Smith & Nephew, Inc. · MTO CANNULA ROTATABLE S/V 5MM · Class 1

Adverse event reports
7
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 03596010605269), received between Oct 1, 2018 and Apr 29, 2019 — classified as 7 injuries.

The most frequently reported specific device problem is “Scratched Material” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Scratched Material4
  • Incomplete or Inadequate Connection2
  • Break1
  • Insufficient Device Problem Information1
  • Positioning Problem1
  • Sharp Edges1

Patient problems

  • Tissue Damage3
  • Injury1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1
  • Patient Problem/Medical Problem1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 29, 2019Break; Scratched Material

    IT WAS REPORTED THAT DURING KNEE ARTHROSCOPIC SURGERY, THE MEDIUM FLOW CANNULA GET A SHARP EDGE, CAUSING A POTENTIALLY CARTILAGE DAMAGE. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE AND NO SIGNIFICANT DELAY OR PATIENT INJURIES.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8562312 ↗
  • InjuryApr 29, 2019Sharp Edges

    IT WAS REPORTED THAT DURING KNEE ARTHROSCOPIC SURGERY, THE MEDIUM FLOW CANNULA GET A SHARP EDGE, CAUSING A POTENTIALLY CARTILAGE DAMAGE. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE AND NO SIGNIFICANT DELAY OR PATIENT INJURIES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8562304 ↗
  • InjuryDec 27, 2018Scratched Material; Incomplete or Inadequate Connection

    IT WAS REPORTED THAT, DURING AN SHOULDER ARTHROSCOPY, THE PATIENT SUFFERED CARTILAGE DAMAGES IN THE HUMERUS AS A CONSEQUENCE OF AN EXISTING GAP BETWEEN THE 5.00 MM CANNULA TIP-END (WHICH HAS A SHARP EDGE) AND THE ARTHROSCOPE WHEN CONNECTED. AS THIS ADVERSE CONSEQUENCE WAS NOTICED AFTER THE PROCEDURE, THE SURGERY HAD BEEN FINISHED WITH THE SAME DEVICE, IN A TIMELY MANNER. IT IS STILL UNKNOWN: (I) THE OUTCOME AND EVOLUTION OF THE PATIENT; AND (II) IF THERE WAS ANY NECESSARY REVISION SURGICAL INTERVENTION TO PRECLUDE ANY PERMANENT DAMAGE TO THE CARTILAGE.

    Patient problems: Tissue Damage

    MAUDE report 8201123 ↗
  • InjuryDec 27, 2018Scratched Material; Incomplete or Inadequate Connection

    IT WAS REPORTED THAT, DURING AN SHOULDER ARTHROSCOPY, THE PATIENT SUFFERED CARTILAGE DAMAGES IN THE HUMERUS AS A CONSEQUENCE OF AN EXISTING GAP BETWEEN THE 5.00 MM CANNULA TIP-END (WHICH HAS A SHARP EDGE) AND THE ARTHROSCOPE WHEN CONNECTED. AS THIS ADVERSE CONSEQUENCE WAS NOTICED AFTER THE PROCEDURE, THE SURGERY HAD BEEN FINISHED WITH THE SAME DEVICE, IN A TIMELY MANNER. IT IS STILL UNKNOWN: (I) THE OUTCOME AND EVOLUTION OF THE PATIENT; AND (II) IF THERE WAS ANY NECESSARY REVISION SURGICAL INTERVENTION TO PRECLUDE ANY PERMANENT DAMAGE TO THE CARTILAGE.

    Patient problems: Tissue Damage

    MAUDE report 8201092 ↗
  • InjuryDec 27, 2018Scratched Material

    IT WAS REPORTED THAT, DURING AN SHOULDER ARTHROSCOPY, THE PATIENT SUFFERED CARTILAGE DAMAGES IN THE HUMERUS AS A CONSEQUENCE OF AN EXISTING GAP BETWEEN THE 5.00 MM CANNULA TIP-END (WHICH HAS A SHARP EDGE) AND THE ARTHROSCOPE WHEN CONNECTED. AS THIS ADVERSE CONSEQUENCE WAS NOTICED AFTER THE PROCEDURE, THE SURGERY HAD BEEN FINISHED WITH THE SAME DEVICE, IN A TIMELY MANNER. IT IS STILL UNKNOWN: (I) THE OUTCOME AND EVOLUTION OF THE PATIENT; AND (II) IF THERE WAS ANY NECESSARY REVISION SURGICAL INTERVENTION TO PRECLUDE ANY PERMANENT DAMAGE TO THE CARTILAGE.

    Patient problems: Tissue Damage

    MAUDE report 8201054 ↗
  • InjuryDec 27, 2018Positioning Problem

    IT WAS REPORTED THAT, DURING AN SHOULDER ARTHROSCOPY, THE PATIENT SUFFERED CARTILAGE DAMAGES IN THE HUMERUS AS A CONSEQUENCE OF AN EXISTING GAP BETWEEN THE 5.00 MM CANNULA TIP-END (WHICH HAS A SHARP EDGE) AND THE ARTHROSCOPE WHEN CONNECTED. AS THIS ADVERSE CONSEQUENCE WAS NOTICED AFTER THE PROCEDURE, THE SURGERY HAD BEEN FINISHED WITH THE SAME DEVICE, IN A TIMELY MANNER. IT IS STILL UNKNOWN: (I) THE OUTCOME AND EVOLUTION OF THE PATIENT; AND (II) IF THERE WAS ANY NECESSARY REVISION SURGICAL INTERVENTION TO PRECLUDE ANY PERMANENT DAMAGE TO THE CARTILAGE.

    Patient problems: Patient Problem/Medical Problem

    MAUDE report 8201001 ↗
  • InjuryOct 1, 2018Insufficient Device Problem Information

    IT WAS REPORTED THAT THE SURGEON GOT A CARTILAGE DAMAGE ON HUMERUS WHEN HE WAS ACCESSING THE SHOULDER WITH THE CANNULA AND ARTHROSCOPE. A BACKUP WAS AVAILABLE TO COMPLETE THE PROCEDURE AND THERE WAS NOT A DELAY REPORTED.

    Patient problems: Injury