Skip to content

Virage OCT Spinal Fixation System 07.01728.001Recalls & reported problems

Highridge Medical, LLC · Standard Closure Top · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03662663112758), received between May 7, 2026 and May 7, 2026 — classified as 4 injuries.

The most frequently reported specific device problem is “Malposition of Device” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 4 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Malposition of Device4
  • Physical Resistance/Sticking2

Patient problems

  • Difficulty Chewing4
  • Dysphasia4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 7, 2026Malposition of Device

    IT WAS REPORTED THAT FOUR DAYS AFTER THE INITIAL SURGERY, A PATIENT UNDERWENT REVISION SURGERY TO ADDRESS TRISMUS AND DYSPHASIA. DURING REMOVAL, A COUPLE OF CLOSURE TOPS BECAME SEIZED AND THE ROD CONNECTOR WAS ALSO CUT OUT USING A HELICOIDAL RASP. A SCREW WAS ALSO FOUND TO HAVE FRACTURED AT THE WELD, AND WAS REMOVED AND REPLACED. THE PATIENT SYMPTOMS WERE ATTRIBUTED TO MALPOSITION OF THE DEVICES. THIS IS REPORT FIVE OF ELEVEN FOR THIS EVENT.

    Patient problems: Difficulty Chewing; Dysphasia

    MAUDE report 25117720 ↗
  • InjuryMay 7, 2026Malposition of Device

    IT WAS REPORTED THAT FOUR DAYS AFTER THE INITIAL SURGERY, A PATIENT UNDERWENT REVISION SURGERY TO ADDRESS TRISMUS AND DYSPHASIA. DURING REMOVAL, A COUPLE OF CLOSURE TOPS BECAME SEIZED AND THE ROD CONNECTOR WAS ALSO CUT OUT USING A HELICOIDAL RASP. A SCREW WAS ALSO FOUND TO HAVE FRACTURED AT THE WELD, AND WAS REMOVED AND REPLACED. THE PATIENT SYMPTOMS WERE ATTRIBUTED TO MALPOSITION OF THE DEVICES. THIS IS REPORT FOUR OF ELEVEN FOR THIS EVENT.

    Patient problems: Difficulty Chewing; Dysphasia

    MAUDE report 25117667 ↗
  • InjuryMay 7, 2026Malposition of Device; Physical Resistance/Sticking

    IT WAS REPORTED THAT FOUR DAYS AFTER THE INITIAL SURGERY, A PATIENT UNDERWENT REVISION SURGERY TO ADDRESS TRISMUS AND DYSPHASIA. DURING REMOVAL, A COUPLE OF CLOSURE TOPS BECAME SEIZED AND THE ROD CONNECTOR WAS ALSO CUT OUT USING A HELICOIDAL RASP. A SCREW WAS ALSO FOUND TO HAVE FRACTURED AT THE WELD, AND WAS REMOVED AND REPLACED. THE PATIENT SYMPTOMS WERE ATTRIBUTED TO MALPOSITION OF THE DEVICES. THIS IS REPORT THREE OF ELEVEN FOR THIS EVENT.

    Patient problems: Difficulty Chewing; Dysphasia

    MAUDE report 25117627 ↗
  • InjuryMay 7, 2026Malposition of Device; Physical Resistance/Sticking

    IT WAS REPORTED THAT FOUR DAYS AFTER THE INITIAL SURGERY, A PATIENT UNDERWENT REVISION SURGERY TO ADDRESS TRISMUS AND DYSPHASIA. DURING REMOVAL, A COUPLE OF CLOSURE TOPS BECAME SEIZED AND THE ROD CONNECTOR WAS ALSO CUT OUT USING A HELICOIDAL RASP. A SCREW WAS ALSO FOUND TO HAVE FRACTURED AT THE WELD, AND WAS REMOVED AND REPLACED. THE PATIENT SYMPTOMS WERE ATTRIBUTED TO MALPOSITION OF THE DEVICES. THIS IS REPORT TWO OF ELEVEN FOR THIS EVENT.

    Patient problems: Difficulty Chewing; Dysphasia

    MAUDE report 25117596 ↗

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices

All APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page