TRIGEN 71674070Recalls & reported problems

Smith & Nephew, Inc. · 7.0MM COMPRESSION SCREW STARTER DRILL · Class 2

All models of TRIGEN (381)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Smith & Nephew Trigen compression screw starter drill is a surgical instrument used to start holes for screws during bone fixation surgery. Most of the 71 reports describe the drill tip breaking off or snapping during use, sometimes leaving a fragment inside the patient that has to be retrieved, which in a few cases delayed the operation. There are no deaths, and 28 reports mention injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
71
Deaths reported
0
Injuries reported
28
Recalls naming it
0

Summary

The FDA's MAUDE database contains 71 adverse event reports that list this exact device (DI 03596010560636), received between Oct 27, 2017 and Apr 8, 2026 — classified as 28 injuries, 43 malfunctions.

The most frequently reported specific device problem is “Fracture” (44 reports).

1 other recall cover devices cleared under the same FDA submission (K122170); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 2 reportsMar 2024: 2 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 2 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture44
  • Break24
  • Adverse Event Without Identified Device or Use Problem1
  • Detachment of Device or Device Component1
  • Material Fragmentation1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions59
  • No Known Impact Or Consequence To Patient5
  • Foreign Body In Patient3
  • No Consequences Or Impact To Patient2
  • No Information2
  • Bone Fracture(s)1

Recent adverse event reports

The 25 most recent of 71 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 8, 2026Fracture

    IT WAS REPORTED THAT, DURING A TRAUMA SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE OFF WHILE IN USE, ALL THE BROKEN PIECES WERE REMOVED AND ACCOUNTED. THE PROCEDURE WAS COMPLETED, AFTER A NON-SIGNIFICANT DELAY, WITH THE SAME S+N PRODUCT. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24820851 ↗
  • MalfunctionMar 20, 2026Fracture

    IT WAS REPORTED THAT DURING A TRAUMA SURGERY, THE TIP OF THE 7.0 COMPRESSION SCREW DRILL BROKE OFF WHILE IN USE. THE TIP WAS RETRIEVED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY. IT IS UNKNOWN HOW WAS COMPLETED THE SURGERY. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24649544 ↗
  • MalfunctionMar 18, 2026Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE OFF WHILE THE DEVICE WAS IN USE, AND THE TIP WAS RETRIEVED. IT IS UNKNOWN HOW THE PROCEDURE WAS PERFORMED; HOWEVER, THERE WAS A 2-HOUR SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24626549 ↗
  • MalfunctionFeb 4, 2026Fracture

    IT WAS REPORTED THAT, DURING A FEMUR INTRAMEDULLARY NAILING SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE WHILE IN USE. THE BROKEN TIP WAS RETRIEVED BUT IT IS UNKNOWN HOW. THE PROCEDURE WAS RESUMED, AFTER A 30 MINUTE DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24259323 ↗
  • MalfunctionJan 19, 2026Fracture

    IT WAS REPORTED THAT DURING A TRAUMA SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE WHILE IN USE AND FELL INSIDE PATIENT. THE PIECE WAS RETRIEVED AND ACCOUNTED FOR. THE PROCEDURE WAS COMPLETED AFTER A 20 MINUTES DELAY USING A SIMILAR S+N BACK UP. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24108986 ↗
  • MalfunctionNov 3, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE OFF AFTER DRILLING, JUST BEFORE USING THE NEXT DRILL. THE BROKEN TIP WAS RETRIEVED. THE PROCEDURE WAS RESUMED, AFTER A 10 MINUTES DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23451281 ↗
  • MalfunctionJun 12, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE. THE PIECE WAS RETRIEVED. IT IS UNKNOWN HOW THE PIECE WAS RETRIEVED. THERE IS NOT SURGICAL REPORTS, OPERATIVE NOTES NOR X-RAYS AVAILABLE. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22199839 ↗
  • MalfunctionMar 14, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL WAS BROKEN OFF WHILE IN USE. ALL PIECES WERE RETRIEVED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21606700 ↗
  • MalfunctionJan 29, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING AN INTERNAL FIXATION SURGERY, THE TIP OF A 7.0 COMPRESSION SCREW DRILL BROKE OFF WHILE IN USE. PIECES WERE REMOVED FROM THE PATIENT WITH AN UNKNOWN METHOD. THERE WAS A NON-SIGNIFICANT DELAY AND IT IS UNKNOWN HOW THE PROCEDURE WAS FINISHED. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21264851 ↗
  • MalfunctionJan 6, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A TRAUMA SURGERY, THE TIP ON THE INTERTAN 7.0MM COMP SCREW STARTER DRILL BROKE WHILE IN USE. THE PIECE WAS RETRIEVED BUT IT IS UNKNOWN HOW. THERE WAS A NON-SIGNIFICANT DELAY AND THE PROCEDURE WAS FINISHED USING THE SAME DEVICE. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21080463 ↗
  • MalfunctionNov 6, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING INTERTAN NAIL SURGERY, UPON USING ONE (1) 7.0 COMPRESSION SCREW DRILL, THE TAPERED PART OF THE DRILL, BROKE OFF INSIDE OF THE PATIENT. THE BROKEN PIECE WAS RECOVERED FROM THE PATIENT BY THE SURGEON USING A LONG HEMOSTAT. THE BROKEN TIP WAS DISCARDED BY THE SCRUB NURSE WITH THE REST OF THE SHARPS AT THE END OF THE CASE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20622324 ↗
  • MalfunctionSep 27, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING A TRAUMA SURGERY, THE TIP OF THE 7.0 COMPRESSION SCREW DRILL BROKE OFF AND PIECES FELL INSIDE THE PATIENT'S WOUND, THE BROKEN PIECE WAS RETRIEVED BUT IT IS UNKNOWN HOW. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20328119 ↗
  • MalfunctionSep 19, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN ORIF SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE OFF. THE PIECE WAS RETRIEVED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING A S+N BACKUP DEVICE. NO INJURY WAS REPORTED AS A RESULT OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20260785 ↗
  • MalfunctionJun 12, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING AN INTERNAL FIXATION SURGERY, THERE WAS A NOISE WHILE USING THE 7.0 COMPRESSION SCREW DRILL AND IT WAS NOTICED THAT THE TIP HAD BROKEN OFF, THE PIECES WERE RETRIEVED, BUT THE METHOD IS UNKNOWN. THERE WAS A NON-SIGNIFICANT DELAY AND THE PROCEDURE WAS FINISHED USING A SMITH & NEPHEW BACK UP. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19520366 ↗
  • MalfunctionJun 11, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING TRAUMA, INTERTAN SURGERY, THE TIP ON ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE WHILE IN USE. THE PIECE WAS RETRIEVED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19510551 ↗
  • MalfunctionApr 22, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN ORIF FEMUR PROCEDURE, DRILLING WAS PERFORMED WITH ONE (1) 7.0 COMPRESSION SCREW DRILL. THEN, WHEN DRILLING WITH THE LAG SCREW DRILL, THE DRILL COULD NOT ADVANCE TO THE NAIL DUE TO METAL INTERFERENCE. THE NAIL WAS REMOVED AND CHECKED, AND IT WAS NOTICED THAT THE TIP OF THE COMPRESSION SCREW DRILL HAD FRACTURED AT THE LAG SCREW HOLE. THE BROKEN TIP WAS REMOVED WITH MIKULICZ' FORCEPS. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY (10 MIN), WITH THE SAME DEVICE. PATIENT WAS NOT HARMED AS A CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19154486 ↗
  • MalfunctionApr 11, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING A INTERNAL FIXATION SURGERY , THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL WAS BROKEN OFF WHILE IN USE AFTER NOISE. THE PIECE WAS RETRIEVED; HOWEVER, IT IS UNKNOWN HOW. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY WITH A SMITH + NEPHEW BACK UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19094032 ↗
  • MalfunctionMar 26, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE OFF WHILE IN USE. THE PIECE WAS RETRIEVED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18977092 ↗
  • InjuryMar 4, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION PROCEDURE, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE OFF INTRAOPERATIVELY AND COULD NOT BE RETRIEVED. THE PIECE WAS LEFT IN THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENTS' CURRENT HEALTH STATUS IS UNKNOWN.

    Patient problems: Foreign Body In Patient

    MAUDE report 18826919 ↗
  • MalfunctionFeb 19, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE OFF. THE BROKEN PIECE WAS RETRIEVED. THE PROCEDURE WAS PERFORMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18732286 ↗
  • InjuryFeb 14, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION PROCEDURE, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL BROKE OFF WHILE IN USE. THE PIECE WAS RETRIEVED. ADDITIONALLY, THERE WAS A FRACTURE LINE ON THE BONE WHICH WAS FIXED USING A CABLE. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH AN S+N BACK-UP DEVICE. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Bone Fracture(s)

    MAUDE report 18707419 ↗
  • MalfunctionJan 30, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING SURGERY, THE TIP OF THE SCREW DRILL WAS BROKEN WHILE IN USE. THE BROKEN TIP WAS RETRIEVED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18607755 ↗
  • MalfunctionJan 15, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING TRAUMA SURGERY, THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL WAS BROKEN INSIDE THE PATIENT WHILE IN USE. THE BROKEN TIP WAS RETRIEVED AND THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18512488 ↗
  • MalfunctionNov 14, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, JUST AFTER THE DRILLING WAS STARTED THE TIP OF A 7.0 COMPRESSION SCREW DRILL WAS BROKEN, THE PIECE WAS RETRIEVED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18133867 ↗
  • MalfunctionNov 13, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, WHEN DRILLING THE TIP OF ONE (1) 7.0 COMPRESSION SCREW DRILL WAS BROKEN. THE PIECE FELL INSIDE THE WOUND WAS RETRIEVED AND ACCOUNTED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18122606 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K122170. One submission often covers many models, so check the product and lot details in each record.

  • Z-0351-2020 · Oct 7, 2019 · Terminated
    LAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted sha

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →