Skip to content

R3 71366432Recalls & reported problems

Smith & Nephew, Inc. · R3 LINER IMPACTOR HEAD II 32MM · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Smith & Nephew R3 Liner Impactor Head II is a tool used during hip replacement surgery to seat an acetabular liner. The reports mostly describe the head cracking, breaking, or chipping while impacting the liner, sometimes leaving fragments that surgeons recovered; no deaths are reported and 7 injuries are reported. There are 40 adverse event reports in total and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
40
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 03596010656063), received between May 30, 2018 and Oct 15, 2025 — classified as 7 injuries, 33 malfunctions.

The most frequently reported specific device problem is “Crack” (17 reports).

2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack17
  • Break12
  • Fracture10
  • Degraded1
  • Material Fragmentation1
  • Naturally Worn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions33
  • No Consequences Or Impact To Patient4
  • No Patient Involvement3
  • No Known Impact Or Consequence To Patient2
  • No Code Available1

Recent adverse event reports

The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 15, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING A THA SURGERY, IT WAS NOTICED THAT ONE (1) R3 LNR MPACTOR HD II 32MM WAS CHIPPED FROM BEING HIT OFF THE RIM OF THE CUP. NO PIECES FELL INSIDE THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23301991 ↗
  • MalfunctionNov 13, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING A THR SURGERY, WHEN IMPACTING THE ACETABULAR LINER, THE TIP OF THE R3 LNR IMPACTOR HD II 32MM FRACTURE. NOTHING FELL INSIDE THE WOUND. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20678543 ↗
  • MalfunctionMay 8, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING A THR, THE R3 LNR MPACTOR HD II 32MM BROKE DURING IMPLANTATION OF THE LINER. ALL THE PIECES WERE REPORTEDLY RECOVERED USING FORCEPS. SURGERY WAS COMPLETED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19270763 ↗
  • MalfunctionFeb 27, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING THR SURGERY, THE HEAD OF THE R3 LNR MPACTOR HD II 32MM CRACKED DURING SURGERY WHILE IMPACTING LINER. PIECES WERE RETRIEVED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18792705 ↗
  • MalfunctionNov 9, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING A THR, THE R3 LNR MPACTOR HD II 32MM BROKE INSIDE THE PATIENT. SURGERY WAS PERFORMED AFTER A NON-SIGNIFICANT DELAY. IT IS UNKNOWN HOW SURGERY WAS FINISHED. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18100545 ↗
  • InjuryDec 9, 2022Fracture; Fracture

    IT WAS REPORTED THAT, DURING A THR, ONE (1) R3 LNR MPACTOR HD II 32MM BROKE WHILE BEING IMPACTED. ALL PIECES WERE RECOVERED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15948811 ↗
  • MalfunctionOct 27, 2022Fracture; Fracture

    IT WAS REPORTED THAT, DURING THR, ONE R3 LNR MPACTOR HD II 32MM BROKE WHILE THE SURGEON WAS INSERTING AN R3 CUP 54MM. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15683465 ↗
  • MalfunctionOct 14, 2022Break; Break

    IT WAS REPORTED THAT, TWO R3 LNR MPACTOR HD II 32MM ARE BROKEN. SINCE IT WAS NOTICED DURING A FIELD INSPECTION, NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15610806 ↗
  • MalfunctionOct 12, 2022Break; Break

    IT WAS REPORTED THAT, ONE (1) R3 LNR MPACTOR HD II 32MM WAS FOUND BROKEN. AS THIS WAS NOTICED UPON FIELD INSPECTION, NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15592130 ↗
  • MalfunctionOct 12, 2022Fracture; Fracture

    IT WAS REPORTED THAT TWO (2) R3 LNR MPACTOR HD II 32MM ARE BROKEN. AS THIS WAS NOTICED DURING FIELD INSPECTION, NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15591449 ↗
  • MalfunctionSep 8, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING THR, WHILE IMPACTING A LINER INTO THE CUP, THE BLUE PLASTIC TIP ON THE END OF THE R3 LNR MPACTOR HD II 32MM BROKE OFF INTO 5 TO 6 PIECES. IT IS UNKNOW IF ANY PIECE FELL INTO THE PATIENT, HOW THE SURGERY WAS FINISHED. NO INJURY OR DELAY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15386135 ↗
  • MalfunctionApr 21, 2022Crack; Crack

    IT WAS REPORTED THAT, DURING THR, THE BLUE PORTION OF THE R3 LNR MPACTOR HD II 32MM CRACKED. NO PIECES BROKE OFF. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. NO DELAYS REPORTED. NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14177168 ↗
  • MalfunctionApr 11, 2022Crack; Crack

    IT WAS REPORTED THAT THE PLASTIC ON THE HEAD OF AN R3 LINER IMPACTOR HEAD II 32MM IS CRACKED. AS THIS WAS NOTICED IN A NON-SURGICAL ENVIRONMENT, THERE WAS NOT ANY PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14075145 ↗
  • MalfunctionFeb 16, 2022Crack; Crack

    IT WAS REPORTED THAT, DURING INSPECTION, IT WAS FOUND THAT AN R3 LNR MPACTOR HD II 32MM HAVE A CRACK IN THEM. NO CASE INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13540185 ↗
  • MalfunctionFeb 16, 2022Crack; Crack

    IT WAS REPORTED THAT, DURING INSPECTION, IT WAS FOUND THAT AN R3 LNR MPACTOR HD II 32MM HAS A CRACK IN IT. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13540180 ↗
  • MalfunctionFeb 10, 2022Crack; Degraded; Crack; Degraded

    IT WAS REPORTED THAT, DURING SET UP FOR A THR SURGERY, IT WAS NOTICED THAT TWO R3 LNR IMPACTOR HD II 32MM HAS CRACKED IN THEM DUE TO WEAR. THE PROCEDURE WAS FINISHED WITH A SMITH AND NEPHEW BACK UP DEVICE WITHOUT DELAY. SINCE THE INCIDENT OCCURRED BEFORE THE PROCEDURE, THE PATIENT WAS NOT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13506166 ↗
  • MalfunctionFeb 2, 2022Crack; Crack

    IT WAS REPORTED THAT, DURING A THR SURGERY, A R3 LNR MPACTOR HD II 32MM HAS DEVELOPED A CRACK IN SIDING. THE PROCEDURE WAS COMPLETED, WITHOUT DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13428955 ↗
  • InjuryJan 4, 2022Crack; Crack

    IT WAS REPORTED THAT, DURING A TOTAL HIP REPLACEMENT, THE TIP OF A R3 LINER IMPACTOR HEAD 32MM CRACKED. SURGERY WAS RESUMED, WITHOUT ANY DELAY, USING A SMITH & NEPHEW BACK-UP DEVICE. PATIENT WAS NOT HARMED BEYOND THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13159515 ↗
  • InjuryDec 10, 2021Crack; Crack

    IT WAS REPORTED THAT DURING IMPACTION OF THE LINER, THE R3 LNR MPACTOR HD II 32MM CRACKED. IT IS UNKNOWN HOW THE PROCEDURE WAS FINISHED OR IF THERE WAS A SURGICAL DELAY. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12974620 ↗
  • MalfunctionNov 17, 2021Break; Break

    IT WAS REPORTED THAT, DURING A THR SURGERY, AN R3 LINER IMPACTOR HEAD II 32MM BROKE OFF. NO BROKEN PIECES FELL INTO THE PATIENT, INSTRUMENT WAS OUTSIDE THE PATIENT. SURGERY WAS COMPLETED WITH A BACKUP DEVICE, WITHOUT ANY DELAY. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12831979 ↗
  • MalfunctionSep 24, 2021Crack; Naturally Worn; Crack; Naturally Worn

    IT WAS REPORTED THAT, A R3 LNR MPACTOR HD II 32MM AND R3 LNR MPACTOR HD II 36MM WERE WORN AND CRACKED. AS THIS WAS NOTICED UPON INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12530577 ↗
  • MalfunctionJun 22, 2021Break; Break

    IT WAS REPORTED THAT A R3 LNR MPACTOR HD II 32MM BROKE. INCIDENT OCCURRED DURING SET UP OR INSPECTION; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12038835 ↗
  • MalfunctionMay 13, 2021Crack; Crack

    IT WAS REPORTED THAT DURING SET UP FOR A THR, IT WAS NOTICED THAT THE LNR IMPACTOR HD II 32MM WAS CRACKED. THERE WAS NO SIGNIFICANT DELAY AND THE PROCEDURE WAS FINISHED USING A SMITH & NEPHEW BACK UP. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11822307 ↗
  • MalfunctionMay 13, 2021Crack; Crack

    IT WAS REPORTED THAT AFTER A THR, IT WAS NOTICE THAT THE R3 LNR MPACTOR HD II 32MM (CASE-(B)(4)) AND THE R3 LNR MPACTOR HD II 36MM (CASE (B)(4)) WERE CRACK. THE PROCEDURE WAS FINISHED WITH THE SAME DEVICES AND NO SURGICAL DELAY WAS REPORTED. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11816878 ↗
  • MalfunctionApr 21, 2021Crack; Crack

    IT WAS REPORTED THAT THE R3 LNR MPACTOR HD II 32MM HEAD IMPACTOR CRACKED. THE EVENT OCCUR DURING USE IN A THR SURGERY OUTSIDE THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A SMITH AND NEPHEW BACK UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11704319 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.

  • Z-0730-2019 · Jan 7, 2019 · Terminated
    SL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
  • Z-1008-2017 · Nov 16, 2016 · Terminated
    REFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116

Other Prosthesis, hip, femoral component, cemented, metal devices

All Prosthesis, hip, femoral component, cemented, metal devices →