STA Satellite® 58104Recalls & reported problems
Diagnostica Stago · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03607450581041), received between May 5, 2020 and May 5, 2020 — classified as 1 injury.
The most frequently reported specific device problem is “Incorrect Measurement” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K082248); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Jul 21, 2025Open, Classified
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
Z-2473-2025 · Root cause: Mixed-up of materials/components · FDA record ↗
Most reported problems
Device problems
- Incorrect Measurement1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMay 5, 2020Incorrect Measurement; Incorrect Measurement
ON (B)(6) 2020, CUSTOMER RELEASED AN INR GREATER THAN 16.0 FOLLOWING TWO REPEATED VMAX ERROR ON THE PT TEST. ACCORDING TO THE CUSTOMER, QUALITY CONTROL BEFORE AND AFTER THE RESULTS WERE IN RANGE. CUSTOMER PERFORMED THE MAINTENANCE OF THE INSTRUMENT AND REPEATED AGAIN THE TEST: INR= 2.4. THE CORRECTED RESULT WAS CALLED TO PROVIDER: NO CHANGE MADE TO PATIENT THERAPY DUE TO THE INITIAL RESULT. THEN ON (B)(6) 2020, QC FOR PT AND PTT WERE IN VMAX ERROR. QC REPEATED WITHIN RANGE BUT AFTER A THIRD RUN, PTT QC WERE IN VMAX AGAIN. INSTRUMENT TAKEN OUT OF SERVICE AND SAMPLES FROZEN. A TECHNICAL INTERVENTION HAS BEEN ASKED AND PERFORMED THE SAME DAY: THE TECHNICIAN FOUND A NEEDLE CLOGGED AND REPLACED NEEDLE FROM CUSTOMER INVENTORY. PRELIMINARY CONCLUSION: THE MOST LIKELY ROOT CAUSE WAS A CLOGGED NEEDLE.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10024815 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K082248. One submission often covers many models, so check the product and lot details in each record.
- Z-0252-2013 · Jun 15, 2012 · TerminatedSTA Satellite Automated Multi-Parameter Analyzer This system is a fully automatic clinical instrument indicated and intended for the performance of tests on human plasma, the resu
Other SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES devices
- ACL TOP 970 CL 00000280097Instrumentation Laboratory Company · System, Multipurpose For In Vitro Coagulation Studies2 recalls0 reports
- TEG5000 HEMOSTASIS ANALYZER 07-022Haemonetics Corporation · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall2 reports
- ACL ELITE 00000880311RInstrumentation Laboratory Company · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall0 reports
- ACL ELITE PRO 00001000311Instrumentation Laboratory Company · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall0 reports
- ACL ELITE PRO 00001000311RInstrumentation Laboratory Company · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall0 reports
- ACL ELITE 00000880311Instrumentation Laboratory Company · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall0 reports
- Hemochron Whole Blood Microcoagulation System ELITEAccriva Diagnostics Holdings, Inc. · System, Multipurpose For In Vitro Coagulation Studies2 deaths13 reports
- Automated Blood Coagulation Analyzer CS-2500 BV981798Sysmex Corporation · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 death8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page