Skip to content

TRIGEN 71674086Recalls & reported problems

Smith & Nephew, Inc. · SUBTROCHANTERIC LAG SCREW HEXDRIVER ROD · Class 2

All models of TRIGEN (381)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 03596010580290), received between Aug 1, 2019 and May 18, 2026 — classified as 3 injuries, 10 malfunctions.

The most frequently reported specific device problem is “Fracture” (7 reports).

1 other recall cover devices cleared under the same FDA submission (K122170); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture7
  • Break6
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • No Consequences Or Impact To Patient2
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 18, 2026Fracture

    IT WAS REPORTED THAT, DURING AN INTRAMEDULLARY NAIL SURGERY, THE NAIL WAS INSERTED AND THE SUBTROCHANTERIC SCREW PLACEMENT TECHNIQUE WAS PERFORMED, BUT AT THE TIME OF GIVING THE LAST TURNS OF INSERTION, IT WAS NOTICED THAT THE SCREW HANDLE CAME LOOSE. UPON REMOVING THE HANDLE, IT WAS NOTICED THAT ONE (1) SUBTROC LAG SCREW HEXDRVR ROD WAS FRACTURED. NO PIECES FELL INSIDE THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25211759 ↗
  • MalfunctionMay 28, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING THE INTERNAL FIXATION SURGERY, THE SCREW WAS BEING REMOVED AND THE SUBTROC LAG SCREW HEXDRVR ROD BROKE. NOTHING FELL INTO THE INCISION. NO DAMAGE CAUSED, THE PATIENT'S HEALTH CONDITION IS STABLE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22102056 ↗
  • MalfunctionJul 10, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING REMOVAL OF INTERTAN LAG SCREW, THE TIP OF THE SUBTROC LAG SCREW HEXDRVR ROD WAS BROKEN. THE PIECE WAS REMOVED BY AN UNKNOWN METHOD. THE PROCEDURE WAS FINISHED USING THE SAME DEVICE, BUT IT IS UNKNOWN IF THERE WAS A DELAY. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19711143 ↗
  • MalfunctionSep 9, 2022Fracture; Fracture

    IT WAS REPORTED THAT, DURING SURGERY, THE TIP OF A SUBTROC LAG SCREW HEXDRVR ROD WAS BROKEN. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH A S+N BACK-UP DEVICE INSTEAD. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15395810 ↗
  • MalfunctionSep 9, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING SET UP OR INSPECTION, IT WAS NOTICED A BROKEN SUBTROC LAG SCREW DRIVER RETAINING ROD. NO PATIENT WAS INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15393948 ↗
  • InjuryApr 13, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING INTERTAN EXPLANT SURGERY, IT WAS DIFFICULT TO REMOVE THE LAG SCREW. THE INNER ROD OF THE SUBTROC LAG SCREW HEXDRVR ROD WAS FRACTURED. THE IMPLANTED DEVICES COULD NOT BE REMOVED. THE PROCEDURE WAS COMPLETED, WITH A SIGNIFICANT DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14097726 ↗
  • MalfunctionSep 17, 2021Break; Break

    IT WAS REPORTED THAT, THE LISTED DEVICES WERE CRACKED OR BROKEN. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12487910 ↗
  • MalfunctionApr 2, 2021Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT AFTER PROCEDURE, THE TOP PART OF THE SUBTROC LAG SCREW HEXDRIVER ROD WAS NOTICED TO BE BENT AND BROKEN. THE PROCEDURE WAS FINISHED USING A BACKUP DEVICE FROM SMITH AND NEPHEW, WITHOUT SURGICAL DELAY AND NO INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11610659 ↗
  • InjuryMar 4, 2021Break; Break

    IT WAS REPORTED THAT DURING SURGERY, THE THREADS OF INNER CAPTURE FOR SUBTROC LAG SCREW HEXDRIVER ROD SNAPPED DURING EXCISION OF THE NAIL, WHILE USING IT INSIDE THE PATIENT. ALL PIECES WERE REMOVED FROM THE PATIENT AND THE PROCEDURE CONTINUED USING A BACKUP DEVICE FROM SMITH AND NEPHEW, WITHOUT SURGICAL DELAY AND NO INJURY TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11408844 ↗
  • InjuryJun 12, 2020Break; Break

    IT WAS REPORTED THAT THE LAG SCREW CAPTURE ROD BREAK IN SURGERY, IT WAS INSIDE THE PATIENT. IT IS UNKNOWN HOW WAS THE PROCEDURE COMPLETED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10149900 ↗
  • MalfunctionJun 8, 2020Break; Break

    IT WAS REPORTED THAT DURING PROCEDURE SUBTROCHANTERIC DRIVER RETAINING ROD BROKE AT THE THREADING. SMITH& NEPHEW BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10131432 ↗
  • MalfunctionNov 5, 2019Break; Break

    IT WAS REPORTED THAT DURING SURGERY THE TIP OFF THE RETENTION ROD BROKE WHILE COMPRESSING THE LAG SCREW. NO DELAY NOR INJURY WAS REPORTED. A BACK UP WAS AVAILABLE BUT NOT REQUIRED TO FINISH THE SURGERY. NO FRAGMENTS FELL INSIDE THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9282332 ↗
  • MalfunctionAug 1, 2019Fracture; Fracture

    IT WAS REPORTED THAT DURING SURGERY, WHILE TIGHTENING A SCREW, THEY HEARD A POP AND THE ROD BROKE. NO DELAY AND NO INJURY. IT IS UNKNOWN HOW THE PROCEDURE WAS CONCLUDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8850804 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K122170. One submission often covers many models, so check the product and lot details in each record.

  • Z-0351-2020 · Oct 7, 2019 · Terminated
    LAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted sha

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page