Skip to content

TRIGEN 71654506Recalls & reported problems

Smith & Nephew, Inc. · GUIDE BOLT · Class 2

All models of TRIGEN (381)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 03596010585868), received between Sep 6, 2018 and Aug 4, 2025 — classified as 2 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Failure to Align” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K122170); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Align2
  • Fracture2
  • Break1
  • Corroded1
  • Device Damaged by Another Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Device Embedded In Tissue or Plaque1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 4, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, WHEN ONE (1) GUIDE BOLT WAS BEING LOCKED THE SCREW BROKE LEAVING HALF OF THE THREAD IN THE NAIL. THE SCREW PORTION COULD NOT BE REMOVED AND REMAINED IMPLANTED IN THE PATIENT, THEREFORE, THE PROXIMAL PLUG COULD NOT BE INSERTED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH THE PREVIOUSLY ADDRESSED CHANGE IN THE SURGICAL TECHNIQUE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE AND THE PATIENT´S HEALTH CONDITION IS STABLE.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 22709870 ↗
  • InjuryNov 2, 2022Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TIBIAL ORIF, WHILE IMPACTING THE NAIL, A 7/16 GUIDE BOLT BROKE. FURTHERMORE, THE DOCTOR ATTEMPTED TO REMOVE THE BROKEN FRAGMENT WITH SEVERAL INSTRUMENTS AND SUCCEEDED. SURGERY WAS RESUMED, AFTER A SIGNIFICANT DELAY, WITH A BACKUP DEVICE. THE PATIENT WAS NOT INJURED AS A CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15715194 ↗
  • MalfunctionOct 21, 2021Failure to Align; Failure to Align

    IT WAS REPORTED THAT, DURING INTERNAL FIXATION SURGERY, THE SURGEON STATED THAT IT WAS DIFFICULT TO LINE UP PROPERLY THE GUIDE BOLT AND A STD DRILL GUIDE DURING THE CASE. A SURGICAL DELAY OF APPROXIMATELY 2 MINUTES OCCURRED DUE TO THIS ISSUE, AND A SMITH AND NEPHEW BACKUP DEVICES WERE AVAILABLE TO CONCLUDE THE PROCEDURE. NO PATIENT INJURY WAS REPORTED DUE TO THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12677279 ↗
  • MalfunctionSep 15, 2021Break; Break

    IT WAS REPORTED THAT, DURING TIBIAL SURGERY, A 3.0MM X 1000MM BALL TIP GUIDE ROD AND A META RETRO FEM NAIL 10 X 34 PUNCHED OUT OF THE PACKAGING, OPENING THE SEAL OF THE PACKAGE. ALSO, THE BALL TIP OF THE GUIDE BOLT BROKE FROM THE ROD AND WEDGED INSIDE THE GUIDE BOLT. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH BACK-UP DEVICES. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12472136 ↗
  • MalfunctionJun 8, 2021Failure to Align; Failure to Align

    IT WAS REPORTED THAT, DURING INTERNAL FIXATION SURGERY, THE SURGEON STATED THAT IT WAS DIFFICULT TO LINE UP PROPERLY A GUIDE BOLT (CASE: (B)(6)) AND A STD DRILL GUIDE (CASE: (B)(6)) DURING THE CASE. A SURGICAL DELAY OF APPROXIMATELY 2 MINUTES OCCURRED DUE TO THIS ISSUE, AND A SMITH AND NEPHEW BACKUP DEVICES WERE AVAILABLE TO CONCLUDE THE PROCEDURE. NO PATIENT INJURY WAS REPORTED DUE TO THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11962593 ↗
  • MalfunctionDec 9, 2020Corroded; Corroded

    IT WAS REPORTED THAT RUST S BUILDING UP INSIDE OF GUIDE BOLT. NO CASE INVOLVED AS THIS WAS FOUND DURING INSPECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10978871 ↗
  • MalfunctionSep 6, 2018Device Damaged by Another Device; Device Damaged by Another Device

    IT WAS REPORTED A MALFUNCTION IN WHICH THE BOLT THAT FIXES THE NAIL TO EXTERNAL TARGETER BROKE DURING INSERTION OF THE NAIL INTO THE MEDULLARY CANAL, THEN EXTRACTED THE NAIL, REMOVED PART OF THE BROKEN BOLT, REASSEMBLED NAIL ON TARGETER WITH SECOND BOLT SUPPLIED TO URBINO AND THEN REINSERTED THE NAIL. TEN MINUTES DELAY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7850478 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K122170. One submission often covers many models, so check the product and lot details in each record.

  • Z-0351-2020 · Oct 7, 2019 · Terminated
    LAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted sha

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page