R3 71366451Recalls & reported problems
Smith & Nephew, Inc. · R3 LINER IMPACTOR HEAD II 50-54MM · Class 2
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 03596010656100), received between Mar 28, 2021 and Mar 20, 2022 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Crack” (6 reports).
2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Crack6
- Scratched Material4
- Naturally Worn3
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 20, 2022Crack; Crack
IT WAS REPORTED THAT A R3 LINER IMPACTOR HEAD II 50-54MM WAS FOUND TO BE CRACKED. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13831437 ↗MalfunctionJul 6, 2021Crack; Naturally Worn; Scratched Material; Crack; Naturally Worn; Scratched Material
IT WAS REPORTED THAT A R3 LINER IMPACTOR HEAD II 50-54MM IS SCRATCHED, CRACKED AND WORN DOWN. AS THIS WAS NOTICED UPON INSPECTION, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12123102 ↗MalfunctionJul 6, 2021Crack; Naturally Worn; Scratched Material; Crack; Naturally Worn; Scratched Material
IT WAS REPORTED THAT A R3 LINER IMPACTOR HEAD II WAS SCRATCHED, CRACKED AND WORN. AS THIS WAS NOTICED UPON INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12121979 ↗MalfunctionJul 5, 2021Crack; Naturally Worn; Scratched Material; Crack; Naturally Worn; Scratched Material
IT WAS REPORTED THAT A R3 LINER IMPACTOR HEAD II WAS SCRATCHED, CRACKED AND WORN. AS THIS WAS NOTICED UPON INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12119350 ↗MalfunctionJul 5, 2021Crack; Scratched Material; Crack; Scratched Material
IT WAS REPORTED THAT, DURING INSPECTION A R3 LNR IMPACTOR HD II WAS NOTICED SCRATCHED AND CRACKED. AS THIS WAS NOTICED IN A NON-SURGICAL ENVIRONMENT, THERE WAS NOT ANY PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12119333 ↗InjuryMay 10, 2021Break; Break
IT WAS REPORTED THAT, DURING A THR, A R3 LNR MPACTOR HD II 50-54MM BROKE WHILE BEING USED AS A SECONDARY CUP IMPACTOR. INSTRUMENT WAS BEING USED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED USING A S+N BACK-UP DEVICE. SURGERY WAS LESS THAN OR EQUAL TO 30 MIN DELAYED. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11797916 ↗MalfunctionMar 28, 2021Crack; Crack
IT WAS REPORTED THAT DURING A THR A R3 LNR MPACTOR HD II 50-54MM CRACKED INTO TWO PIECES. THE SURGERY WAS COMPLETED WITH A BACK UP, NO DELAYS AND NO INJURIES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11580407 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.
- Z-0730-2019 · Jan 7, 2019 · TerminatedSL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
- Z-1008-2017 · Nov 16, 2016 · TerminatedREFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116
Other Prosthesis, hip, femoral component, cemented, metal devices
- GENESIS II 71440005Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal599 reports
- NA 71364073Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal436 reports
- Peri-Loc 71173377Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal401 reports
- R3 71368569Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal398 reports
- NA 71364006Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal197 reports
- NA 71364011Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal176 reports
- Accord 71360046Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal171 reports
- REFLECTION 71362279Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal155 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.