TRIGEN 71675224Recalls & reported problems
Smith & Nephew, Inc. · TRIGEN INTERTAN 10MM X 36CM 125DEGREE RIGHT · Class 2 · Implant
All models of TRIGEN (381)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03596010561091), received between May 16, 2017 and Jan 18, 2022 — classified as 2 injuries.
The most frequently reported specific device problem is “Fracture” (2 reports).
6 other recalls cover devices cleared under the same FDA submission (K040212); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fracture2
- Insufficient Device Problem Information1
Patient problems
- Ambulation or Postural Difficulties1
- Discomfort1
- Failure of Implant1
- Injury1
- Loss of Range of Motion1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJan 18, 2022Fracture; Fracture
IT WAS REPORTED THAT, AFTER AN INTERNAL FIXATION SURGERY HAD BEEN PERFORMED ON (B)(6) 2015 IN ORDER TO TREAT A COMMINUTED FEMUR FRACTURE, THE PATIENT EXPERIENCED A RUPTURE OF THE INTERTAN 10MM X 36CM 125D RT USED. FUNCTIONAL DISCOMFORT WITH PERSISTENT LIMP AND PROXIMAL RIGHT HIP IMPINGEMENT WERE REPORTED. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2021 IN ORDER TO REMOVE THE BROKEN CENTROMEDULLARY NAIL. ALSO, AN ISOLATED BONE GRAFTING WAS PERFORMED AS THE FRACTURE WAS CONSOLIDATED. THE DEVICE IS NOT AVAILABLE FOR ANALYSIS. THE CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Ambulation or Postural Difficulties; Discomfort; Failure of Implant; Loss of Range of Motion
MAUDE report 13282441 ↗InjuryMay 16, 2017Fracture; Insufficient Device Problem Information; Fracture; Insufficient Device Problem Information
IT WAS REPORTED A REVISION SURGERY WAS PERFORMED LESS THAN FOUR MONTHS AFTER SURGERY AFTER THE PATIENT FELT A SNAP AND WAS UNABLE TO WEIGHT-BEAR.
Patient problems: Injury
MAUDE report 6570969 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K040212. One submission often covers many models, so check the product and lot details in each record.
- Z-1133-2022 · Apr 22, 2022 · Open, ClassifiedINTERTAN 1.5 NAIL 13mm X 42cm 130DEG. Left INTERTAN 1.5 NAIL 13mm X 42cm 130DEG. Right
- Z-0772-2021 · Nov 2, 2020 · TerminatedBone fixation nail. TRIGEN INTERTAN - Product Usage: indicated for fractures of the femur including simple shaft fractures, comminuted shaft fractures, spiral shaft fractures, long
- Z-0097-2018 · Oct 24, 2017 · Open, Classifiedsmith&nephew TRIGEN INTERTAN 10S, REF 71675384, 10 MM X 18 CM NAIL, 130o (degrees), SET SCREW PREASSEMBLED IN NAIL
- Z-2258-2014 · Jul 2, 2014 · TerminatedTRIGEN(R) INTERTAN(TM) 10MM X 18 CM NAIL, 125o, REF 71675201, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 18 CM NAIL, 125o, REF 71
- Z-1636-2014 · Apr 4, 2014 · TerminatedTRIGEN INTERTAN NAIL: (1) REF 71675201, 10 MM X 18 CM, 125 degree, (2) REF 71675202, 11.5 MM X 18 CM, 125 degree, (3) REF 71675203, 13 MM X 18 CM, 125 degree, (4) REF 71675204,
- Z-1637-2014 · Apr 4, 2014 · TerminatedTRIGEN INTERTAN 1.5 NAIL: (1) REF 71676515, 10 MM X 10 CM, 125 degree, LEFT, (2) REF 71676556, 11.5 MM X 36 CM, 125 degree, RIGHT, (3) REF 71676558, 11.5 MM X 38 CM, 125 degree
Other NAIL, FIXATION, BONE devices
- NA 03.043.024Synthes GmbH · NAIL, FIXATION, BONE234 reports
- NA 03.010.440Synthes GmbH · NAIL, FIXATION, BONE70 reports
- TRIGEN 71676627Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 recall2 reports
- TRIGEN 71676628Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 recall0 reports
- NA 03.043.018Synthes GmbH · NAIL, FIXATION, BONE33 reports
- TRIGEN 71675383Smith & Nephew, Inc. · NAIL, FIXATION, BONE29 reports
- NA 03.010.441Synthes GmbH · NAIL, FIXATION, BONE27 reports
- NA 04.043.240SSynthes GmbH · NAIL, FIXATION, BONE24 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page