The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03596010597069), received between Oct 26, 2022 and Oct 29, 2024 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Foreign Body In Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 29, 2024Break; Break

    IT WAS REPORTED THAT DURING AN SHOULDER PROCEDURE, A 0.5MM PIECE OF THE DISTAL TIP OF THE CLEAR-TRAC CANNULA 5.5MM BROKE WHEN INTRODUCE IN THE PATIENT. THE PIECE COULD NOT BE FOUND WITHIN THE SURGICAL SITE. THE PROCEDURE WAS COMPLETED WITH A S+N BACK UP DEVICE. THERE WAS NO SURGICAL DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 20561690 ↗
  • MalfunctionAug 21, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, A WHITE PART, POSSIBLY PART OF THE CANNULA DEVICE, FELL INTO THE PATIENT'S BODY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NON-SIGNIFICANT SURGICAL DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17592598 ↗
  • MalfunctionJan 26, 2023Break; Break

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, IT WAS NOT POSSIBLE TO REMOVE THE OBTURATOR FROM THE CANNULA. THE CUSTOMER REMOVED THE DEVICE, BUT THE OBTURATOR BROKE IN THE CANNULA. THIS PROBLEM OCCURRED OUTSIDE THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SURGICAL DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16247522 ↗
  • MalfunctionOct 26, 2022Break; Break

    IT WAS REPORTED THAT, DURING AN ARTHROSCOPY, IT WAS NOT POSSIBLE TO REMOVE THE OBTURATOR FROM THE CANNULA, THE CUSTOMER REMOVED IT FORCIBLY, BUT THE OBTURATOR WAS BROKEN IN THE CANNULA. THE PROBLEM OCCURRED OUTSIDE THE PATIENT. SURGERY WAS COMPLETED WITHOUT DELAY, USING A BACK-UP DEVICE INSTEAD. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15677443 ↗