R3 71366052Recalls & reported problems
Smith & Nephew, Inc. · R3 MINIMALLY INVASIVE SURGERY X-BAR · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Smith & Nephew R3 minimally invasive surgery X-Bar is part of a cemented metal hip prosthesis used in joint replacement. Reports mostly describe the bar snapping or breaking during hip surgery or setup, sometimes leaving the device in two pieces or its weld coming apart; surgeries continued with a backup part and no pieces were left inside patients. There are 33 reports, no deaths and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 33
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 03596010580207), received between Mar 26, 2019 and Dec 12, 2025 — classified as 33 malfunctions.
The most frequently reported specific device problem is “Break” (23 reports).
2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break23
- Material Twisted/Bent3
- Fracture2
- Material Deformation2
- Solder Joint Fracture2
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions29
- No Patient Involvement4
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 12, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING A THA SURGERY, ONE (1) R3 MIS X-BAR BROKE. THE PROCEDURE WAS COMPLETED, AFTER NO DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23789223 ↗MalfunctionFeb 2, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A THR SURGERY, THE R3 MIS X-BAR SNAPPED. NO PIECES FELL INSIDE THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18630937 ↗MalfunctionMay 18, 2022Break; Break
IT WAS REPORTED THAT, THAT THE METAL WELD OF AN R3 MIS X-BAR HAS BROKEN, DEVICE IS IN TWO PIECES RATHER THAN ONE UNIT. AS THIS WAS NOTICED IN A NON-SURGICAL ENVIRONMENT, NO PATIENT WAS INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14431917 ↗MalfunctionFeb 27, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE BAR OF A R3 MINIMALLY INVASIVE SURGERY X-BAR FELL OUT OF TOP PORTION OF THE DEVICE, DUE TO BAD WELD. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE INSTEAD. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13615289 ↗MalfunctionJan 27, 2022Break; Break
IT WAS REPORTED THAT, AFTER A THR SURGERY, AN R3 MIS X-BAR BROKE. SINCE INCIDENT OCCURRED AFTER PROCEDURE, PATIENT WAS NOT LONGER INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13365210 ↗MalfunctionJan 6, 2022Break; Break
IT WAS REPORTED THAT, DURING THR SURGERY, A R3 MIS X-BAR BROKE WHILE OUTSIDE THE PATIENT. NO PIECES FELL INTO PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK UP DEVICE. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13181113 ↗MalfunctionNov 22, 2021Break; Break
IT WAS REPORTED THAT, DURING A TKA SURGERY SETUP, THE WELD ON A R3 MIS X-BAR BROKE OFF. SURGERY WAS PERFORMED WITH A BACKUP DEVICE, WITHOUT ANY DELAY. AS THIS PROBLEM OCCURRED BEFORE USE IN PATIENT, THE PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12852558 ↗MalfunctionNov 12, 2021Break; Break
IT WAS REPORTED THAT, DURING SET UP/INSPECTION SURGERY, THE PIECE THAT ATTACHES TO THE R3 MIS X-BAR BROKE OFF. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12805758 ↗MalfunctionNov 10, 2021Break; Break
IT WAS REPORTED THAT, DURING A THR SURGERY, AN R3 MINIMALLY INVASIVE SURGERY X-BAR BROKE, NO PIECES FELL INSIDE THE PATIENT. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12787906 ↗MalfunctionOct 19, 2021Break; Break
IT WAS REPORTED THAT, DURING SET UP/INSPECTION FOR A THR PROCEDURE, THE CLIP OF A R3 MIS X-BAR BROKE OFF. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, THE PROCEDURE WAS FINISHED, BUT THE DEVICE WAS NOT USED. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12657358 ↗MalfunctionSep 8, 2021Break; Break
IT WAS REPORTED THAT ONE R3 MIS X-BAR BROKE DURING WASH. AS THIS WAS NOTICED UPON CLEANING AND STERILIZATION ACTIVITIES, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12440865 ↗MalfunctionJul 13, 2021Break; Break
IT WAS REPORTED THAT, DURING A TKA, A R3 MIS X-BAR BROKE. NO PIECES FELL INSIDE THE PATIENT. THE PROCEDURE WAS FINISHED WITHOUT DELAY USING A SMITH AND NEPHEW BACK UP DEVICE. NO INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12158911 ↗MalfunctionJun 22, 2021Break; Break
IT WAS REPORTED THAT, DURING A THR SURGERY, A R3 MIS X-BAR BROKE OUTSIDE OF THE PATIENT. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12042929 ↗MalfunctionJun 14, 2021Break; Break
IT WAS REPORTED THAT, DURING A THR, AN R3 MIS X-BAR BROKE OFF OUTSIDE THE PATIENT. NO DELAYS NOR INJURIES WERE REPORTED. PROCEDURE WAS COMPLETED WITH A BACK UP DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11996096 ↗MalfunctionJun 14, 2021Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT AFTER A THR SURGERY, IT WAS NOTICED THAT THE R3 MIS X-BAR WAS BENT. THERE WAS NO SIGNIFICANT DELAY AND THE PROCEDURE WAS FINISHED USING A SMITH & NEPHEW BACK UP. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11991474 ↗MalfunctionMay 28, 2021Break; Break
IT WAS REPORTED THAT THE END OF THE R3 MIS X-BAR HAS FALLEN OFF. NO CASE INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11904471 ↗MalfunctionMay 27, 2021Break; Break
IT WAS REPORTED THAT, DURING SURGERY, A MIS X-BAR BROKE. DEVICE OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11898141 ↗MalfunctionApr 26, 2021Break; Break
IT WAS REPORTED THAT DURING INSPECTION, IT WAS FOUND THAT THE R3 MIS X-BAR WAS BROKEN AT THE BASE INTO TWO SEPARATE PIECES. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11729678 ↗MalfunctionMar 8, 2021Break; Break
IT WAS REPORTED THAT, DURING THR SURGERY A PIECE OF THE MIS X-BAR BROKE OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A SMITH AND NEPHEW BACK-UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11431625 ↗MalfunctionMar 8, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING INSPECTION PROCESS, THE R3 MIS X-BAR WAS NOTICED TO BE BENT. THERE WAS NO CASE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11431079 ↗MalfunctionMar 8, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING INSPECTION PROCESS WAS FOUND THAT THE R3 MIS X-BAR IS BENT. THERE WAS NO CASE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11431075 ↗MalfunctionMar 8, 2021Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING INSPECTION IT WAS NOTICED THAT THE R3 MIS X-BAR IS BENT, THE R3 OFFSET IMPACTOR IS LOCKED UP AND WILL NO RELEASE. NO CASE INVOLVED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11431073 ↗MalfunctionMar 8, 2021Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT THE R3 MIS X-BAR ARE BENT. NO CASE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11431070 ↗MalfunctionMar 8, 2021Solder Joint Fracture; Solder Joint Fracture
IT WAS REPORTED THAT, DURING RECEIVING INSPECTION A R3 MIS X-BAR WAS FOUND BROKEN AT THE WELDING SEAM. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11431003 ↗MalfunctionFeb 19, 2021Break; Break
IT WAS REPORTED THAT, THE R3 MIS X-BAR'S WELD WAS CRACKED AT THE BASE. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11347083 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.
- Z-0730-2019 · Jan 7, 2019 · TerminatedSL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
- Z-1008-2017 · Nov 16, 2016 · TerminatedREFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116
Other Prosthesis, hip, femoral component, cemented, metal devices
- GENESIS II 71440005Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal599 reports
- NA 71364073Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal436 reports
- Peri-Loc 71173377Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal401 reports
- R3 71368569Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal398 reports
- NA 71364006Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal197 reports
- NA 71364011Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal176 reports
- Accord 71360046Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal171 reports
- REFLECTION 71362279Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal155 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page