Skip to content

NA 71563436Recalls & reported problems

Smith & Nephew, Inc. · 4.8 MM QUICK CONNECT DRILL BIT · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 03596010573261), received between Dec 9, 2019 and May 22, 2025 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3
  • Break2
  • Insufficient Device Problem Information1
  • Physical Resistance/Sticking1
  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Information1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 22, 2025Fracture; Separation Failure; Fracture; Separation Failure

    IT WAS REPORTED THAT, DURING PFJ SURGERY, ONE (1) 4.8 MM QUICK CONNECT DRILL BIT WAS PLACED IN THE WRONG HOLE OF THE JOURNEY PFJ IM DRILL GUIDE, IT COLD WELDED CAUSING THE DEVICE TO SNAPBACK AGAINST THE SURGEON AND THE DRILL BIT BROKE. ALL PARTS WERE RECOVERED. ALSO, ONE (1) MINI CONNECTOR WAS FAULTY, NOT LOCKING THE DRILL FULLY. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, IT IS UNKNOWN IF THERE WAS A BACK UP DEVICE AVAILABLE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22069393 ↗
  • MalfunctionMar 4, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING A PATELLOFEMORAL REPLACEMENT (PFJ) SURGERY, ONE (1) 4.8 MM QUICK CONNECT DRILL BIT FRACTURED AT THE CONNECTOR END AFTER COMPLETELY DRILLING THE HOLE. DEVICE WAS CONNECTED TO A CHUCK TO REMOVE IT. THE PROCEDURE WAS COMPLETED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21514598 ↗
  • MalfunctionFeb 19, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING DRILLING IN A PFJ SURGERY, ONE (1) 4.8 MM QUICK CONNECT DRILL BIT SNAPPED IN TWO. NO PIECES FELL INTO THE PATIENT. THE PROCEDURE WAS RESUMED WITHOUT DELAY USING A SMITH AND NEPHEW BACKUP DEVICE, AND NO INJURIES WERE REPORTED AS A RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21413071 ↗
  • MalfunctionDec 27, 2021Break; Break

    IT WAS REPORTED THAT, DURING A TKA SURGERY, A 4.8 MM QUICK CONNECT DRILL BIT BROKE OFF IN FEMUR. ALL PIECES WERE RETRIEVED. SURGERY WAS RESUMED, WITH A NON-SIGNIFICANT DELAY, USING A SN BACK-UP DEVICE. PATIENT WAS NOT HARMED BEYOND THE REPORTED PROBLEM. NO FURTHER COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13087828 ↗
  • InjuryDec 9, 2021Break; Break

    IT WAS REPORTED THAT, DURING TKA SURGERY, A 4.8 MM QUICK CONNECT DRILL BIT SNAPPED OFF IN FEMORAL CANAL. THE DEVICE WAS INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED USING A S+N BACK-UP DEVICE. SURGERY WAS DELAYED FOR GREATER THAN 30 MINUTES. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12972391 ↗
  • MalfunctionAug 25, 2021Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT, DURING A PFA SURGERY, A 4.8 MM QUICK CONNECT DRILL BIT BECAME STUCK IN THE MINI CONNECTOR. SURGERY WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12367641 ↗
  • MalfunctionDec 9, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT DURING SURGERY, 4.8 MM DRILL SEEMED TO ¿CATCH¿ IN THE BONE WHEN DRILLING THE FEMUR AND CAUSED THE POWER TOOL TO ALMOST FLIP OUT OF THE SURGEON¿S HAND. AS A RESULT, THE CONNECTING END OF THE DRILL BIT WAS SEVERELY TWISTED. NO DELAY. NO BACKUP DEVICE AVAILABLE. NO INJURY OR IMPACT TO PATIENT REPORTED.

    Patient problems: No Information

    MAUDE report 9440689 ↗

Other PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page