Skip to content

TRIGEN 71655135Recalls & reported problems

Smith & Nephew, Inc. · TRIGEN META TIBIAL NAIL 10MM X 35CM · Class 2 · Implant

All models of TRIGEN (380)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 03596010585578), received between May 2, 2023 and Sep 29, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Handling Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Device Handling Problem1

Patient problems

  • Nonunion/Delayed-union Bone Fracture2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjurySep 29, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER A LEFT TIBIAL ORIF PROCEDURE WITH PLACEMENT OF ANTIBIOTIC CEMENT SPACER PERFORMED ON (B)(6) 2019 DUE TO INFECTION, THE PATIENT EXPERIENCED NON-UNION ON AN UNKNOWN DATE. NO FURTHER DETAILS ARE AVAILABLE REGARDING THE MEASURES TAKEN TO TREAT THIS SYMPTOM. THIS INFORMATION WAS COLLECTED RETROSPECTIVELY AS PART OF A POST-MARKET CLINICAL FOLLOW UP (PMCF) ACTIVITY AND IS PROVIDED IN AN ANONYMIZED FORMAT; THEREFORE, NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 20329515 ↗
  • InjurySep 29, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER AN IMN PERFORMED ON (B)(6) 2020 DUE TO A MOTORCYCLE ACCIDENT WHERE THE RIGHT TIBIA FRACTURED, THE PATIENT UNDERWENT AN IMN EXCHANGE WITH ALLOGRAFT ON (B)(6) 2021 DUE TO PAINFUL NON-UNION. THIS INFORMATION WAS COLLECTED RETROSPECTIVELY AS PART OF A POST-MARKET CLINICAL FOLLOW UP (PMCF) ACTIVITY AND IS PROVIDED IN AN ANONYMIZED FORMAT; THEREFORE, NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 20329497 ↗
  • MalfunctionMay 2, 2023Device Handling Problem; Device Handling Problem

    IT WAS REPORTED THAT DURING A TRAUMA SURGERY A SEMI-EXTENDED TECHNIQUE WAS PERFORMED, STARTING WITH CHANNEL RHYME, USING THE SEMIEXTENDED ENTRY TUBE WITH THE CLAMPING HANDLE. THE SURGEON PROCEEDED TO INSERT THE META-NAIL TIBIAL 10MM X 35CM MANUALLY, THEN HE STRONGLY IMPACTED IT WITH THE HAMMER IMPACTOR AND IT WAS FOUND THAT THE HANDLE WAS RELEASED FROM THE GUIDE. THE SURGEON WAS INFORMED AND HE SAID THAT AT THE END HE WOULD REMOVED IT. SALES REP SUGGESTED TO CHECK BEFORE CONTINUING AND IF POSSIBLE REMOVE THE NAIL A LITTLE TO KNOW WHY THE HANDLE CAME LOOSE AND WHY THE GUIDE WAS BEING INTRODUCED, BUT THE SURGEON IGNORED IT AND REPEATED THAT AFTER INSERTING THE NAIL HE WOULD CHECK. EACH TIME IT WAS EVIDENT THAT THE GUIDE DID NOT TOUCH THE PATELLAR REGION, SALES REP SHOWED HIM WHAT HAPPENED AT THE END OF IMPACTING THE NAIL TO ITS POINT, THE SPECIALIST NOTICED THAT THE GUIDE ENTERED THE MEDULLARY CANAL AND ITS REMOVAL WAS NOT SO EASY, HE HIT THE ARC OF CONNECTION OF THE SEMIEXTENDED DRILL GUIDE, WHERE THE NAIL WAS ATTACHED TO TRY TO REMOVE IT, SALES REP SHOWED HIM THAT CAREFULLY BECAUSE THE BLOWS WITH THE HAMMER WERE ON THE LEGS THAT MAKE CONNECTION WITH THE BLACK GUIDE OF PROXIMAL BLOC

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16854451 ↗

FDA clearance and classification

The FDA regulates product code JDS (nail, fixation, bone) under 21 CFR 888.3030 as a Class II device. 21 CFR 888.3030 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the TRIGEN 71655135

Has the TRIGEN 71655135 been recalled?

No FDA recall record in our data names this exact device identifier (DI 03596010585578), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the TRIGEN 71655135?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between May 2, 2023 and Sep 29, 2024. In those reports, the most common specific device problem is “Device Handling Problem” (1); the most common patient problem is “Nonunion/Delayed-union Bone Fracture” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the TRIGEN 71655135?

None of the reports are classified as deaths, 2 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the TRIGEN 71655135 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the TRIGEN 71655135, and is it still sold?

The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand TRIGEN, model 71655135). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other NAIL, FIXATION, BONE devices

All NAIL, FIXATION, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page