TRIGEN 71677090Recalls & reported problems
Smith & Nephew, Inc. · INTERTAN LAG/COMPRESSION SCREW KIT 90MM / 85MM · Class 2
All models of TRIGEN (380)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 03596010563309), received between Nov 22, 2019 and Nov 16, 2023 — classified as 7 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
10 other recalls cover devices cleared under the same FDA submission (K040212, K210980); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Break1
- Defective Device1
- Device Dislodged or Dislocated1
- Fracture1
- Inadequacy of Device Shape and/or Size1
- Incomplete or Inadequate Connection1
- Material Disintegration1
Patient problems
- Failure of Implant5
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient2
- Anemia1
- Bone Fracture(s)1
- Deformity/ Disfigurement1
- Injury1
- Insufficient Information1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryNov 16, 2023Fracture; Fracture
IT WAS REPORTED THAT, A PLAINTIFF UNDERWENT A LEFT INTERTAN FEMORAL NAIL PROCEDURE, ON (B)(6) 2023, DUE TO A FALL AND SUBSEQUENT SUB TROCHANTERIC FRACTURE. THEY LATER DEVELOPED A VARUS DEFORMITY AND WERE DIAGNOSED WITH A SUB TROCHANTERIC / INTERTROCHANTERIC FEMUR NONUNION WITH FAILED FIXATION, AND THE INTERTAN 1.5 11.5MMX38CM 125D LT NAIL AND THE LAG/COMP SCREW KIT 90/85 WERE FOUND TO BE FRACTURED. THEREFORE, THEY WERE ELECTED FOR A REVISION SURGERY ON (B)(6) 2023, IN WHICH THE NAIL WAS EXPLANTED AND A SYNTHES COMPETITOR DEVICE WAS IMPLANTED. PATIENT WAS TAKEN TO PACU IN STABLE CONDITION, NO OTHER COMPLICATIONS WERE REPORTED.
Patient problems: Anemia; Deformity/ Disfigurement; Failure of Implant; Muscle Weakness/Atrophy; Nonunion/Delayed-union Bone Fracture; Pain
MAUDE report 18156709 ↗InjurySep 29, 2022Positioning Failure; Positioning Failure
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE LAG SCREW OF A INTERTAN LAG/COMPRESSION SCREW KIT 90MM / 85MM PULLED BACK BEFORE THE COMPRESSION SCREW SEATED. THE SURGEON REMOVED BOTH COMPONENTS OF THE INTERTAN LAG/COMPRESSION SCREW KIT 90MM / 85MM AND RE-DRILLED, BUT IT DID NOT HELP. SURGERY WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH THE SAME DEVICE. PATIENT WAS NOT HARMED BEYOND THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15502153 ↗InjuryJul 11, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, AFTER AN INTERNAL FIXATION WAS PERFORMED ON (B)(6) 2022, THE PATIENT EXPERIENCED EARLY COMPRESSION, CAUSING THE LAG/COMP SCREW KIT 90/85 TO BEGIN PULLING FROM THE HEAD OF THE FEMUR. THIS EVENT WAS TREATED BY A REVISION SURGERY ON (B)(6) 2022, IN WHICH THE INTERSTRAIN 10S 10MM X 18CM 130D WAS REMOVED. DURING THE REMOVAL, THE TRIGEN LOW PROFILE SCREW 5.0MM X 30MM STRIPPED OUT. SEVERAL TECHNIQUES WERE USED FOR THE REMOVAL, INCLUDING DRIVING THE SCREW TO THE FAR CORTEX, FRACTURING THE FEMUR. CURRENT HEALTH STATUS OF PATIENT IS UNKNOWN.
Patient problems: Bone Fracture(s); Failure of Implant; Insufficient Information
MAUDE report 14974285 ↗InjuryAug 31, 2021Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT, AFTER AN INTERNAL FIXATION SURGERY HAD BEEN PERFORMED ON (B)(6) 2020, POST-OP X-RAYS SHOWED THAT THE LAG/COMP SCREW KIT 90/85 OF AN INTERTAN 1.5 11.5MMX40CM 125D RT NAIL ARE BACKING OUT, DESPITE BEING LOCKED IN THE CASE. IT IS UNKNOWN HOW THIS ADVERSE EVENT WAS SOLVED. CURRENT HEALTH STATUS OF PATIENT IS ALSO UNKNOWN.
Patient problems: Failure of Implant
MAUDE report 12399958 ↗InjuryAug 31, 2021Device Dislodged or Dislocated; Migration; Device Dislodged or Dislocated; Migration
IT WAS REPORTED THAT, AFTER AN INTERTROCHANTERIC FRACTURE CORRECTION SURGERY WAS PERFORMED ON (B)(6) 2021, THE LAG/COMP SCREW KIT 90/85, THAT WAS ORIGINALLY INSERTED INTO A INTERTAN 10S 10MM X 20CM 130D, FAILED DUE TO A CUT OUT AND THE SCREWS BACKING OUT DESPITE BEING LOCKED IN THE CASE. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2021 TO TREAT THIS ADVERSE EVENT. THE PATIENT'S OUTCOME IS UNKNOWN.
Patient problems: Failure of Implant
MAUDE report 12394906 ↗MalfunctionMay 10, 2021Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT, DURING A TRAUMA HIP SURGERY THE SCRUB TECH INSERTED THE WRONG CAPTURE STICK INTO THE LAG SCREW DRIVER AND NOW IT¿S STUCK ON THE LAG/COMP SCREW KIT 90/85 AND WON'T COME OUT. IT WAS UNKNOWN HOW THE PROCEDURE WAS FINISHED AND IF THERE WAS A SURGICAL DELAY. NO PATIENT INJURIES OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11797484 ↗InjuryJan 2, 2021Break; Break
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO THE NAIL BREAKAGE. NO DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11103619 ↗InjuryJun 1, 2020Material Disintegration; Defective Device; Material Disintegration; Defective Device
IT WAS REPORTED THAT DURING AN INTERTAN CASE, AS THE COMPRESSION SCREW WAS BEING INSERTED UNDER THE LAG SCREW, ONCE COMPRESSION WAS BEING APPLIED, THE SCREW WOULD TIGHTEN AND THEN BOUNCE BACKWARDS. THE SCREW WAS REMOVED FROM THE PATIENT, AND INSPECTED TO SEE SMALL METAL SHAVINGS ON THE THREADS OF THE SCREW. BOTH SCREWS WERE REMOVED AND REPLACED WITH A NEW INTERLOCKING SCREW KIT, WHICH FUNCTIONED AS NORMAL AND COMPRESSED THE FRACTURE. A DELAY WAS PRESENTED LESS OR EQUAL TO 30 MINUTES. IT IS UNKNOWN THE STATE OF THE PATIENT.
Patient problems: Failure of Implant; Injury
MAUDE report 10105413 ↗MalfunctionFeb 14, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING SURGERY, INSERTED THE COMPRESSION SCREW AND TRACTION WAS REMOVED COMPLETELY AND WHEN COMPRESSION SCREW WAS ALMOST ALL THE WAY DOWN, SURGEON SWITCHED TO THE T HANDLE AND THEN TO BE FINISHED BY HAND. THE SCREW KEPT GETTING TIGHT, BUT DID NOT SHOW ANY SIGN OF COMPRESSING. THEN ALL OF A SUDDEN THE COMPRESSION SCREW WOULD BECOME LOOSE AGAIN. PROCEDURE WAS COMPLETED WITH A NEW SET OF SCREWS THAT WERE INSERTED ONLY BY HAND.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9714950 ↗MalfunctionDec 29, 2019Incomplete or Inadequate Connection; Incomplete or Inadequate Connection
IT WAS REPORTED THAT DURING SURGERY, THE COMPRESSION SCREW DID NOT ENTER THE BROCADE GUIDE AND SHOWED REBOUND TO ENTER THE SLIDING SCREW. AT THE END, NO COMPRESSION SCREW IS PLACED, DISTAL LOCKING WAS NOT PERFORMED, NAIL INSERTION ARCH WAS REMOVED, THE SLIDING SCREW BLOCKED AND 0 MM PLUG PLACEMENT WAS PERFORMED, WHICH ALSO DOES NOT FIT PROPERLY INTO THE INTERTAN INTRAMEDULLARY NAIL. NO INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9528704 ↗MalfunctionNov 22, 2019Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
IT WAS REPORTED THAT DURING SURGERY, AFTER INSERTING THE LAG/COMP SCREW, IT WAS FOUND THAT THE LAG SCREW WAS LONGER THAN EXPECTED. BACKUP DEVICE FROM SMITH AND NEPHEW WAS AVAILABLE. NO HARM TO PATIENT. DELAY OF LESS THAN 30 MINUTES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9363696 ↗
FDA clearance and classification
Cleared through 510(k) K210980 on Nov 14, 2022 (decision: Substantially Equivalent). Submitted by Smith & Nephew, Inc. as “Smith & Nephew Intramedullary Nail Systems”. The FDA review took 592 days. FDA record ↗
Cleared through Special 510(k) K040212 on Feb 20, 2004 (decision: Substantially Equivalent). Submitted by Smith & Nephew, Inc. as “Trigen Intertan”. The FDA review took 21 days. FDA record ↗
The FDA regulates product code JDS (nail, fixation, bone) under 21 CFR 888.3030 as a Class II device. 21 CFR 888.3030 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K040212, K210980. One submission often covers many models, so check the product and lot details in each record.
- Z-2456-2026 · Mar 16, 2026 · Open, ClassifiedTRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
- Z-1806-2025 · Mar 18, 2025 · Open, ClassifiedMETA-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
- Z-2443-2023 · Jul 24, 2023 · Open, ClassifiedTRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT, REF 71647336, Femur nail
- Z-2444-2023 · Jul 24, 2023 · Open, ClassifiedTRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT, REF 71647340, Femur nail
- Z-1133-2022 · Apr 22, 2022 · Open, ClassifiedINTERTAN 1.5 NAIL 13mm X 42cm 130DEG. Left INTERTAN 1.5 NAIL 13mm X 42cm 130DEG. Right
- Z-0772-2021 · Nov 2, 2020 · TerminatedBone fixation nail. TRIGEN INTERTAN - Product Usage: indicated for fractures of the femur including simple shaft fractures, comminuted shaft fractures, spiral shaft fractures, long
- Z-0097-2018 · Oct 24, 2017 · Open, Classifiedsmith&nephew TRIGEN INTERTAN 10S, REF 71675384, 10 MM X 18 CM NAIL, 130o (degrees), SET SCREW PREASSEMBLED IN NAIL
- Z-2258-2014 · Jul 2, 2014 · TerminatedTRIGEN(R) INTERTAN(TM) 10MM X 18 CM NAIL, 125o, REF 71675201, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 18 CM NAIL, 125o, REF
- Z-1636-2014 · Apr 4, 2014 · TerminatedTRIGEN INTERTAN NAIL: (1) REF 71675201, 10 MM X 18 CM, 125 degree, (2) REF 71675202, 11.5 MM X 18 CM, 125 degree, (3) REF 71675203, 13 MM X 18 CM, 125 degree, (4) REF 71675204,
- Z-1637-2014 · Apr 4, 2014 · TerminatedTRIGEN INTERTAN 1.5 NAIL: (1) REF 71676515, 10 MM X 10 CM, 125 degree, LEFT, (2) REF 71676556, 11.5 MM X 36 CM, 125 degree, RIGHT, (3) REF 71676558, 11.5 MM X 38 CM, 125 degree
Common questions about the TRIGEN 71677090
Has the TRIGEN 71677090 been recalled?
No FDA recall record in our data names this exact device identifier (DI 03596010563309), based on the FDA recall file dated Nov 27, 2026. However, 10 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the TRIGEN 71677090?
The FDA's MAUDE database lists 11 adverse event reports for this device, received between Nov 22, 2019 and Nov 16, 2023. In those reports, the most common specific device problems are “Break” (1), “Defective Device” (1) and “Device Dislodged or Dislocated” (1); the most common patient problems are “Failure of Implant” (5), “Anemia” (1) and “Bone Fracture(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the TRIGEN 71677090?
None of the reports are classified as deaths, 7 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the TRIGEN 71677090 FDA approved?
It was cleared through Special 510(k) K040212 on Feb 20, 2004; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the TRIGEN 71677090, and is it still sold?
The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand TRIGEN, model 71677090). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other NAIL, FIXATION, BONE devices
- NA 03.043.024Synthes GmbH · NAIL, FIXATION, BONE233 reports
- NA 03.010.440Synthes GmbH · NAIL, FIXATION, BONE69 reports
- TRIGEN 71676627Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 recall2 reports
- TRIGEN 71676628Smith & Nephew, Inc. · NAIL, FIXATION, BONE1 recall0 reports
- NA 03.043.018Synthes GmbH · NAIL, FIXATION, BONE33 reports
- TRIGEN 71675383Smith & Nephew, Inc. · NAIL, FIXATION, BONE29 reports
- NA 03.010.441Synthes GmbH · NAIL, FIXATION, BONE27 reports
- NA 03.010.435Synthes GmbH · NAIL, FIXATION, BONE23 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page