CAPI 3 IMMUNOTYPING 2600Recalls & reported problems
Sebia · Device designed for the qualitative detection and the characterization of monoclonal proteins by capillary electrophoresis · Class 1
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03607360026007), received between Sep 16, 2021 and Sep 16, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Component Misassembled” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Component Misassembled1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2021Component Misassembled; Component Misassembled
18 AUGUST 2021: THE CUSTOMER OBSERVED THAT ONE KIT OF THE CAPI 3 IMMUNOTYPING ANTISERA CARTRIDGE OUT OF THE 50 KITS RECEIVED, LOT 11021/01, PRESENTED WITH DISCORDANT RESULTS. VISUAL INSPECTION IDENTIFIED THAT TWO OF THE ANTISERA ANTI-IGA AND ANTI-IGM WERE REVERSED IN THE POSITIONS WITHIN THE SEALED CARTRIDGE. NO DISCREPANT PATIENT LABORATORY REPORTS WERE RELEASED. THERE WERE NO REPORTS OF IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12482986 ↗
Other Multi-Analyte Controls, All Kinds (Assayed) devices
- LIQUID ASSAYED CHEMISTRY CONTROL PREMIUM PLUS - LEVEL 3 (LIQ CHEM ASY PREMIUM PL LAE4215Randox Laboratories Limited · Multi-Analyte Controls, All Kinds (Assayed)1 recall0 reports
- Dried Blood Control 0193Chromsystems Instruments + Chemicals GmbH · Multi-Analyte Controls, All Kinds (Assayed)1 recall0 reports
- MAS Omni-CARDIO Liquid Assayed Integrated Cardiac Control OCRD-SPMicrogenics Corporation · Multi-Analyte Controls, All Kinds (Assayed)1 recall0 reports
- LIQUID ASSAYED CHEMISTRY CONTROL PREMIUM – LEVEL 3 (LIQ CHEM ASY PREMIUM 3) LAE4217Randox Laboratories Limited · Multi-Analyte Controls, All Kinds (Assayed)1 recall0 reports
- MAS Omni-CARDIO Liquid Assayed Integrated Cardiac Control OCRD-303Microgenics Corporation · Multi-Analyte Controls, All Kinds (Assayed)1 recall0 reports
- IMMUNOASSAY PREMIUM - TRI-LEVEL (IA PREMIUM 1, 2 AND 3) IA2633Randox Laboratories Limited · Multi-Analyte Controls, All Kinds (Assayed)1 recall0 reports
- IMMUNOASSAY PREMIUM - LEVEL 3 (IA PREMIUM 3) IA2640Randox Laboratories Limited · Multi-Analyte Controls, All Kinds (Assayed)1 recall0 reports
- MAS Omni-CARDIO Liquid Assayed Integrated Cardiac Control OCRD-202Microgenics Corporation · Multi-Analyte Controls, All Kinds (Assayed)1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.