OSTEORAPTOR 72201995Recalls & reported problems
Smith & Nephew, Inc. · OSTEORAPTOR 2.9 SUTURE ANCHOR WITH TWO ULTRABRAID No.2 SUTURE WHITE/COBRAID BLUE · Class 2 · Implant
All models of OSTEORAPTOR (8)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Smith & Nephew Osteoraptor suture anchor is used to attach tissue to bone during surgery. Its 255 reports mostly describe the anchor breaking or fracturing during surgery, often forcing a switch to a backup device, with some fragments reported in patients. There were no deaths and 91 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 248
- Deaths reported
- 0
- Injuries reported
- 91
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 248 adverse event reports that list this exact device (DI 03596010615565), received between Aug 14, 2017 and Jun 21, 2026 — classified as 91 injuries, 157 malfunctions.
The most frequently reported specific device problem is “Break” (153 reports).
40 other recalls cover devices cleared under the same FDA submission (K151105); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break153
- Device Dislodged or Dislocated43
- Fracture26
- Crack13
- Material Separation10
- Material Deformation6
- Detachment of Device or Device Component3
- Insufficient Device Problem Information3
Patient problems
- No Clinical Signs, Symptoms or Conditions192
- Perforation22
- No Known Impact Or Consequence To Patient17
- Rupture9
- Foreign Body In Patient8
- No Information4
- Unspecified Tissue Injury4
- Insufficient Information3
Recent adverse event reports
The 25 most recent of 248 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 21, 2026Break
IT WAS REPORTED THAT, DURING A SURGERY, THE OSTEORAPTOR ANCHOR BROKE. THE PROCEDURE WAS RESUMED, USING AN EQUIVALENT SN BACK UP DEVICE. IT IS UNKNOWN IF THERE WAS A DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25572653 ↗MalfunctionJun 7, 2026Break
IT WAS REPORTED THAT, DURING A SHOULDER GLENOID LABRUM REPAIR PROCEDURE, THE OSTEORAPTOR'S SUTURE BROKE AS THE ANCHOR WAS HAMMERED IN. THE ANCHOR WAS REMOVED USING TWEEZERS. THE PROCEDURE WAS RESUMED, WITH A NON SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACK-UP. AN ADDITIONAL HOLE WAS REQUIRED FOR THE BACK UP DEVICE, THERE WAS A VOID LEFT IN THE PATIENT. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Perforation
MAUDE report 25422739 ↗MalfunctionMay 21, 2026Fracture
IT WAS REPORTED THAT, DURING AN ANKLE LIGAMENT REPAIR SURGERY, AFTER THE INSERTION SITE WAS CLEARED OF ALL DEBRIS WITH A 2.9MM ANCHOR MATCHING INSTRUMENT, THE SIDE OF ONE (1) OSTEORAPTOR SUTURE ANCHOR WAS FRACTURED INSIDE THE PATIENT. THE ANCHOR AND ALL THE PIECES WERE REMOVED FROM THE PATIENT WITH STRAIGHT FORCEPS. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACK-UP. THE BACK-UP DEVICE WAS USED IN THE ORIGINALLY DRILLED BONE HOLE. ADDITIONAL ANESTHESIA WAS NOT REQUIRED AS CONSEQUENCE OF THE SURGICAL PROLONGATION. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25237968 ↗MalfunctionApr 9, 2026Break; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING A SHOULDER JOINT GLENOID LABRUM REPAIR SURGERY, THE OSTEORAPTOR ANCHOR FELL OFF AS THE IMPLANT ENTERED THE BONE TUNNEL, THE ANCHOR APPEARED TO BE BROKEN. THE PROCEDURE WAS COMPLETED WITH A S+N EQUIVALENT DEVICE. IT IS UNKNOWN IF THERE WAS SURGICAL DELAY, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24832533 ↗MalfunctionMar 30, 2026Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC DISLOCATION SURGERY OF THE SHOULDER JOINT, THE OSTEORAPTOR ANCHOR BROKE. ALL TH EPIECES WERE REMOVED WITH FORCEPS AND SUCTION. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY USING AN EQUIVALENT SN BACK-UP. THE BACK UP WAS USED IN THE ORIGINAL DRILLED HOLE, NO VOIDS WERE LEFT, THE INSTRUMENT TO PREPARE THE INSERTION SITE WAS A 2.9 MM ANCHOR MATCHING INSTRUMENT. THE PATIENT STATUS WAS FINE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24714583 ↗MalfunctionMar 26, 2026Break
IT WAS REPORTED THAT, DURING A SURGERY, AN OSTEORAPTOR ANCHOR BROKE DURING IMPLANTATION. THE PROCEDURE WAS RESUMED, USING AN EQUIVALENT S+N BACK-UP. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24707130 ↗MalfunctionMar 4, 2026Fracture
IT WAS REPORTED THAT, DURING A HIP ARTHROSCOPIC PROCEDURE, THE OSTEORAPTOR'S THREE PIECE SET FUNCTIONED NORMALLY AND ACHIEVED NORMAL OPENING, HOWEVER, THE SCREW COULD NOT ADVANCE INTO POSITION AND FRACTURED DURING IMPACTION. THE PROCEDURE WAS RESUMED USING AN EQUIVALENT S+N BACK-UP. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24511950 ↗MalfunctionMar 2, 2026Break
IT WAS REPORTED THAT, DURING SURGERY, ONE (1) OSTEORAPTOR SUTURE ANCHOR BROKE. THE PROCEDURE WAS RESUMED, USING AN EQUIVALENT S+N BACKUP; HOWEVER, IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24479919 ↗MalfunctionFeb 8, 2026Fracture
IT WAS REPORTED THAT DURING A REPAIR OF LATERAL HUMERAL LIGAMENT OF ELBOW JOINT PROCEDURE, IT WAS NOTICED THAT OSTEORAPTOR 2.9 ANCHOR GOT FRACTURED DURING IMPLANTATION. THE PROCEDURE WAS RESUMED, USING AN S+N EQUIVALENT BACK UP PRODUCT, HOWEVER IT IS UNKNOWN IF THERE WAS ANY SURGICAL DELAY, NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24288443 ↗MalfunctionFeb 7, 2026Break
IT WAS REPORTED THAT, DURING A BRISTOW BANKART PROCEDURE, WHEN THE OSTEORAPTOR 2.9 W/ 2 UB WHITE / BLUE WAS INSERTED INTO THE BONE HOLE, IT FRACTURED. THE INSERTION SITE WAS CLEARED OF ALL DEBRIS AND ALL THE BROKEN PIECES WERE REMOVED FROM THE PATIENT. THE PROCEDURE WAS RESUMED, AFTER A 5-MINUTE DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24285987 ↗MalfunctionFeb 5, 2026Material Deformation
IT WAS REPORTED THAT, DURING A PROCEDURE, WHEN INSERTING THE OSTEORAPTOR ANCHOR, IT DEVIATED FROM THE ORIGINAL TRAJECTORY, CAUSING THE BENDING TO BE UNUSABLE. THE PROCEDURE WAS RESUMED, USING AN EQUIVALENT SN BACK-UP. IT IS UNKNOWN IF THERE WAS A DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24268263 ↗MalfunctionFeb 4, 2026Break
IT WAS REPORTED THAT DURING AN ANKLE LIGAMENT REPAIR SURGERY, THE OSTEORAPTOR ANCHOR BROKE. THE BROKEN PIECES WERE REMOVED WITH TWEEZERS. THE PROCEDURE WAS RESUMED, AFTER A NON SIGNIFICANT DELAY, USING AN EQUIVALENT SN BACK-UP ON THE ORIGINALLY DRILLED HOLE, PREPARED WITH ANCHOR MATCHING INSTRUMENT. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24259292 ↗MalfunctionFeb 3, 2026Break
IT WAS REPORTED THAT DURING AN ANKLE LIGAMENT REPAIR SURGERY, THE OSTEORAPTOR ANCHOR BROKE. ALL THE PIECES WERE REMOVED WITH TWEEZERS. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACK-UP IN THE ORIGINALLY DRILLED BONE HOLE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24247456 ↗MalfunctionJan 23, 2026Break
IT WAS REPORTED THAT, DURING A ROTATOR CUFF AND LABRUM REPAIR SURGERY, THREE (3) UNITS OF A OSTEORAPTOR 2.9 W/ 2 UB WHITE / BLUE WERE FRACTURED. THE PROCEDURE WAS RESUMED USING AN EQUIVALENT S+N BACK UP. IT IS UNKNOWN IF THERE WAS A DELAY, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24167755 ↗MalfunctionJan 21, 2026Break
IT WAS REPORTED THAT, DURING A ROTATOR CUFF AND LABRUM REPAIR SURGERY, AN OSTEORAPTOR 2.9 W/ 2 UB WHITE / BLUE BROKE. THE PROCEDURE WAS RESUMED USING AN EQUIVALENT S+N BACK UP. IT IS UNKNOWN IF THERE WAS A DELAY, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24130479 ↗MalfunctionJan 15, 2026Break; Fracture; Material Deformation
IT WAS REPORTED THAT, DURING A GLENOID LABRUM REPAIR, AN OSTEORAPTOR ANCHOR BROKE. THE PROCEDURE WAS RESUMED, USING AN EQUIVALENT S+N BACK UP. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24080799 ↗MalfunctionJan 13, 2026Break
IT WAS REPORTED THAT DURING A SHOULDER DISLOCATION PROCEDURE, IT WAS NOTICED THAT THE OSTEORAPTOR 2.9 ANCHOR'S TIP GOT BROKEN WHILE BEING IMPLANTED. THE BROKEN PARTS WERE REMOVED USING TWEEZERS. THE PROCEDURE WAS AFTER RESUMED AFTER A NON-SIGNIFICANT SURGICAL DELAY, USING AN EQUIVALENT SMITH AND NEPHEW BACK UP DEVICE ON THE ORIGINAL DRILLED BONE HOLE, LEAVING NO VOID IN PATIENT. PATIENT IS FINE AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24054571 ↗MalfunctionJan 12, 2026Fracture
IT WAS REPORTED THAT, DURING A LABRUM REPAIR SURGERY, THE OSTEORAPTOR ANCHOR WAS FRACTURED. THE PROCEDURE WAS COMPLETED WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES USING AN EQUIVALENT S&N BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24048485 ↗MalfunctionJan 12, 2026Break; Fracture
IT WAS REPORTED THAT, DURING A GLENOID LABRUM REPAIR, AN OSTEORAPTOR ANCHOR BROKE. THE PROCEDURE WAS RESUMED, USING AN EQUIVALENT S+N BACK UP. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24039182 ↗MalfunctionDec 26, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT, DURING AN ARTHROSCOPY FOR SHOULDER INSTABILITY, WHILE THE DOCTOR WAS MANEUVERING, THE OSTEORAPTOR 2.9 W/ 2 UB WHITE / BLUE TIP CAME LOOSE INSIDE. IT WAS REMOVED FROM THE PATIENT WITH A GRASPER FORCEPS. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23897852 ↗MalfunctionDec 22, 2025Fracture; Fracture
IT WAS REPORTED THAT, DURING A SHOULDER JOINT LABRUM REPAIR SURGERY, THE TOP OF ONE (1) OSTEORAPTOR ANCHOR FRACTURED DURING IMPLANTATION. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23871850 ↗MalfunctionDec 18, 2025Break; Break
IT WAS REPORTED THAT, DURING SHOULDER SURGERY, THE OSTEORAPTOR ANCHOR WAS BROKEN. THE BROKEN PIECES WERE REMOVED FROM THE PATIENT USING TWEEZERS. THE PROCEDURE WAS COMPLETED WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES USING AN EQUIVALENT SN BACK-UP DEVICE IN THE ORIGINAL BONE HOLE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23841083 ↗MalfunctionDec 12, 2025Fracture; Premature Activation; Fracture; Premature Activation
IT WAS REPORTED THAT DURING THE ANKLE LIGAMENT REPAIR SURGERY, THE OSTEORAPTOR ANCHOR FELL OFF FROM THE INSERTER AND WAS PHYSICALLY BROKEN. THE PROCEDURE WAS RESUMED, AFTER A NON SIGNIFICANT DELAY, USING AN EQUIVALENT SN BACK-UP ON THE ORIGINALLY DRILLED BONE HOLE, NO BROKEN PIECES WERE LEFT IN PATIENT , THEY WERE REMOVED WITH TWEEZERS. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23791589 ↗MalfunctionDec 2, 2025Crack; Crack
IT WAS REPORTED THAT DURING SURGERY, AFTER DRILLING THE BONE HOLE WITH THE 2.9 MM TOOL, ONE (1) OSTEORAPTOR SUTURE ANCHOR CRACKED WHILE BEING HAMMERED IN. THE PROCEDURE WAS PERFORMED USING AN EQUIVALENT S+N BACK-UP. IT IS UNKNOWN IF THERE WAS ANY SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23700905 ↗MalfunctionDec 1, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING A FOOT AND ANKLE SURGERY, A GENTLE PULL CAUSED THE OSTEORAPTOR ANCHOR TO COME OUT AFTER IT WAS IMPLANTED. THE PROCEDURE WAS RESUMED USING AN EQUIVALENT S+N BACKUP INTO AN ADDITIONAL BONE HOLE, AFTER A NON-SIGNIFICANT DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Perforation
MAUDE report 23687178 ↗
FDA clearance and classification
Cleared through 510(k) K151105 on Jul 28, 2015 (decision: Substantially Equivalent). Submitted by Smith & Nephew, Inc. as “Smith and Nephew Suture Anchors - 5.0mm Absorbable Polymer Anchor, Smith and Nephew Suture Anchors - TAG Bioraptor Suture Anchor 7209317, 7209318, Smith and Nephew Suture Anchors - Bioraptor 2.3 PK Suture Anchor, Smith and Nephew Suture Anchors - Osteoraptor Suture Anchor, Smith and Nephew Suture Anchors - Twinfix Ultra HA Suture Anchor, Smith and Nephew Suture Anchors - OSTEORAPTOR Curved Suture Anchors, Smith and Nephew Suture Anchors - Ultra Fast-Fix Meniscal Repair System, Ultra Fa”. The FDA review took 95 days. FDA record ↗
The FDA regulates product code HWC (screw, fixation, bone) under 21 CFR 888.3040 as a Class II device. 21 CFR 888.3040 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K151105. One submission often covers many models, so check the product and lot details in each record.
- Z-0147-2021 · Sep 9, 2020 · Open, ClassifiedRAPTORMITE 3.7MM AB W/ NEEDLES AND TWO-Absorbable Suture anchor Product Number: 72201805
- Z-0138-2021 · Sep 9, 2020 · Open, ClassifiedTWINFIX 5.0MM AB SUTURE ANCHOR-Absorbable Suture anchors Product Number: 7210711
- Z-0165-2021 · Sep 9, 2020 · Open, ClassifiedOSTEORAPTOR CRVD 2.3 SA UB COBRD BLUE- Absorbable Suture anchor Product Number: 72203290
- Z-0166-2021 · Sep 9, 2020 · Open, ClassifiedOSTEORAPTOR CRVD 2.3 SA UB COBRD BLACK- Absorbable Suture anchor Product Number: 72203291
- Z-0169-2021 · Sep 9, 2020 · Open, ClassifiedHEALICOIL RSB SA 4.75MM W/1 UT & 1 UB BK-- Absorbable Suture anchor Product Number: 72203705
- Z-0171-2021 · Sep 9, 2020 · Open, ClassifiedHEALICOIL RG SA 5.5MM W/3 UB-BL CB BL BK- Absorbable Suture anchor Product Number: 72203707
- Z-0159-2021 · Sep 9, 2020 · Open, ClassifiedTWINFIX ULTRA PLLA HA 4.5MM W/ NEEDLES- Absorbable Suture anchor Product Number: 72202612
- Z-0162-2021 · Sep 9, 2020 · Open, ClassifiedTWINFIX ULTRA HA 4.5 W/2 UB BLUE & BLK- Absorbable Suture anchor Product Number: 72202624
- Z-0167-2021 · Sep 9, 2020 · Open, ClassifiedHEALICOIL RSB SA 4.75MM W/1 UT & 1 UB BL- Absorbable Suture anchor Product Number: 72203697
- Z-0160-2021 · Sep 9, 2020 · Open, ClassifiedTWINFIX ULTRA PLLA HA 5.5MM W/ NEEDLES- Absorbable Suture anchor Product Number: 72202616
- Z-0144-2021 · Sep 9, 2020 · Open, ClassifiedULTRA FAST-FIX AB ASSEMBLY - CURVED-Absorbable Suture anchors Product Number: 72201494
- Z-0173-2021 · Sep 9, 2020 · Open, ClassifiedHEALICOIL RSB SA 5.5MM W/1 UT & 1 UB BL- Absorbable Suture anchor Product Number: 72203801
- Z-0175-2021 · Sep 9, 2020 · Open, ClassifiedMICRORAPTOR REGENESORB SUTURE ANCHOR- Absorbable Suture anchor Product Number: 72204984
- Z-0161-2021 · Sep 9, 2020 · Open, ClassifiedTWINFIX ULTRA PLLA HA 6.5MM W/ NEEDLES- Absorbable Suture anchor Product Number: 72202620
- Z-0146-2021 · Sep 9, 2020 · Open, ClassifiedBIORAPTOR 2.9MM SUTURE ANCHOR-Absorbable Suture anchor Product Number: 72201702
…and 25 more.
Common questions about the OSTEORAPTOR 72201995
Has the OSTEORAPTOR 72201995 been recalled?
No FDA recall record in our data names this exact device identifier (DI 03596010615565), based on the FDA recall file dated Nov 27, 2026. However, 40 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the OSTEORAPTOR 72201995?
The FDA's MAUDE database lists 248 adverse event reports for this device, received between Aug 14, 2017 and Jun 21, 2026. In those reports, the most common specific device problems are “Break” (153), “Device Dislodged or Dislocated” (43) and “Fracture” (26); the most common patient problems are “Perforation” (22), “Rupture” (9) and “Foreign Body In Patient” (8). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the OSTEORAPTOR 72201995?
None of the reports are classified as deaths, 91 as injuries and 157 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the OSTEORAPTOR 72201995 FDA approved?
It was cleared through 510(k) K151105 on Jul 28, 2015. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the OSTEORAPTOR 72201995 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.
Who makes the OSTEORAPTOR 72201995, and is it still sold?
The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand OSTEORAPTOR, model 72201995). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone498 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-200Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-325Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-550Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-425Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-600Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page