AEQUALIS REVERSED DWD144Recalls & reported problems
Tornier · Class 2 · Implant
All models of AEQUALIS REVERSED (41)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 03700386917127), received between Dec 29, 2021 and Sep 21, 2026 — classified as 5 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Break3
- Fracture2
- Osseointegration Problem1
- Unstable1
Patient problems
- Implant Pain2
- No Clinical Signs, Symptoms or Conditions2
- Bacterial Infection1
- Fall1
- Inadequate Osseointegration1
- Joint Laxity1
- Muscle/Tendon Damage1
- Post Operative Wound Infection1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjurySep 21, 2026Adverse Event Without Identified Device or Use Problem
A revision surgery was performed as a result of a suspected infection after the patient fell.
Patient problems: Fall; Post Operative Wound Infection
MAUDE report 50841157 ↗InjuryJul 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, THE PATIENT WAS REVISED DUE TO AN INFECTION. ALL THE IMPLANTS WERE REMOVED. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Unspecified Infection
MAUDE report 22668366 ↗InjuryJul 29, 2025Break; Fracture; Osseointegration Problem; Break; Fracture; Osseointegration Problem
IT WAS REPORTED THAT, 2 SCREWS BROKE (CATALOG# DWD138 AND DWD144) AND THE ENTIRE PROSTHESIS (BASEPLATE) WAS LOOSE. THE ISSUE WAS NOTICED BY X RAY. REQUIRED A REVISION PROCEDURE.
Patient problems: Inadequate Osseointegration
MAUDE report 22649387 ↗InjuryJan 9, 2025Unstable; Unstable
PATIENT UNDERWENT PRIMARY LEFT REVERSE SHOULDER REPLACEMENT ON (B)(6) 2019. SURGEON STATES THAT PROSTHETIC JOINT HAS BEEN FUNCTIONING WELL, UNTIL APPROXIMATELY 6 MONTHS AGO, WHEN PATIENT EXPERIENCED SUDDEN ONSET OF PAIN, AND INCREASING INSTABILITY (PATIENT HAD BEEN DOING EXERCISES WITH WEIGHTS). DURING SURGERY, PROSTHETIC JOINT WAS FOUND TO BE VERY LAX, EASILY DISLOCATED, WITH SUB-SCAPULARIS TENDON RUPTURED. SURGEON REPAIRED SUB-SCAP., PLUS TENDON GRAFT / REPAIR, THEN REMOVED THE 9MM HUMERAL POLY INSERT, REPLACING THIS WITH A 9MM SPACER PLUS A NEW 9MM POLY INSERT. ADDITIONAL INFORMATION: THE LOCKING SCREW THAT CAME WITH THE 36MM SPACER, WOULD NOT ENGAGE WITH THE IMPLANT. THE 42MM SPACER WAS OPENED TO ACCESS ANOTHER LOCKING SCREW. THIS SCREW WAS FOUND TO ENGAGE THE IMPLANTED METAPHYSIS, LOCKING THE IMPACTED SPACER TO THE METAPHYSIS. THE FAULTY LOCKING SCREW IS AVAILABLE FOR RETURN. JOINT WAS FOUND TO BE STABLE THROUGH FULL RANGE OF MOVEMENT, SOFT TISSUE ENVELOP RECONSTRUCTED (SEPARATE COMPLAINT RAISED FOR SCREW).
Patient problems: Implant Pain; Joint Laxity; Muscle/Tendon Damage
MAUDE report 21110806 ↗InjuryApr 22, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "PATIENT UNDERWENT LEFT PRIMARY SHOULDER REPLACEMENT ON (B)(6) 2020. PATIENT PRESENTED TO SURGEON IN 2023, WITH A HISTORY OF A PAINFUL LEFT SHOULDER, AND A DISCHARGING SINUS (LEFT AXILLA). SWABS GREW THE BACTERIA C. ACNES. PLAN: REMOVE ALL PROSTHESIS (1ST STAGE) / INSERT AN ANTIBIOTIC CEMENT SPACER; RETURN IN SEVERAL MONTHS, WHEN CLINICAL MARKERS INDICATE INFECTION RESOLVED, AND IMPLANT A NEW PROSTHETIC LEFT SHOULDER JOINT(2ND STAGE). TODAY, ALL PROSTHESIS REMOVED, DISCARDED. MULTIPLE SPECIMENS TAKEN FOR PATHOLOGY"
Patient problems: Bacterial Infection; Implant Pain
MAUDE report 19153479 ↗MalfunctionFeb 25, 2022Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THE MULTI DIRECTIONAL LOCKING SCREW BROKE 4.5MM X 44 ON THE REVERSED 2 GLENOID SYSTEM. THE SCREW WAS PUT IN ON HAND AND THE HEX HEAD SNAPPED OFF WITH THE MOST PROXIMAL LOCKING THREADS STILL ATTACHED. IF THE BASEPLATE IS TO BE REVISED THE SCREW WILL HAVE NO MEANS OF BEING BACKED OUT WITHOUT MIDAS REX/ DIAMOND TIPPED BURR. THE SCREW IS IN SITU AND THE LOCKING HEAD FELL ONTO THE FLOOR AFTER BEING REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13600168 ↗MalfunctionDec 29, 2021Break; Break
LOCKING SCREW FROM REVERSE 2 GLENOID SYSTEM BROKE OFF WHEN TIGHTENING. THE 4.5MM X 44MM MULTIDIRECTIONAL SCREW USED WITH THE REVERSE 2 BASEPLATE WAS POWERED INTO THE BASEPLATE AT A SLOW SPEED AND ONCE THE LOCKING THREADS STARTED TO ENGAGE THE DRIVER WAS REMOVED, HOWEVER THE HEAD OF THE SCREW CAME OFF WITH THE DRIVER. THE LOCKING THREADS ON THE SCREW DID ENGAGE WITH THE LOCKING ISLET, HOWEVER IN A REVISION SITUATION THERE IS NO HEAD OF THE SCREW TO AID THE REMOVAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13107528 ↗
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026