Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 03596010653666), received between Mar 15, 2019 and May 22, 2026 — classified as 4 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Crack2
  • Material Separation1
  • Mechanical Jam1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Patient Involvement2
  • No Information1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 22, 2026Break

    IT WAS REPORTED THAT, DURING A ROTATOR CUFF PROCEDURE, THE HANDLE OF THE SUTURE CUTTER BROKE. ALL THE PIECES WERE REMOVED. THE PROCEDURE WAS RESUMED, WITH A SIGNIFICANT DELAY, USING AN ARTHREX COMPETITOR DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25257381 ↗
  • InjurySep 7, 2022Crack; Crack

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE SUTURE CUTTER METAL CRACKED AT DISTAL END OF TOOL, AS IT WAS UNABLE TO CUT THE SUTURE. THERE WAS A PHYSICAL BREAKAGE PRESENT. THERE WAS NO LOSS OF MATERIAL, BUT THE CUTTER PUSHED THE SUTURE THROUGH THE CRACK IN MATERIAL. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15375570 ↗
  • InjuryOct 27, 2021Crack; Crack

    IT WAS REPORTED THAT, DURING A SURGERY, THE FLUSH SUTURE CUTTER LOWER JAW WAS CRACKED SO THE SUTURE COULD NOT BE CUT. NO DELAY WAS REPORTED. IT IS UNKNOWN IF THERE WAS SIGNIFICANT DELAY. IT IS UNKNOWN IF THERE WAS SIGNIFICANT DELAY. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12707895 ↗
  • InjuryFeb 22, 2021Break; Mechanical Problem; Break; Mechanical Problem

    IT WAS REPORTED THAT DURING A SHOULDER LABRAL REPAIR SURGERY, THE FLUSH CUTTER WOULD NOT OPEN AND CLOSE PROPERLY AND A SMALL PART AT THE FRONT END BECAME SEPARATED FROM THE MAIN SHAFT. THE PROCEDURE WAS COMPLETED WITHOUT SIGNIFICANT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11362046 ↗
  • MalfunctionFeb 5, 2021Break; Break

    IT WAS REPORTED THAT THE TIP OF THE FLUSH SUTURE CUTTER WAS SPLIT AND THE SUTURE IS SLIPPING THROUGH. NO CASE WAS INVOLVED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11283830 ↗
  • MalfunctionOct 18, 2020Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE FLUSH SUTURE CUTTER HAD JAW ISSUES, IT DID NOT OPEN. NO CASE WAS INVOLVED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Patient Involvement

    MAUDE report 10694710 ↗
  • InjuryAug 14, 2020Break; Break

    IT WAS REPORTED THAT DURING SURGERY, THE FLUSH SUTURE CUTTER BROKE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A BACK-UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Information

    MAUDE report 10410800 ↗
  • MalfunctionJul 30, 2020Material Separation; Material Separation

    IT WAS REPORTED THAT DURING THE SET UP INSPECTION, THE CUTTING PIECE BROKE OFF OF THE SUTURE CUTTER. BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE. NO DELAY AND NO PATIENT INJURIES WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Patient Involvement

    MAUDE report 10343863 ↗
  • MalfunctionMar 15, 2019Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE DOCTOR PROCEEDED TO CUT A SUTURE AND THE BASE OF THE CUTTER BROKE INSIDE THE PATIENT, ALL THE PIECES WERE REMOVED. THERE WERE NO DELAYS OR PROBLEMS AS THERE WAS ANOTHER CUTTER AVAILABLE.