BIOSURE 72201888Recalls & reported problems
Smith & Nephew, Inc. · BIOSURE RATCHET DRIVER · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 03596010615442), received between Oct 23, 2018 and Jul 10, 2025 — classified as 2 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Connection Problem1
- Flare or Flash1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 10, 2025Break; Break
IT WAS REPORTED THAT DURING AN ACL SURGERY, THE TIP OF THE BIOSURE DRIVER (CASE (B)(4)) HAS BECOME FLARED OUT DUE TO CONTINUED USE. THIS CAUSED THREE BIOSURE REGENESORB SCREWS TO BREAK DURING INSERTION. THE BROKEN PIECES WERE RETRIEVED USING FORCEPS, HOMEOSTAT AND SUCTION. THE SURGERY WAS COMPLETED WITH A CHANGE IN THE SURGICAL TECHNIQUE. THERE WAS NO SURGICAL DELAY, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22473620 ↗MalfunctionAug 3, 2024Break; Break
IT WAS REPORTED THAT DURING AN ACL PROCEDURE, THE BIOSURE SCREWDRIVER TIP BROKE WHEN IT WAS BEING PLACED. ALL THE BROKEN PIECES WERE REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH NON-SIGNIFICANT SURGICAL DELAY USING A BACK-UP DEVICE THROUGH THE SAME BONE HOLE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19901872 ↗MalfunctionJan 17, 2023Break; Break
IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY, WHEN THE SURGEON WAS PUTTING AN INTERFERENCE SCREW IN THE TIBIA, THE BIOSURE RATCHET DRIVER BROKE TOWARDS THE BOTTOM WHEN THE SIZE CHANGES. THE DEVICE BROKE OUTSIDE THE PATIENT. THE SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16184789 ↗InjurySep 8, 2022Break; Break
IT WAS REPORTED THAT AFTER A KNEE PROCEDURE, IT WAS NOTED THAT THE BIOSURE RATCHET DRIVER WAS FRACTURED, POTENTIALLY LEAVING A SMALL AMOUNT OF METAL IN-SITE. X-RAYS WERE REVIEWED AND NO FOREIGN BODY WAS FOUND. THE PATIENT IS FINE AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15383843 ↗MalfunctionJan 10, 2022Break; Connection Problem; Break; Connection Problem
IT WAS REPORTED THAT, DURING AN UNKNOWN PROCEDURE. IT WAS NOTICED THAT THE BIOSURE RATCHET DRIVER HAD A BROKEN MOUTH WHICH MADE HARD TO FITTING THE SCREW. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13203939 ↗MalfunctionDec 30, 2021Break; Break
IT WAS REPORTED THAT, DURING SURGERY USING A BIOSURE RATCHET DRIVER, IT WAS NOTICED THAT THE PRODUCT KEY WAS BROKEN. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. NO DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13120340 ↗InjuryAug 19, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING LCP, LCA AND LCM ARTHROSCOPY PROCEDURE, THE BIOSURE RATCHET DRIVER BROKE IN THE DOCTOR'S HANDS. THE INSTRUMENT WAS INSIDE THE PATIENT. NO DELAY AND THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Information
MAUDE report 10428786 ↗MalfunctionOct 23, 2018Flare or Flash
IT WAS REPORTED THAT THE TIP OF THE SCREW DRIVER FLARE UP DUE TO IT WAS NECESSARY TO PUT TOO MUCH TORQUE ON THE SCREW. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO DELAY OR PATIENT INJURIES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7995805 ↗
Other ACCESSORIES,ARTHROSCOPIC devices
- Arthrex® AR-12990Arthrex, Inc. · Accessories, arthroscopic409 reports
- NA 228302Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC287 reports
- FIRSTPASS 72290128Smith & Nephew, Inc. · ACCESSORIES,ARTHROSCOPIC232 reports
- NA 251723Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC145 reports
- COBRA 3910900093Stryker Corporation · Accessories, arthroscopic138 reports
- FirstPass 72290129Smith & Nephew, Inc. · Accessories, arthroscopic82 reports
- FirstPass 72290130Smith & Nephew, Inc. · Accessories, arthroscopic80 reports
- MITEK 242025Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC61 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page