TRIGEN 71674092Recalls & reported problems
Smith & Nephew, Inc. · ENTRY PORTAL HANDLE · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03596010560834), received between Aug 11, 2021 and Apr 18, 2023 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break4
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 18, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING THE BEGINNING OF A TRAUMA SURGERY, THE ENTRY PORTAL HANDLE DEVICE WAS IMPACTED AND THE IMPACTION AND MANIPULATION HANDLE FRACTURED, SLOWING DOWN THE PROCEDURE, THE BROKEN PIECE WAS REMOVED AND REPLACED BY THE 3.2MM THREADED PIN. THE PROCEDURE WAS RESUMED, AFTER A SIGNIFICANT DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16765418 ↗MalfunctionFeb 23, 2022Break; Break
IT WAS REPORTED THAT THE ENTRY PORTAL IS BROKEN AT SHAFT AND HEAD CONNECTION. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT ANY PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13583497 ↗MalfunctionNov 10, 2021Break; Break
IT WAS REPORTED THAT, DURING INTERNAL FIXATION SURGERY, THE TUBE A TRIGEN ENTRY PORTAL HANDLE BECAME LODGED IN HANDLE COUPLING AND BROKE OUTSIDE THE PATIENT. SURGERY WAS RESUMED WITH A SMITH & NEPHEW BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM. IT IS UNKNOWN IF SURGERY WAS DELAYED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12791771 ↗MalfunctionOct 19, 2021Break; Break
IT WAS REPORTED THAT A ENTRY PORTAL HANDLE IS BROKEN. AS THIS WAS NOTICED UPON INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12656954 ↗MalfunctionAug 11, 2021Break; Break
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION PROCEDURE, A GRIPPER AND A ENTRY PORTAL HANDLE BROKE. NO FURTHER DETAILS AVAILABLE AT THIS TIME. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12307393 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.
- Z-0730-2019 · Jan 7, 2019 · TerminatedSL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
- Z-1008-2017 · Nov 16, 2016 · TerminatedREFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116
Other Prosthesis, hip, femoral component, cemented, metal devices
- GENESIS II 71440005Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal599 reports
- NA 71364073Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal436 reports
- Peri-Loc 71173377Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal401 reports
- R3 71368569Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal398 reports
- NA 71364006Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal197 reports
- NA 71364011Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal176 reports
- Accord 71360046Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal171 reports
- REFLECTION 71362279Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal155 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page