TRIGEN 71674071Recalls & reported problems
Smith & Nephew, Inc. · 7/16 GUIDE BOLT · Class 2
All models of TRIGEN (381)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 03596010560643), received between Nov 26, 2018 and Sep 28, 2021 — classified as 2 injuries, 9 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (4 reports).
1 other recall cover devices cleared under the same FDA submission (K122170); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Deformation4
- Break2
- Connection Problem2
- Failure to Align2
- Naturally Worn2
- Mechanical Jam1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjurySep 28, 2021Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT, DURING TRAUMA SURGERY, THE SURGEON WAS UNABLE TO UNSCREW A TRIGEN 7/16 GUIDE BOLT AND REMOVE THE ALIGNMENT DRILL GUIDE. EXCESSIVE FORCE WAS USED TO LOOSEN THE TRIGEN 7/16 GUIDE BOLT, WHICH LED TO BREAKAGE OF A TRIGEN GUIDE BOLT WRENCH. THE TIP OF THE TRIGEN GUIDE BOLT WRENCH GOT STUCK INSIDE THE TRIGEN 7/16 GUIDE BOLT. THE ISSUE WAS SOLVED BY DRILLING A HOLE THROUGH THE TRIGEN 7/16 GUIDE BOLT WITH A METAL BUR. THE ALIGNMENT GUIDE WAS REMOVED WITH A BACK-UP GUIDE BOLT WRENCH. A SIGNIFICANT DELAY WAS CAUSED AS CONSEQUENCE OF THIS PROBLEM. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12543934 ↗MalfunctionSep 10, 2021Failure to Align; Material Deformation; Failure to Align; Material Deformation
IT WAS REPORTED THAT A DRILL GUIDE HANDLE IS WARPED AND ITS TARGETING IS OFF AND THAT A 7/16 GUIDE BOLT IS CROSS-THREADED AND DOES NOT ATTACH PROPERLY. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12456448 ↗MalfunctionAug 11, 2021Failure to Align; Failure to Align
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE DISTAL SCREW HOLES DID NOT ALIGN PROPERLY. IT IS UNKNOWN IF THE 7/16 GUIDE BOLT, 125 RAD DRILL GDE DROP, 130 RAD DRILL GDE DROP OR DRILL GUIDE HANDLE CAUSED THE MISALIGNMENT. PROCEDURE WAS COMPLETED WITH A BACK-UP DEVICE WITHOUT ANY DELAY. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12307696 ↗MalfunctionMay 24, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING NAIL CASE, THE 7/16 GUIDE BOLT WAS STRIPPED. A SMITH AND NEPHEW BACK UP DEVICE WAS AVAILABLE. NO INJURY TO PATIENT OR DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11871833 ↗MalfunctionApr 28, 2021Break; Naturally Worn; Break; Naturally Worn
IT WAS REPORTED THAT AFTER AN INTERTAN NAIL INSERTION PROCEDURE, IT WAS NOTICED THAT THE 7/16 GUIDE BOLT THREADS ARE WEARING AND BREAKING DOWN. THIS INCIDENT WAS REALIZED AFTER THE PROCEDURE; HENCE, SURGERY WAS COMPLETED WITHOUT DELAY USING THE SAME DEVICE. NO OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11742454 ↗InjuryApr 21, 2021Connection Problem; Connection Problem
IT WAS REPORTED THAT, DURING SURGERY, THE TIP 7/16 GUIDE BOLT CROSS-THREADED INTO THE NAIL RESULTING IN NOT BEING ABLE TO RELEASE THE NAIL FROM THE INTERTAN DRILL GUIDE. SURGERY WAS COMPLETED BY USING A COMPETITOR DEVICE. SURGERY WAS NOT SIGNIFICANTLY DELAYED. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11699934 ↗MalfunctionMar 24, 2021Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE END OF THE 7/16 GUIDE BOLT WHICH IS TIGHTENED USING A GUIDE BOLT WRENCH IS WORN INTO A CIRCLE INSTEAD OF A HEXAGONAL SHAPE, SO THE WRENCH DOESN'T NOT CATCH TO TIGHTEN THE BOLT. NO CASE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11566422 ↗MalfunctionMar 19, 2021Connection Problem; Connection Problem
IT WAS REPORTED THAT DURING SURGERY, THE 7/16 GUIDE BOLT WAS NOT LOCKING INTO THE IMPLANT. PROCEDURE CONTINUED USING A BACKUP DEVICE FROM SMITH AND NEPHEW, WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES AND NO INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11522600 ↗MalfunctionNov 24, 2020Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING THE PROCEDURE, THE GUIDE BOLT HEAD WAS STRIPPED. THIS CAUSED A DELAY BETWEEN 0-30 MIN AND IT WAS NECESSARY TO USED A BACKUP FROM SMITH & NEPHEW TO FINISH THE SURGERY. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10892350 ↗MalfunctionNov 5, 2020Material Deformation; Material Deformation
IT WAS REPORTED THAT IN A 7/16 GUIDE BOLT, THREADS ARE STRIPPED. NO HARM OR INJURY REPORTED ON PATIENT. PROCEDURE WAS FINISHED WITH A SMITH AND NEPHEW BACK UP DEVICE. NO DELAY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10791087 ↗MalfunctionNov 26, 2018Break; Break
IT WAS REPORTED THAT DURING SURGERY THE GUIDE BOLT WRENCH HEAD SNAPPED OFF INSIDE THE GUIDE BOLT. ALL PIECES ACCOUNTED FOR. DELAY OF OVER 2 HOURS REPORTED. NO INJURY REPORTED. PROCEDURE FINISHED WITH COMPETITOR DEVICE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8104357 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K122170. One submission often covers many models, so check the product and lot details in each record.
- Z-0351-2020 · Oct 7, 2019 · TerminatedLAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted sha
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page