Skip to content

AEQUALIS GLENOID DWB216Recalls & reported problems

Tornier · Class 2 · Implant

All models of AEQUALIS GLENOID (11)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 03700386915468), received between Dec 19, 2016 and Mar 28, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Migration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Migration1
  • Naturally Worn1
  • Osseointegration Problem1

Patient problems

  • Implant Pain1
  • Osteolysis1
  • Patient Problem/Medical Problem1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT REQUIRED REVISION SURGERY DUE TO PAIN.

    Patient problems: Implant Pain

    MAUDE report 18999680 ↗
  • InjuryMay 9, 2023Naturally Worn; Osseointegration Problem; Migration; Naturally Worn; Osseointegration Problem; Migration

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN APRIL 2012 ¿ MAY 2022. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2022 A PATIENT REQUIRED REVISION SURGERY DUE "ASEPTIC LOOSENING GLENOID/GLENOID IMPLANT WEAR/LYSIS - HUMERUS/LYSIS - GLENOID", WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER. REVISION PROCEDURE TYPE : "SINGLE STAGE REVISION¿.

    Patient problems: Osteolysis

    MAUDE report 16898439 ↗
  • InjuryDec 19, 2016Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT REVISION SURGERY AT 10 YEARS POST-OPERATIVE FOR ROTATOR CUFF INSUFFICIENCY. THE IMPLANTED STEM, HUMERAL HEAD AND GLENOID COMPONENT WERE REMOVED. CONVERSION INTO RSA.

    Patient problems: Patient Problem/Medical Problem

    MAUDE report 6184866 ↗

Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page