USE. THE PIECE WAS RETRIEVED BUT IT IS UNKNOWN HOW. THERE WAS A NON-SIGNIFICANT DELAY AND THE PROCEDURE WAS FINISHED USING THE SAME DEVICE. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21080463 ↗MalfunctionNov 6, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING INTERTAN NAIL SURGERY, UPON USING ONE (1) 7.0 COMPRESSION SCREW DRILL, THE TAPERED PART OF THE DRILL, BROKE OFF INSIDE OF THE PATIENT. THE BROKEN PIECE WAS RECOVERED FROM THE PATIENT BY THE SURGEON USING A LONG HEMOSTAT. THE BROKEN TIP WAS DISCARDED BY THE SCRUB NURSE WITH THE REST OF THE SHARPS AT THE END OF THE CASE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20622324 ↗MalfunctionSep 27, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A TRAUMA SURGERY, THE TIP OF THE 7.0 COMPRESSION SCREW DRILL BROKE OFF AND PIECES FELL INSIDE THE PATIENT'S WOUND, THE BROKEN PIECE WAS RETRIEVED BUT IT IS UNKNOWN HOW. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20328119 ↗MalfunctionSep 19, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING AN ORIF SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE OFF. THE PIECE WAS RETRIEVED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING A S+N BACKUP DEVICE. NO INJURY WAS REPORTED AS A RESULT OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20260785 ↗MalfunctionJun 12, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING AN INTERNAL FIXATION SURGERY, THERE WAS A NOISE WHILE USING THE 7.0 COMPRESSION SCREW DRILL AND IT WAS NOTICED THAT THE TIP HAD BROKEN OFF, THE PIECES WERE RETRIEVED, BUT THE METHOD IS UNKNOWN. THERE WAS A NON-SIGNIFICANT DELAY AND THE PROCEDURE WAS FINISHED USING A SMITH & NEPHEW BACK UP. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19520366 ↗MalfunctionJun 11, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING TRAUMA, INTERTAN SURGERY, THE TIP ON ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE WHILE IN USE. THE PIECE WAS RETRIEVED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19510551 ↗MalfunctionApr 22, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING AN ORIF FEMUR PROCEDURE, DRILLING WAS PERFORMED WITH ONE (1) 7.0 COMPRESSION SCREW DRILL. THEN, WHEN DRILLING WITH THE LAG SCREW DRILL, THE DRILL COULD NOT ADVANCE TO THE NAIL DUE TO METAL INTERFERENCE. THE NAIL WAS REMOVED AND CHECKED, AND IT WAS NOTICED THAT THE TIP OF THE COMPRESSION SCREW DRILL HAD FRACTURED AT THE LAG SCREW HOLE. THE BROKEN TIP WAS REMOVED WITH MIKULICZ' FORCEPS. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY (10 MIN), WITH THE SAME DEVICE. PATIENT WAS NOT HARMED AS A CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19154486 ↗MalfunctionApr 11, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A INTERNAL FIXATION SURGERY , THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL WAS BROKEN OFF WHILE IN USE AFTER NOISE. THE PIECE WAS RETRIEVED; HOWEVER, IT IS UNKNOWN HOW. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY WITH A SMITH + NEPHEW BACK UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19094032 ↗MalfunctionMar 26, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE OFF WHILE IN USE. THE PIECE WAS RETRIEVED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18977092 ↗InjuryMar 4, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION PROCEDURE, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE OFF INTRAOPERATIVELY AND COULD NOT BE RETRIEVED. THE PIECE WAS LEFT IN THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENTS' CURRENT HEALTH STATUS IS UNKNOWN.
Patient problems: Foreign Body In Patient
MAUDE report 18826919 ↗MalfunctionFeb 19, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE OFF. THE BROKEN PIECE WAS RETRIEVED. THE PROCEDURE WAS PERFORMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18732286 ↗InjuryFeb 14, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION PROCEDURE, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE OFF WHILE IN USE. THE PIECE WAS RETRIEVED. ADDITIONALLY, THERE WAS A FRACTURE LINE ON THE BONE WHICH WAS FIXED USING A CABLE. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH AN S+N BACK-UP DEVICE. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Bone Fracture(s)
MAUDE report 18707419 ↗MalfunctionJan 30, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING SURGERY, THE TIP OF THE SCREW DRILL WAS BROKEN WHILE IN USE. THE BROKEN TIP WAS RETRIEVED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18607755 ↗MalfunctionJan 15, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING TRAUMA SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL WAS BROKEN INSIDE THE PATIENT WHILE IN USE. THE BROKEN TIP WAS RETRIEVED AND THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18512488 ↗MalfunctionNov 14, 2023Fracture; Fracture
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, JUST AFTER THE DRILLING WAS STARTED THE TIP OF A 7.0 COMPRESSION SCREW DRILL WAS BROKEN, THE PIECE WAS RETRIEVED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18133867 ↗MalfunctionNov 13, 2023Fracture; Fracture
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, WHEN DRILLING THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL WAS BROKEN. THE PIECE FELL INSIDE THE WOUND WAS RETRIEVED AND ACCOUNTED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18122606 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K122170. One submission often covers many models, so check the product and lot details in each record.
- Z-0351-2020 · Oct 7, 2019 · TerminatedLAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted sha
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page