Skip to content

CLEAR-TRAC 72200905Recalls & reported problems

Smith & Nephew, Inc. · CLEAR-TRAC 7.0 MM X 72 MM THREADED CANNULA, WITH DISPOSABLE OBTURATOR, GREY, STERILE · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 03596010595133), received between May 28, 2019 and Mar 25, 2026 — classified as 1 injury, 8 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Delivered as Unsterile Product2
  • Material Fragmentation2
  • Contamination /Decontamination Problem1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Known Impact Or Consequence To Patient2
  • Foreign Body In Patient1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 25, 2026Delivered as Unsterile Product

    IT WAS REPORTED THAT DURING A ROTATOR CUFF PROCEDURE, THE CLEARTRAC CANNULA PACKAGE WAS RECEIVED ALREADY OPEN. THE PROCEDURE WAS RESUMED, AFTER A DELAY GREATER THAN 30 MINUTES, USING A SIMILAR S+N BACK-UP PRODUCT. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24690361 ↗
  • MalfunctionFeb 5, 2026Material Fragmentation

    IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY, PLASTIC PIECES OF THE CLEARTRAC THREADED CANNULA WERE IDENTIFIED FLOATING AROUND. CLEAR PLASTIC FROM THE TIP AND WHITE PLASTIC FROM THE CANNULA WERE IDENTIFIED. IT IS UNCLEAR HOW THE PROCEDURE WAS COMPLETED, AND IT IS UNKNOWN WHETHER THERE WAS A SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24268366 ↗
  • MalfunctionFeb 4, 2026Material Fragmentation

    IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY, PLASTIC PIECES OF THE CLEARTRAC THREADED CANNULA WERE IDENTIFIED FLOATING AROUND. CLEAR PLASTIC FROM THE TIP AND WHITE PLASTIC FROM THE CANNULA WERE IDENTIFIED. IT IS UNCLEAR HOW THE PROCEDURE WAS COMPLETED, AND IT IS UNKNOWN WHETHER THERE WAS A SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24264907 ↗
  • MalfunctionAug 11, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT DURING SET UP FOR A KNEE ARTHROSCOPY, THE CLEARTRAC CANNULA HAD BROKEN PACKAGING AND THEREFORE, WAS NOT STERILE. THERE WAS A SMITH AND NEPHEW BACK-UP DEVICE AVAILABLE AND NO DELAY WAS REPORTED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22771564 ↗
  • InjuryFeb 18, 2021Break; Break

    IT WAS REPORTED THAT DURING A SHOULDER PROCEDURE THE LITTLE PORTIONS OF THE CANNULA TRIPLE SEAL WERE BREAKING OFF WHEN INTRODUCING THE INSTRUMENTS INTO THE CANNULA. THE DEVICE BREAK INSIDE THE PATIENT, THE PIECES WERE REMOVED USING GRASPER AND SHAVER. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT A SIGNIFICANT DELAY USING A BACK-UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11342864 ↗
  • MalfunctionOct 11, 2020Break; Break

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE CANNULA BROKE. IT DID NOT CONNECT INTO THE EQUIPMENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10660173 ↗
  • MalfunctionAug 10, 2020Break; Break

    IT WAS REPORTED THAT DURING SHOULDER SURGERY, A 7.0 MM CANNULA CLEARTRAC WAS OPENED AND BEFORE USING IT, IT WAS NOTICED THAT THE RUBBER AROUND THE TIP WAS BROKEN. A S+N BACKUP WAS USED TO COMPLETED THE PROCEDURE. THERE WAS NO DELAY AND NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10384081 ↗
  • MalfunctionFeb 20, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT DURING A SHOULDER PROCEDURE, THE CANNULA HAD SMALL PIECES OF PLASTIC. THIS DEVICE WAS USED IN THE PATIENT AND THE PLASTIC PIECE FELL OFF INTO THE PATIENT BUT IT WAS RETRIEVED FROM. THE PROCEDURE WAS COMPLETED WITH A BACK UP DEVICE WITH NO DELAY OR PATIENT INJURY REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: Foreign Body In Patient

    MAUDE report 9734899 ↗
  • MalfunctionMay 28, 2019Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE CANNULA HAD BEEN BENT. A BACK UP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE , A DELAY GREATER THAN 30 MINUTES WAS REPORTED. NO PATIENT INJURIES WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8647039 ↗

Other ACCESSORIES,ARTHROSCOPIC devices

All ACCESSORIES,ARTHROSCOPIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page