Skip to content

R3 71366052Recalls & reported problems

Smith & Nephew, Inc. · R3 MINIMALLY INVASIVE SURGERY X-BAR · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Smith & Nephew R3 minimally invasive surgery X-Bar is part of a cemented metal hip prosthesis used in joint replacement. Reports mostly describe the bar snapping or breaking during hip surgery or setup, sometimes leaving the device in two pieces or its weld coming apart; surgeries continued with a backup part and no pieces were left inside patients. There are 33 reports, no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
33
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 03596010580207), received between Mar 26, 2019 and Dec 12, 2025 — classified as 33 malfunctions.

The most frequently reported specific device problem is “Break” (23 reports).

2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break23
  • Material Twisted/Bent3
  • Fracture2
  • Material Deformation2
  • Solder Joint Fracture2
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions29
  • No Patient Involvement4
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 12, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A THA SURGERY, ONE (1) R3 MIS X-BAR BROKE. THE PROCEDURE WAS COMPLETED, AFTER NO DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23789223 ↗
  • MalfunctionFeb 2, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING A THR SURGERY, THE R3 MIS X-BAR SNAPPED. NO PIECES FELL INSIDE THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18630937 ↗
  • MalfunctionMay 18, 2022Break; Break

    IT WAS REPORTED THAT, THAT THE METAL WELD OF AN R3 MIS X-BAR HAS BROKEN, DEVICE IS IN TWO PIECES RATHER THAN ONE UNIT. AS THIS WAS NOTICED IN A NON-SURGICAL ENVIRONMENT, NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14431917 ↗
  • MalfunctionFeb 27, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE BAR OF A R3 MINIMALLY INVASIVE SURGERY X-BAR FELL OUT OF TOP PORTION OF THE DEVICE, DUE TO BAD WELD. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE INSTEAD. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13615289 ↗
  • MalfunctionJan 27, 2022Break; Break

    IT WAS REPORTED THAT, AFTER A THR SURGERY, AN R3 MIS X-BAR BROKE. SINCE INCIDENT OCCURRED AFTER PROCEDURE, PATIENT WAS NOT LONGER INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13365210 ↗
  • MalfunctionJan 6, 2022Break; Break

    IT WAS REPORTED THAT, DURING THR SURGERY, A R3 MIS X-BAR BROKE WHILE OUTSIDE THE PATIENT. NO PIECES FELL INTO PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK UP DEVICE. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13181113 ↗
  • MalfunctionNov 22, 2021Break; Break

    IT WAS REPORTED THAT, DURING A TKA SURGERY SETUP, THE WELD ON A R3 MIS X-BAR BROKE OFF. SURGERY WAS PERFORMED WITH A BACKUP DEVICE, WITHOUT ANY DELAY. AS THIS PROBLEM OCCURRED BEFORE USE IN PATIENT, THE PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12852558 ↗
  • MalfunctionNov 12, 2021Break; Break

    IT WAS REPORTED THAT, DURING SET UP/INSPECTION SURGERY, THE PIECE THAT ATTACHES TO THE R3 MIS X-BAR BROKE OFF. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12805758 ↗
  • MalfunctionNov 10, 2021Break; Break

    IT WAS REPORTED THAT, DURING A THR SURGERY, AN R3 MINIMALLY INVASIVE SURGERY X-BAR BROKE, NO PIECES FELL INSIDE THE PATIENT. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12787906 ↗
  • MalfunctionOct 19, 2021Break; Break

    IT WAS REPORTED THAT, DURING SET UP/INSPECTION FOR A THR PROCEDURE, THE CLIP OF A R3 MIS X-BAR BROKE OFF. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, THE PROCEDURE WAS FINISHED, BUT THE DEVICE WAS NOT USED. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12657358 ↗
  • MalfunctionSep 8, 2021Break; Break

    IT WAS REPORTED THAT ONE R3 MIS X-BAR BROKE DURING WASH. AS THIS WAS NOTICED UPON CLEANING AND STERILIZATION ACTIVITIES, THERE WAS NOT PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12440865 ↗
  • MalfunctionJul 13, 2021Break; Break

    IT WAS REPORTED THAT, DURING A TKA, A R3 MIS X-BAR BROKE. NO PIECES FELL INSIDE THE PATIENT. THE PROCEDURE WAS FINISHED WITHOUT DELAY USING A SMITH AND NEPHEW BACK UP DEVICE. NO INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12158911 ↗
  • MalfunctionJun 22, 2021Break; Break

    IT WAS REPORTED THAT, DURING A THR SURGERY, A R3 MIS X-BAR BROKE OUTSIDE OF THE PATIENT. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12042929 ↗
  • MalfunctionJun 14, 2021Break; Break

    IT WAS REPORTED THAT, DURING A THR, AN R3 MIS X-BAR BROKE OFF OUTSIDE THE PATIENT. NO DELAYS NOR INJURIES WERE REPORTED. PROCEDURE WAS COMPLETED WITH A BACK UP DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11996096 ↗
  • MalfunctionJun 14, 2021Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT AFTER A THR SURGERY, IT WAS NOTICED THAT THE R3 MIS X-BAR WAS BENT. THERE WAS NO SIGNIFICANT DELAY AND THE PROCEDURE WAS FINISHED USING A SMITH & NEPHEW BACK UP. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11991474 ↗
  • MalfunctionMay 28, 2021Break; Break

    IT WAS REPORTED THAT THE END OF THE R3 MIS X-BAR HAS FALLEN OFF. NO CASE INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11904471 ↗
  • MalfunctionMay 27, 2021Break; Break

    IT WAS REPORTED THAT, DURING SURGERY, A MIS X-BAR BROKE. DEVICE OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11898141 ↗
  • MalfunctionApr 26, 2021Break; Break

    IT WAS REPORTED THAT DURING INSPECTION, IT WAS FOUND THAT THE R3 MIS X-BAR WAS BROKEN AT THE BASE INTO TWO SEPARATE PIECES. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11729678 ↗
  • MalfunctionMar 8, 2021Break; Break

    IT WAS REPORTED THAT, DURING THR SURGERY A PIECE OF THE MIS X-BAR BROKE OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A SMITH AND NEPHEW BACK-UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11431625 ↗
  • MalfunctionMar 8, 2021Material Deformation; Material Deformation

    IT WAS REPORTED THAT DURING INSPECTION PROCESS, THE R3 MIS X-BAR WAS NOTICED TO BE BENT. THERE WAS NO CASE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11431079 ↗
  • MalfunctionMar 8, 2021Material Deformation; Material Deformation

    IT WAS REPORTED THAT DURING INSPECTION PROCESS WAS FOUND THAT THE R3 MIS X-BAR IS BENT. THERE WAS NO CASE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11431075 ↗
  • MalfunctionMar 8, 2021Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT DURING INSPECTION IT WAS NOTICED THAT THE R3 MIS X-BAR IS BENT, THE R3 OFFSET IMPACTOR IS LOCKED UP AND WILL NO RELEASE. NO CASE INVOLVED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11431073 ↗
  • MalfunctionMar 8, 2021Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE R3 MIS X-BAR ARE BENT. NO CASE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11431070 ↗
  • MalfunctionMar 8, 2021Solder Joint Fracture; Solder Joint Fracture

    IT WAS REPORTED THAT, DURING RECEIVING INSPECTION A R3 MIS X-BAR WAS FOUND BROKEN AT THE WELDING SEAM. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11431003 ↗
  • MalfunctionFeb 19, 2021Break; Break

    IT WAS REPORTED THAT, THE R3 MIS X-BAR'S WELD WAS CRACKED AT THE BASE. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11347083 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.

  • Z-0730-2019 · Jan 7, 2019 · Terminated
    SL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
  • Z-1008-2017 · Nov 16, 2016 · Terminated
    REFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116

Other Prosthesis, hip, femoral component, cemented, metal devices

All Prosthesis, hip, femoral component, cemented, metal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page