TRIGEN 71674040Recalls & reported problems
Smith & Nephew, Inc. · INTERTAN LAG SCREW DRILL · Class 2
All models of TRIGEN (381)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03596010596994), received between Aug 10, 2018 and Mar 5, 2025 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
1 other recall cover devices cleared under the same FDA submission (K122170); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Crack1
- Degraded1
- Insufficient Device Problem Information1
- Naturally Worn1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 5, 2025Degraded; Degraded
IT WAS REPORTED THAT DURING TRAUMA SURGERY, IT WAS FOUND THE COLOR CODE ON THE INTERTAN LAG SCREW DRILL HAD BEEN GRADUALLY REMOVED. AFTER REMOVING THE COLOR CODE, THE DEVICE WAS STERILIZED AGAIN. THE PROCEDURE WAS ABLE TO BE COMPLETED WITH THE SAME DEVICE AFTER A SIGNIFICANT DELAY. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21531983 ↗InjuryMar 10, 2022Crack; Crack
IT WAS REPORTED THAT, DURING TRAUMA SURGERY, THE INTERTAN LAG SCREW DRILL - REAMER SPLIT NOTICED ON X-RAY WHILE IN PATIENT. DEVICE INSIDE THE PATIENT. THE PROCEDURE WAS ABLE TO BE COMPLETED WITH THE SAME DEVICE. SURGERY WAS NOT DELAYED. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13720785 ↗MalfunctionDec 30, 2019Break; Break
IT WAS REPORTED THAT DURING SURGERY THE SCREW DRILL INTERFERED WITH SOMETHING INSIDE THE PATIENT, SO THE DEVICE WAS REMOVED FROM THE SURGICAL FIELD AND IT WAS FOUND THE TIP ON THE DEVICE WAS MISSING. ANY PIECE WAS FOUND ON THE X RAY IMAGE. 15 MINUTES DELAY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9528943 ↗MalfunctionAug 10, 2018Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information
IT WAS REPORTED A MALFUNCTION IN WHICH THERE WAS A DELAY LONGER THAN 30 MINUTES. THE DEVICE WAS WORN OUT. THERE WAS NO BACK DURING THE SURGERY.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7774047 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K122170. One submission often covers many models, so check the product and lot details in each record.
- Z-0351-2020 · Oct 7, 2019 · TerminatedLAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted sha
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page