Medicrea|UNiD ROD B33126099-02Recalls & reported problems
Medicrea International · UNID EXP B33126099-02 6.0 TI ROD 4UP L · Class 2 · Implant
All models of Medicrea|UNiD ROD (6)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03613720289531), received between Nov 14, 2023 and Mar 6, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Device Handling Problem2
- Adverse Event Without Identified Device or Use Problem1
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Bone Fracture(s)1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 6, 2026Fracture
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING POSTERIOR SPINAL FUSION THERAPY AT T9 - PELVIS LEVEL. IT WAS REPORTED THAT A ROD FRACTURE OCCURRED AT THE L2-3 LEVEL, AND THE FRACTURED ROD REMAINS IMPLANTED. POSTOPERATIVE X-RAYS WERE CONDUCTED, AND THE ROD FRACTURES WERE IDENTIFIED DURING MEASUREMENT PROTOCOLS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24528425 ↗MalfunctionSep 8, 2025Break; Device Handling Problem; Break; Device Handling Problem
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL FUSION. IT WAS REPORTED THAT, THE UNID ROD WAS BROKEN AND THE ROD REMAINS IN PATIENT. LEVELS IMPLANTED WERE T10-ILIAC. THERE WAS NO REVISION SURGERY THAT IS CURRENTLY PLANNED TO EXPLANT THE BROKEN RODS. PATIENT SYMPTOMS WERE ASKED BUT UNKNOWN. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22990211 ↗MalfunctionSep 8, 2025Break; Device Handling Problem; Break; Device Handling Problem
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL FUSION. IT WAS REPORTED THAT, THE UNID ROD WAS BROKEN AND THE ROD REMAINS IN PATIENT. LEVELS IMPLANTED WERE T10-ILIAC. THERE WAS NO REVISION SURGERY THAT IS CURRENTLY PLANNED TO EXPLANT THE BROKEN RODS. PATIENT SYMPTOMS WERE ASKED BUT UNKNOWN. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22990184 ↗MalfunctionApr 30, 2024Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT THE SURGEON WAS USING A SAGITTAL BENDER AT THE TOP OF THE ROD AND THE ROD BROKE WHILE BENDING. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19210647 ↗InjuryNov 14, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT UNDERWENT POSTERIOR SPINAL FUSION PROCEDURE FOR UNKNOWN SPINAL INDICATION. IT WAS REPORTED THAT, INITIALLY THE RODS WERE FRACTURED POST-OPERATIVELY. PROCEDURE OR TECHNIQUE PERFORMED DURING IMPLANTATION WAS POSTERIOR LUMBAR INTERBODY FUSION AT L2-L3, ANTERIOR LUMBAR INTERBODY FUSION AT L5-S1. THE LEVELS IMPLANTED WERE T11-PELVIS. REVISION SURGERY WAS PLANNED FOR (B)(6), 2023. THE INDICATED RODS WERE PARTIALLY REMOVED DURING REVISION. THE RODS WERE CUT ROUGHLY 2 INCHES BELOW THE TOPS TO ACCOMMODATE DOMINOS TO EXTEND THE FUSION CONSTRUCT UP TO T4 FROM T10. IT WAS CONFIRMED THAT THE ORIGINAL RODS WERE NOT FRACTURED BUT THE PATIENT NEEDED TO BE EXTENDED UP. THE CUT OFF ROD PIECES WERE DISPOSED OF. IN POST-OP X-RAYS, THEY DISCOVERED A T10 COMPRESSION FRACTURE WITH BILATERAL PEDICLE FRACTURES WHICH REQUIRED TO EXTEND THE CONSTRUCT TO T4. THERE WERE NO PATIENT SYMPTOMS AS RESULT, NO FURTHER COMPLICATIONS REPORTED.
Patient problems: Bone Fracture(s)
MAUDE report 18128969 ↗
Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices
- EXPEDIUM 179702000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL439 reports
- EXPEDIUM VERSE 199721001Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL227 reports
- XIA 3 48230000Stryker Corporation · APPLIANCE, FIXATION, SPINAL INTERLAMINAL193 reports
- VIPER 186715000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 death167 reports
- NA 04.614.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL80 reports
- NA 04.615.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- EXPEDIUM VIPER 196789480Depuy Spine, LLC · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- Arsenal Deformity Spinal Fixation System 47127Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL66 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page