Skip to content

ELITE 72201213Recalls & reported problems

Smith & Nephew, Inc. · ELITE PREMIUM FULL LOOP KNOT MANIPULATOR, SILVER HANDLE 2 PER BOX · Class 1

All models of ELITE (9)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03596010591272), received between Oct 9, 2017 and May 3, 2019 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break2
  • Device Operates Differently Than Expected2
  • Cut In Material1

Patient problems

  • No Known Impact Or Consequence To Patient4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 3, 2019Break

    IT WAS REPORTED THAT, DURING AN ACL RECONSTRUCTION PROCEDURE, WHEN KNOTTING, THE KNOT MANIPULATOR CUT THE ULTRATAPE SUTURE. THE SUTURE WAS REMOVED FROM THE BODY. AFTER SUTURE BREAKAGE, THE IMPLANT WAS RE-SECURED WITH THE BACK-UP ULTRA TAPE DEVICE AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH EXTENDED SURGICAL TIME OF 15 MIN. THE PATIENT WAS NOT INJURED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8578495 ↗
  • MalfunctionJun 27, 2018Cut In Material

    IT WAS REPORTED THAT THE KNOT MANIPULATORS CUT THE ULTRABRAID #2 SUTURE DURING THE KNOTTING. NO SIGNIFICANT DELAY OR PATIENT INJURY WERE REPORTED. A BACK UP WAS AVAILABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7642100 ↗
  • MalfunctionDec 4, 2017Break; Device Operates Differently Than Expected

    IT WAS REPORTED THAT THE KNOT PUSHER SEEMS TO HAVE A BURR AND CUT THE SUTURE. A BACK UP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE. THERE WAS NO SIGNIFICANT DELAY OR PATIENT INJURY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7080625 ↗
  • MalfunctionOct 9, 2017Device Operates Differently Than Expected

    IT WAS REPORTED THAT THE KNOT PUSHER CUT THE SUTURE WHILE KNOTTING IT. A BACK-UP DEVICE WAS AVAILABLE AND THERE WAS NO DELAY DURING SURGERY. NO PATIENT INJURY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6931095 ↗

Other ACCESSORIES,ARTHROSCOPIC devices

All ACCESSORIES,ARTHROSCOPIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page