Medicrea|UNiD ROD B33125599-02Recalls & reported problems
Medicrea International · UNID EXP B33125599-02 5.5 TI ROD 4UP L · Class 2 · Implant
All models of Medicrea|UNiD ROD (6)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 03613720289494), received between Feb 24, 2025 and Jul 17, 2026 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture3
- Break2
- Inadequacy of Device Shape and/or Size1
- No Apparent Adverse Event1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 17, 2026Inadequacy of Device Shape and/or Size
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25853557 ↗MalfunctionFeb 5, 2026Fracture
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT, POSTOPERATIVE X-RAYS WERE MEASURED AND NOTICED THE ROD FRACTURES DURING MEASUREMENT PROTOCOLS. PROCEDURE PERFORMED WAS POSTERIOR SPINAL FUSION AT LEVELS L2-S1. THERE WERE NO PATIENT SYMPTOMS REPORTED. NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24266363 ↗InjuryDec 8, 2025Break; No Apparent Adverse Event; Break; No Apparent Adverse Event
SUBSEQUENT TO SUBMISSION OF THE INITIAL MDR, ADDITIONAL INFORMATION WAS RECEIVED INDICATING THE DEVICE INVOLVED WAS NOT MANUFACTURED BY MEDTRONIC. THE INITIAL REPORT WAS SUBMITTED BASED ON PRELIMINARY INFORMATION. NO MEDTRONIC DEVICE FAILURE IS ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23739664 ↗MalfunctionSep 2, 2025Fracture; Fracture
ADDITIONAL INFORMATION WAS RECEIVED REGARDING THE EVENT. THE PATIENT WAS HAVING POSTERIOR SPINAL FUSION PROCEDURE. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22936692 ↗MalfunctionJun 11, 2025Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. PRE-OP DIAGNOSIS OF PATIENT WAS DEGENERATIVE SPINE, STENOSIS, SCOLIOSIS. IT WAS REPORTED THAT, THE UNID ROD WAS BROKEN. THE UNID ROD BROKE ABOVE THE RIGHT S2 SCREW. THE SMALL FRAGMENT THAT WAS IN S2 WAS REMOVED. THE REMAINING ROD FROM T10-S1 REMAINS IN THE PATIENT AND A LATERAL CONNECTOR AND DOMINOS WERE USED TO AFFIX A SUPPLEMENTAL ROD AND THE S2 SCREW TO THE REMAINDER OF THE CONSTRUCT. PROCEDURE PERFORMED WAS T10-PELVIS, L2/3 DIRECT LATERAL LUMBAR FUSION, L3/4 AND L4/L5 TRANSFORAMINAL LUMBAR INTERBODY FUSION. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22182218 ↗InjuryFeb 24, 2025Fracture; Fracture
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING T10-PELVIS POST ERIOR SACROILIAC FUSION (PSIF). IT WAS REPORTED THAT THE ROD WAS FRACTURED POST-OPERATIVELY AND A REVISION SURGERY WAS PERFORMED ON (B)(6) 2025 TO EXPLANT THE BROKEN ROD. THE BROKEN ROD WAS REPLACED IN REVISION SURGERY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21443650 ↗
Other Thoracolumbosacral pedicle screw system devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page