Skip to content

NA 03.043.029Recalls & reported problems

Synthes GmbH · AIMING ARM/ RADIOLUCENT · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Synthes radiolucent aiming arm is used in bone surgery, and its 501 reports mostly describe parts breaking, such as clips and hooks, or parts not fitting with other devices. Some describe a drill hitting a nail and a connecting screw that would not come off, causing a delay. There are no deaths or injuries reported, and no recalls name this device. Reports fell to 65 in the past year from 138 in the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
498
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 498 adverse event reports that list this exact device (DI 10886982297396), received between Aug 18, 2021 and Jun 9, 2026 — classified as 498 malfunctions.

The most frequently reported specific device problem is “Break” (284 reports).

2 other recalls cover devices cleared under the same FDA submission (K201336); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 10 reportsJan 2024: 13 reportsFeb 2024: 6 reportsMar 2024: 9 reportsApr 2024: 7 reportsMay 2024: 5 reportsJun 2024: 7 reportsJul 2024: 5 reportsAug 2024: 6 reportsSep 2024: 7 reportsOct 2024: 9 reportsNov 2024: 13 reportsDec 2024: 6 reportsJan 2025: 16 reportsFeb 2025: 13 reportsMar 2025: 7 reportsApr 2025: 14 reportsMay 2025: 15 reportsJun 2025: 8 reportsJul 2025: 16 reportsAug 2025: 12 reportsSep 2025: 8 reportsOct 2025: 10 reportsNov 2025: 7 reportsDec 2025: 16 reportsJan 2026: 11 reportsFeb 2026: 7 reportsMar 2026: 7 reportsApr 2026: 5 reportsMay 2026: 6 reportsJun 2026: 3 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202616
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break284
  • Device-Device Incompatibility212
  • Crack13
  • Material Integrity Problem8
  • Device Slipped4
  • Material Deformation4
  • Material Twisted/Bent4
  • No Apparent Adverse Event4

Patient problems

  • No Clinical Signs, Symptoms or Conditions430
  • Insufficient Information71
  • Bone Notching1
  • Foreign Body In Patient1

Recent adverse event reports

The 25 most recent of 498 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 9, 2026Device-Device Incompatibility

    IT WAS REPORTED ON AN UNKNOWN DATE, DURING THE PROCEDURE, THE DRILL BIT HIT THE TIBIAL NAIL TWICE WHEN DRILLING TWO DIFFERENT HOLES. AT THE END OF THE PROCEDURE THE CONNECTING SCREW WOULD NOT COME OFF. THERE WAS 20 MINUTES OF DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25445117 ↗
  • MalfunctionJun 4, 2026Break

    IT WAS REPORTED THAT THEY OPENED SET FOR A TIBIA NAIL CASE (TNA) AND FOUND THAT BOTH HOOKS WERE BROKEN OFF. THIS HADN'T BEEN AN ISSUE IN THE PREVIOUS CASE, BUT WAS NOW BROKEN WHEN OPENED FOR THIS CASE, DAYS LATER. ANOTHER TNA SET WITH THE ARM WAS RETRIEVED. WHEN IT WAS OPENED ONE OF THE HOOKS WAS BROKEN OFF OF IT. THE SETS ARE NOT LOCKED WITH ANY INSTRUMENTATION TO CREATE EXCESSIVE PRESSURE. ALL BY HAND AND NOT EXCESSIVE BY HAND. THESE ARE FROM SEPARATE FACILITIES AND DIFFERENT SURGEONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25399131 ↗
  • MalfunctionJun 4, 2026Break

    IT WAS REPORTED THAT THEY OPENED SET FOR A TIBIA NAIL CASE (TNA) AND FOUND THAT BOTH HOOKS WERE BROKEN OFF. THIS HADN'T BEEN AN ISSUE IN THE PREVIOUS CASE, BUT WAS NOW BROKEN WHEN OPENED FOR THIS CASE, DAYS LATER. ANOTHER TNA SET WITH THE ARM WAS RETRIEVED. WHEN IT WAS OPENED ONE OF THE HOOKS WAS BROKEN OFF OF IT. THE SETS ARE NOT LOCKED WITH ANY INSTRUMENTATION TO CREATE EXCESSIVE PRESSURE. ALL BY HAND AND NOT EXCESSIVE BY HAND. THESE ARE FROM SEPARATE FACILITIES AND DIFFERENT SURGEONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25398984 ↗
  • MalfunctionMay 19, 2026Break

    IT WAS REPORTED THE CLIPS ON THE AIMING ARMS BROKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25218941 ↗
  • MalfunctionMay 19, 2026Break

    IT WAS REPORTED THE CLIPS ON THE AIMING ARMS BROKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25218908 ↗
  • MalfunctionMay 7, 2026Break

    IT WAS REPORTED THAT THE AIM-ARM RADIOLUC EXPERIENCED A SNAPPED-OFF KNOB, WITH THE DAMAGE IDENTIFIED DURING A LOAN KIT INSPECTION. THE EVENT WAS DISCOVERED BY A TECHNICIAN DURING INSPECTION, AND THERE WAS NO INVOLVEMENT OF A SURGICAL PROCEDURE OR PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25113773 ↗
  • MalfunctionMay 7, 2026Device-Device Incompatibility

    WHEN DRILLING THE PROXIMAL MEDIAL TO LATERAL HOLES ON THE NAIL, THE CARBON FIBER AIMING ARM WAS MISALIGNED, LEADING TO A FAIL IN DRILLING USING THE INSTRUMENTATION. THIS CAUSED DELAY IN SURGERY TIME. NO COMPLICATIONS FOR PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25112117 ↗
  • MalfunctionMay 1, 2026Break

    IT WAS REPORTED ON(B)(6) 2026, DURING A PROCEDURE TWO DEVICES BROKE. WHEN THEY WERE CLIPPING THE AIMING ARM ON, A PIECE OF THE CARBON FIBER BROKE OFF. ON INSERTION OF THE TIBIAL NAIL, THE METAL PIN ON THE IMPACTOR BROKE AND THE SPRING FELL OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25058419 ↗
  • MalfunctionMay 1, 2026Break

    IT WAS REPORTED THAT ON (B)(6) 2026, THE SURGEON WAS HAMMERING IN A TIBIA NAIL WITH THE AIMING ARM ATTACHED TO THE INSERTION HANDLE WHICH MUST'VE WEAKENED ONE OF THE HOOKS THAT HOOKS ONTO ONE OF THE PEGS OF THE AIMING ARM BECAUSE WHEN HE WENT TO TAKE THE AIMING ARM OFF ONE OF THOSE HOOKS SNAPPED OFF. THE SURGEON NOTICED THAT THE HOOK BROKE OFF AND THINKS IT HAS MESSED WITH THE CONNECTION BETWEEN THE AIMING ARM AND THE INSERTION HANDLE. THERE WAS NO SURGICAL DELAY. SURGEON HAD TO MANIPULATE THE AIMING ARM IN ORDER FOR THE DRILL BITS GOING THROUGH THE TROCARS TO GO THROUGH THE NAIL AND NOT HIT THE NAIL. PROCEDURE WAS SUCCESSFULLY COMPLETED. THE AIMING ARM HOOK BROKE OFF AND WENT ONTO THE FLOOR OF THE OPERATING ROOM. THERE WAS NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25056611 ↗
  • MalfunctionApr 29, 2026Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, HANDLE OF AIMING ARM BROKE DURING HANDLING. THERE WAS NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25027724 ↗
  • MalfunctionApr 23, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT SURGEON EXPERIENCED MISTARGETING WHILST PERFORMING DRILLING OF PROXIMAL LOCKING HOLES THROUGH SUPRAPATELLAR AIMING ARM. KIT INCLUDED THE NEWLY DESIGNED SUPRAPATELLAR INSERTION HANDLE. SURGEON INDICATED THAT HE NEEDED TO MOVE THE AIMING ARM TO ALLOW THE DRILL BIT TO PASS THROUGH THE NAIL. THIS OCCURRED IN BOTH PROXIMAL STATIC HOLES. SURGEON EXPLAINED THAT HE FELT THE ISSUE WAS IN THE CONNECTION BETWEEN THE AIMING ARM AND INSERTION HANDLE. SURGEON WAS ABLE TO COMPLETE THE PROCEDURE AND THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24973272 ↗
  • MalfunctionApr 21, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT AIMING ARM/ RADIOLUCENT AND INSERT-HANDLE RADIOLUC SHORT WERE INVOLVED IN A REPORTED EVENT. THE SURGEON EXPERIENCED MISTARGETING DURING DRILLING OF PROXIMAL LOCKING HOLES THROUGH THE TNA AIMING ARM, NOTING THAT THE DRILL BIT WAS NOTCHING THE PROXIMAL HOLE AND REQUIRED REPOSITIONING OF THE AIMING ARM TO PASS THE DRILL BIT THROUGH THE NAIL. THE SURGEON SUSPECTED AN ALIGNMENT ISSUE BETWEEN THE AIMING ARM AND THE INSERTION HANDLE. THE SURGERY WAS COMPLETED WITHOUT ANY REPORTED PATIENT HARM OR CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24939204 ↗
  • MalfunctionApr 9, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2026 DURING A PROCEDURE, SURGEON USED A TIBIAL NAIL AND THE PROXIMAL INTERLOCK "MISSED" THE HOLE IN THE NAIL WHEN INSERTING. SURGEON REMOVED THE SCREW, REPOSITIONED AIMING ARM AND INSERTION HANDLE, REDRILLED, THEN REINSERTED THE SCREW THROUGH THE NAIL. THE SURGERY WAS DELAYED BY 10 MINUTES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED AND THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24832915 ↗
  • MalfunctionApr 1, 2026Break

    IT WAS REPORTED THAT THE TNA AIMING ARM BROKE DURING REMOVAL FROM THE INSERTIONS HANDLE AFTER SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24755753 ↗
  • MalfunctionMar 30, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT INSERTION HANDLE/ RADIOLUCENT MEDIUM AND AIM-ARM RADIOLUC WERE INVOLVED IN AN EVENT DURING A PARAPATELLAR TIBIA NAIL INSERTION PROCEDURE. WHILE PLACING PROXIMAL INTERLOCKING SCREWS USING THE AIMING ARM, THE DRILL BIT STRUCK THE NAIL, RESULTING IN A SLIGHT PROCEDURAL DELAY OF 3-5 MINUTES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY, AND THERE WAS NO REPORTED PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24726998 ↗
  • MalfunctionMar 25, 2026Break

    IT WAS REPORTED ON (B)(6) 2026 THE SURGEON WAS INSERTING A TIBIA NAIL. THE NAIL WAS INSERTED AND THE JIG WAS THEN HOOKED ONTO THE NAIL INSERTOR. WHEN THE AIMING ARM WAS CLICKED ONTO THE JIG THE LATCH BROKE. NO PATIENT HARM WAS CREATED AND WE OPENED ANOTHER SET TO COMPLETE THE PROCEDURE. NO PATIENT HARM

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24691899 ↗
  • MalfunctionMar 24, 2026Break

    ADDITIONAL INFORMATION RECEIVED: A. DATE OF EVENT (IF AVAILABLE): WHAT IS THE TIMING OF THE REPORTED EVENT? (PRE-OP, INTRA-OP, POST-OP)? - NO ONE SPECIFIC EVENT. JUST DEGRADATION OF THE GRAPHIC CASE AND WORRIES WITH THE HOSPITAL THAT STERILIZATION MIGHT NOT BE GREAT WITH FLAKING OF TRAY. NO CERTAIN OCCURRENCE. NO PATIENT ISSUES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24681562 ↗
  • MalfunctionMar 20, 2026Crack; Device-Device Incompatibility

    IT WAS REPORTED THAT AIM-ARM RADIOLUC AND UNK - NAIL INSERTION HANDLES WERE INVOLVED IN AN INTRA-OPERATIVE EVENT. DURING A TIBIAL NAIL PROCEDURE, THE PROXIMAL LOCKING SCREWS IN THE MEDIAL ASPECT OF THE AIMING ARM DID NOT ALIGN WITH THE HOLES IN THE TIBIAL NAIL. TROUBLESHOOTING STEPS, INCLUDING CHECKING THE AIMING ARM CONNECTION AND USING A MALLET TO ENSURE A TIGHT FIT, DID NOT RESOLVE THE MISALIGNMENT. ULTIMATELY, THE SURGICAL TEAM RESORTED TO FREEHAND DRILLING AND INSERTION OF THE LOCKING SCREWS. AFTER THE PROCEDURE, THE AIMING ARM AND INSERTION HANDLE WERE FOUND TO BE TIGHTLY CONNECTED AND REQUIRED PRYING APART, AND A SPLIT WAS OBSERVED IN ONE OF THE CLIPS ON THE AIMING ARM. THERE WAS NO REPORTED PATIENT HARM, BUT A SIGNIFICANT PROCEDURAL DELAY OF 30 MINUTES OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24647823 ↗
  • MalfunctionMar 16, 2026Break

    IT WAS REPORTED THE PEGS THAT ATTACH TO THE INSERTION ARMS BROKE. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24607167 ↗
  • MalfunctionMar 11, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT DURING A TIBIAL NAIL ADVANCE PROCEDURE, AFTER USING THE COMPRESSION FEATURE WITHIN THE NAIL HANDLE, AND AFTER THE DYNAMIC PROXIMAL SCREW WAS INSERTED, SURGEON ATTEMPTED TO HOLD THE COMPRESSION WITH THE 2ND PROXIMAL STATIC SCREW, BUT THE DRILL BIT HIT THE NAIL AND WAS NOT GOING THROUGH IN ANY MEANS. SURGEON TRIED MINIMAL MANIPULATION MOVING HIS HAND UP AND DOWN, AND EVENTUALLY THE DRILL WENT THROUGH BUT OUTSIDE THE NAIL. THE SURGEON HAD TO REMOVE THE GUIDE ARM AND DRILLED THE HOLE FREE-HAND UNDER X-RAY GUIDANCE. THE EVENT RESULTED IN A 20-MINUTE SURGICAL DELAY AND REQUIRED ADDITIONAL X-RAY IMAGING. THE PROCEDURE WAS SUCCESSFULLY COMPLETED, AND THERE WERE NO REPORTED PATIENT CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24570645 ↗
  • MalfunctionMar 10, 2026Break

    IT WAS REPORTED ON AN UNKNOWN DATE THAT WHEN INSERTING THE TIBIAL NAIL, MALLET WAS USED TO INSERT PROPERLY AND THE CLIPS THAT CONNECTING AIMING ARM TO INSERTER CHIPPED OFF DURING SURGERY. NO FRAGMENTS WERE GENERATED. THERE WAS NO SURGICAL DELAY DUE TO THE REPORTED EVENT. IT DID NOT EFFECT CASE AND THERE WERE NO PATIENT STATUS/OUTCOME/CONSEQUENCES. PROCEDURE WAS NOT COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24556814 ↗
  • MalfunctionFeb 25, 2026Break; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) , 2026, DURING INSERTION OF A TIBIAL IMN, THE SURGEON WAS MALL EATING THE DRIVING CAP ATTACHED TO THE AIMING ARM AND THE WINGS OF THE AIMING ARM BROKE OFF, PREVENTING THE AIMING ARM FROM BEING SECURED TO THE INSERTION HANDLE. THERE WAS 10 MINUTES SURGICAL DELAY. SURGEON HAD TO HOLD THE AIMING ARM TO THE INSERTION HANDLE WHEN INSERTING THE SCREWS. PROCEDURE WAS SUCCESSFULLY COMPLETED. PATIENT OUTCOME WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24444440 ↗
  • MalfunctionFeb 17, 2026Device-Device Incompatibility

    IT WAS REPORTED ON AN UNKNOWN DATE, THAT TNA WITH SUPRAPATELLA AIMING ARM, THE JIG AND A MISSING PROXIMAL STATIC HOLE 1 BY A FEW MM. (THE MOST DISTAL PROXIMAL HOLE FROM MEDIAL TO LATERAL). IT SEEMS TO BE CONSISTENTLY TARGETING A FEW MM POSTERIOR AND SUPERIOR OF THE HOLE IN THE NAIL - CAN SEE THE START OF THE DRILL HOLE ON X-RAY AFTER SHOOTING PERFECT CIRCLES AFTER REMOVING AIMING ARM. ISSUES WITH THE AIMING ARM MISSING OCCURRED WITH THE ORIGINAL CARBON FIBER SUPRAPATELLA INSERTION HANDLE AS WELL AS THE FULLY METAL/NEW STYLE INSERTION HANDLE. IN ALL SCENARIOS THE NEW CONNECTION SCREW WITH WASHER WAS USED. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24364302 ↗
  • MalfunctionFeb 17, 2026Device-Device Incompatibility

    IT WAS REPORTED ON AN UNKNOWN DATE, THAT TNA WITH SUPRAPATELLA AIMING ARM, THE JIG AND A MISSING PROXIMAL STATIC HOLE 1 BY A FEW MM. (THE MOST DISTAL PROXIMAL HOLE FROM MEDIAL TO LATERAL). IT SEEMS TO BE CONSISTENTLY TARGETING A FEW MM POSTERIOR AND SUPERIOR OF THE HOLE IN THE NAIL - CAN SEE THE START OF THE DRILL HOLE ON X-RAY AFTER SHOOTING PERFECT CIRCLES AFTER REMOVING AIMING ARM. ISSUES WITH THE AIMING ARM MISSING OCCURRED WITH THE ORIGINAL CARBON FIBER SUPRAPATELLA INSERTION HANDLE AS WELL AS THE FULLY METAL/NEW STYLE INSERTION HANDLE. IN ALL SCENARIOS THE NEW CONNECTION SCREW WITH WASHER WAS USED. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24364242 ↗
  • MalfunctionFeb 17, 2026Material Integrity Problem

    IT WAS REPORTED THAT DURING A ROUTINE INSPECTION, THE REPORTER IDENTIFIED DAMAGE TO TWO AIMING ARM INSTRUMENTS. BOTH ITEMS WERE FOUND TO HAVE DAMAGE ON THE CONNECTING HINGES THAT ATTACH TO THE TNA INSERTION HANDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24363324 ↗

FDA clearance and classification

Cleared through 510(k) K201336 on Sep 3, 2020 (decision: Substantially Equivalent). Submitted by Synthes (USA) Products, LLC as “DePuy Synthes Tibial Nail Advanced, Depuy Synthes Locking Screws Medullary Nails, 4.0 and 5.0mm”. The FDA review took 106 days. FDA record ↗

The FDA regulates product code HWC (screw, fixation, bone) under 21 CFR 888.3040 as a Class II device. 21 CFR 888.3040 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K201336. One submission often covers many models, so check the product and lot details in each record.

  • Z-0229-2026 · Sep 26, 2025 · Open, Classified
    LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Adva
  • Z-0228-2026 · Sep 26, 2025 · Open, Classified
    LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Adva

Common questions about the NA 03.043.029

Has the NA 03.043.029 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10886982297396), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the NA 03.043.029?

The FDA's MAUDE database lists 498 adverse event reports for this device, received between Aug 18, 2021 and Jun 9, 2026. In those reports, the most common specific device problems are “Break” (284), “Device-Device Incompatibility” (212) and “Crack” (13); the most common patient problems are “Bone Notching” (1) and “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA 03.043.029?

None of the reports are classified as deaths, 0 as injuries and 498 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA 03.043.029 FDA approved?

It was cleared through 510(k) K201336 on Sep 3, 2020. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the NA 03.043.029, and is it still sold?

The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 03.043.029). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page