Skip to content

CRF II - RADIAL PROSTHESIS DTJ019Recalls & reported problems

Tornier · Radial Head · Class 2 · Implant

All models of CRF II - RADIAL PROSTHESIS (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 03700386901232), received between Apr 4, 2023 and Sep 12, 2024 — classified as 6 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated4
  • Adverse Event Without Identified Device or Use Problem1
  • Inadequacy of Device Shape and/or Size1
  • Unstable1

Patient problems

  • Subluxation4
  • Implant Pain2
  • Cyst(s)1
  • Joint Laxity1
  • Loss of Range of Motion1
  • Osteolysis1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjurySep 12, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    SUBJECT: (B)(6); (WELBOW CLINICAL STUDY). FEW DAYS AFTER THE SURGERY INVOLVING THE CRF II IMPLANT, A DISLOCATION OF THE ELBOW (IN THE EVENT OF MAJOR INSTABILITY) WAS OBSERVED. IT WAS OBSERVED A AVULSION OF THE REATTACHMENT ANCHOR OF THE LATERAL LIGAMENTOUS PLAN RESULTED IN POSTERIOR DISLOCATION OF THE RADIAL HEAD. THIS EVENT WAS REPORTED AS A SERIOUS ADVERSE EVENT ¿REQUIRED INTERVENTION TO PREVENT LIFE-THREATENING ILLNESS OR INJURY, OR PERMANENT IMPAIRMENT OR DAMAGE TO ADDRESS THIS ISSUE, THREE INTERVENTIONS WERE REQUIRED : ON 27 JUN 2022 : RE-OPERATION WITHOUT IMPLANT CHANGE : LIGAMENT RECONSTRUCTION USING HALF OF THE FLEXOR CARPI RADIALIS TENDON. THIS RECORD CORRESPONDS TO THIS EVENT ABOVE. ON 12 JUL 2022 : RE-OPERATION WITHOUT IMPLANT CHANGE : LIGAMENT RECONSTRUCTION USING THE CONTROLATERAL PALMARIS LONGUS TENDON (BECAUSE OF RECURRENCE OF POSTEROLATERAL RADIAL DISLOCATION AFTER THE FIRST LIGAMENT). THIS EVENT CORRESPONDS TO 20243733255 AND 20243733256. ON 13 JUL 2022 : RE-OPERATION WITHOUT IMPLANT CHANGE : LIGAMENT RECONSTRUCTION (BECAUSE OF PERSISTENCE OF THE POSTEROLATERAL RADIAL HEAD DISLOCATION) THIS EVENT WAS RESOLVED WITHOUT SEQUELAE AFTER LAST INTERVENTION. THIS EVENT COR

    Patient problems: Subluxation

    MAUDE report 20200608 ↗
  • InjurySep 12, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    SUBJECT: (B)(6) ((B)(6) CLINICAL STUDY). FEW DAYS AFTER THE SURGERY INVOLVING THE CRF II IMPLANT, A DISLOCATION OF THE ELBOW (IN THE EVENT OF MAJOR INSTABILITY) WAS OBSERVED. IT WAS OBSERVED A AVULSION OF THE REATTACHMENT ANCHOR OF THE LATERAL LIGAMENTOUS PLAN RESULTED IN POSTERIOR DISLOCATION OF THE RADIAL HEAD. THIS EVENT WAS REPORTED AS A SERIOUS ADVERSE EVENT ¿REQUIRED INTERVENTION TO PREVENT LIFE-THREATENING ILLNESS OR INJURY, OR PERMANENT IMPAIRMENT OR DAMAGE TO ADDRESS THIS ISSUE, THREE INTERVENTIONS WERE REQUIRED : ON (B)(6) 2022 : RE-OPERATION WITHOUT IMPLANT CHANGE : LIGAMENT RECONSTRUCTION USING HALF OF THE FLEXOR CARPI RADIALIS TENDON. THIS EVENT CORRESPONDS TO (B)(4) AND (B)(4). ON (B)(6) 2022 : RE-OPERATION WITHOUT IMPLANT CHANGE : LIGAMENT RECONSTRUCTION USING THE CONTROLATERAL PALMARIS LONGUS TENDON (BECAUSE OF RECURRENCE OF POSTEROLATERAL RADIAL DISLOCATION AFTER THE FIRST LIGAMENT). THIS EVENT CORRESPONDS TO (B)(4) AND (B)(4). ON (B)(6) 2022 : RE-OPERATION WITHOUT IMPLANT CHANGE : LIGAMENT RECONSTUCTION (BECAUSE OF PERSISTENCE OF THE POSTEROLATERAL RADIAL HEAD DISLOCATION) THIS EVENT WAS RESOLVED WITHOUT SEQUELAE AFTER LAST INTERVENTION. THIS RECORD COVERS THIS EV

    Patient problems: Subluxation

    MAUDE report 20200580 ↗
  • InjurySep 12, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    SUBJECT: (B)(6) ((B)(6) CLINICAL STUDY). FEW DAYS AFTER THE SURGERY INVOLVING THE CRF II IMPLANT, A DISLOCATION OF THE ELBOW (IN THE EVENT OF MAJOR INSTABILITY) WAS OBSERVED. IT WAS OBSERVED A AVULSION OF THE REATTACHMENT ANCHOR OF THE LATERAL LIGAMENTOUS PLAN RESULTED IN POSTERIOR DISLOCATION OF THE RADIAL HEAD. THIS EVENT WAS REPORTED AS A SERIOUS ADVERSE EVENT ¿REQUIRED INTERVENTION TO PREVENT LIFE-THREATENING ILLNESS OR INJURY, OR PERMANENT IMPAIRMENT OR DAMAGE TO ADDRESS THIS ISSUE, THREE INTERVENTIONS WERE REQUIRED: ON (B)(6) 2022 : RE-OPERATION WITHOUT IMPLANT CHANGE : LIGAMENT RECONSTRUCTION USING HALF OF THE FLEXOR CARPI RADIALIS TENDON. THIS EVENT CORRESPONDS TO (B)(4) AND (B)(4). ON (B)(6) 2022 : RE-OPERATION WITHOUT IMPLANT CHANGE : LIGAMENT RECONSTRUCTION USING THE CONTROLATERAL PALMARIS LONGUS TENDON (BECAUSE OF RECURRENCE OF POSTEROLATERAL RADIAL DISLOCATION AFTER THE FIRST LIGAMENT). THIS RECORD CORRESPONDS TO THIS EVENT. ON (B)(6) 2022 : RE-OPERATION WITHOUT IMPLANT CHANGE : LIGAMENT RECONSTUCTION (BECAUSE OF PERSISTENCE OF THE POSTEROLATERAL RADIAL HEAD DISLOCATION) THIS EVENT WAS RESOLVED WITHOUT SEQUELAE AFTER LAST INTERVENTION. THIS EVENT CORRESPONDS TO (B)(4)

    Patient problems: Subluxation

    MAUDE report 20200565 ↗
  • InjuryMay 21, 2024Unstable; Device Dislodged or Dislocated; Unstable; Device Dislodged or Dislocated

    AS REPORTED: "DISLOCATION OF THE ELBOW (IN THE EVENT OF MAJOR INSTABILITY) PROSTHESIS PLACEMENT ON (B)(6) 2022, SUBLUXATION NOTED ON THE VISIT OF (B)(6) 2022, PROSTHESIS CHANGE ON (B)(6) 2022. TO ADDRESS THIS MATTER COMPREHENSIVELY, A CORRECTIVE RE-OPERATION WAS CONDUCTED. THIS INVOLVED THE EXTRACTION OF THE EXISTING IMPLANT AND THE SUBSEQUENT RE-IMPLANTATION OF A NEW PROSTHESIS."

    Patient problems: Joint Laxity; Subluxation

    MAUDE report 19366918 ↗
  • InjuryMar 1, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED: "POST OPERATIVE COMPLICATION: BONE CYST / OSTEOLYSIS / BONE RESORPTION. SEVERE PAIN OF THE LEFT ELBOW POSTERO-LATERAL. PARTIAL LOSS OF RANGE OF MOTION." UPDATE FOLLOWING CLINICAL TEAM FEEDBACK: "TWO YEARS POST-SURGERY, THE PATIENT ENCOUNTERED INTENSE PAIN IN THE POSTERO-LATERAL REGION OF THE LEFT ELBOW, ACCOMPANIED BY A PARTIAL LOSS OF RANGE OF MOTION. SUBSEQUENT TO A FOLLOW-UP VISIT, A BONE CYST WAS CONCLUSIVELY DIAGNOSED MARKING A SIGNIFICANT ADVERSE EVENT (SAE) ¿PERMANENT IMPAIRMENT OF A BODY STRUCTURE OR FUNCTION¿. ACCORDING TO THE SURGEON, THIS EVENT HAS A CAUSAL RELATIONSHIP TO THE IMPLANT AND A POSSIBLE RELATIONSHIP TO THE SURGERY. TO ADDRESS THIS, REVISION SURGERY WAS PERFORMED ON (B)(6) 2024 TO REMOVE THE PROSTHESIS, BUT WITHOUT ANY REPLACEMENT. THE EVENT WAS RESOLVED WITHOUT SEQUELAE ON 16 JAN 2024. THE REMOVAL OF THE PROSTHESIS RESULTED IN COMPLETE ALLEVIATION OF PAIN, RENDERING THE LEFT ELBOW PAIN-FREE, STABLE, AND FULLY RESTORED TO ITS FULL RANGE OF MOTION."

    Patient problems: Cyst(s); Implant Pain; Loss of Range of Motion; Osteolysis

    MAUDE report 18817473 ↗
  • InjuryApr 4, 2023Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    IT WAS REPORTED THAT THE PATIENT HAD PAIN FROM THE INITIAL IMPLANTATION OF AN ELBOW PROSTHESIS 4 MONTHS PRIOR. THE SURGEON ANALYZED THAT THE PROSTHESIS WAS OVERSTUFFED AND HAD CREATED PAIN ON THE HUMERAL CARTILAGE. REVISION WAS PERFORMED WITH A MOPYC HEAD SMALL, NECK MEDIUM AND STEM SMALL. SURGEON HAD DIFFICULTLY TO REMOVE CEMENT FROM BONE SHAFT, BUT AFTER THIS STEP IMPLANTATION OF THE MOPYC WAS EASY.

    Patient problems: Implant Pain

    MAUDE report 16673618 ↗

Other PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER devices

All PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER devices →

Sources and updates