TEG5000 HEMOSTASIS ANALYZER 07-022Recalls & reported problems
Haemonetics Corporation · TEG ANALYZER, 5000, 120V, 60HZ, FG · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 30812747018184), received between Jan 28, 2020 and May 18, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Temperature Problem” (1 report).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Apr 19, 2022Class IITerminated
When the TEG 5000 Analyzer including TEG Analytical Software is used with PlateletMapping (ADP or AA) Assays and TEG Mangement Software, incorrect values of the PlateletMapping ADP or AA % Inhibition and % Aggretagion may be displayed on TEG Manager for the affected assays.
Z-1298-2022 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Temperature Problem1
- Thermal Decomposition of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Patient Involvement1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 18, 2021Temperature Problem; Temperature Problem
HAEMONETICS WAS NOTIFIED THAT A CUSTOMER REPORTED A TEG DEVICE IS EMITTING SMOKE FROM WHERE CHANNELS ARE CONNECTED, CUSTOMER NEEDS FSE FOR REPAIR AND THIS WAS HAPPENED AFTER THE PM HAS DONE ON THAT DEVICE. THERE WAS NO DONOR INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11843483 ↗MalfunctionJan 28, 2020Thermal Decomposition of Device; Thermal Decomposition of Device
ON (B)(6) 2019 HAEMONETICS WAS INFORMED THAT DURING A QUALITY CHECK (QC) A TEG ANALYZR, 5000, 60HZ, FG STARTED SMOKING AND TURNED ITSELF OFF. THE ANALYZER WAS THEN UNPLUGGED. A PATIENT WAS NOT AT THE TIME THE EVENT OCCURRED.
Patient problems: No Patient Involvement
MAUDE report 9636263 ↗
Other SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES devices
- ACL TOP 970 CL 00000280097Instrumentation Laboratory Company · System, Multipurpose For In Vitro Coagulation Studies2 recalls0 reports
- STA Satellite® 58104Diagnostica Stago · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall1 report
- ACL ELITE 00000880311Instrumentation Laboratory Company · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall0 reports
- ACL ELITE PRO 00001000311RInstrumentation Laboratory Company · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall0 reports
- ACL ELITE PRO 00001000311Instrumentation Laboratory Company · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall0 reports
- ACL ELITE 00000880311RInstrumentation Laboratory Company · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall0 reports
- Hemochron Whole Blood Microcoagulation System ELITEAccriva Diagnostics Holdings, Inc. · System, Multipurpose For In Vitro Coagulation Studies2 deaths13 reports
- Automated Blood Coagulation Analyzer CS-2500 BV981798Sysmex Corporation · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 death8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page