AEQUALIS REVERSED II DWD180Recalls & reported problems
Tornier · Class 2 · Implant
All models of AEQUALIS REVERSED II (46)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Tornier Aequalis Reversed II shoulder prosthesis has 108 adverse event reports, with no deaths and 94 injuries. Many reports describe the joint dislocating or partly slipping out of place, the glenoid part failing or loosening, and some describe infection requiring revision surgery. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 108
- Deaths reported
- 0
- Injuries reported
- 94
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 108 adverse event reports that list this exact device (DI 03700386935824), received between Nov 22, 2016 and Sep 18, 2026 — classified as 94 injuries, 14 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (30 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem45
- Device Dislodged or Dislocated30
- Device Difficult to Setup or Prepare9
- Mechanical Jam7
- Loosening of Implant Not Related to Bone-Ingrowth5
- Unstable5
- Inadequacy of Device Shape and/or Size4
- Migration4
Patient problems
- Unspecified Infection28
- Joint Dislocation17
- Failure of Implant16
- Patient Problem/Medical Problem13
- Subluxation11
- Implant Pain8
- Joint Laxity8
- No Clinical Signs, Symptoms or Conditions8
Recent adverse event reports
The 25 most recent of 108 reports, in the reporter's own words as published by the FDA.
InjurySep 18, 2026Device Dislodged or Dislocated
The manufacturer received a report from the National Joint Registry that contains unpublished collected data on the usage and the outcomes of the shoulder joint replacement. The report details analysis provided for revision procedures performed until August 2026.During the review of the report, it was identified that on (b)(6) 2025 a patient required revision surgery due to DISLOCATION/SUBLUXATION, which was not previously reported to the manufacturer.
Patient problems: Subluxation
MAUDE report 50815721 ↗InjuryJun 8, 2026Device Dislodged or Dislocated
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2026. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2023 A PATIENT REQUIRED REVISION SURGERY DUE TO DISLOCATION/SUBLUXATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Subluxation
MAUDE report 25426222 ↗InjuryMar 3, 2026Unstable
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (b)(6) 2024.DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (b)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO INSTABILITY WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Joint Laxity
MAUDE report 24494103 ↗InjuryFeb 16, 2026Device Dislodged or Dislocated
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO DISLOCATION/SUBLUXATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Subluxation
MAUDE report 24342758 ↗InjuryFeb 16, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 24342504 ↗InjuryFeb 16, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL AUG. 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 24342417 ↗InjuryFeb 13, 2026Device Dislodged or Dislocated
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED BETWEEN MARCH 2023 - SEPTEMBER 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO CUFF INSUFFICIENCY, OTHER, DISLOCATION, SUBLUXATION, TRAUMATIC DISLOCATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Subluxation
MAUDE report 24331299 ↗InjuryFeb 12, 2026Device Dislodged or Dislocated
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED BETWEEN MARCH 2023 - SEPTEMBER 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO COMPONENT DISSOCIATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Subluxation
MAUDE report 24323706 ↗InjuryFeb 12, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED IN 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2025 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION AND PERIPROSTHETIC FRACTURE, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Bone Fracture(s); Unspecified Infection
MAUDE report 24323133 ↗InjuryJan 28, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2023 A PATIENT REQUIRED REVISION SURGERY DUE TO HETEROTOPIC OSSIFICATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Extraskeletal ossification
MAUDE report 24198187 ↗InjuryJan 28, 2026Unstable
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024, A PATIENT REQUIRED REVISION SURGERY DUE TO INSTABILITY, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Joint Laxity
MAUDE report 24197891 ↗InjuryJan 6, 2026Device Dislodged or Dislocated
IT WAS REPORTED THAT, THE PATIENT HAD A REVISION FOR A BIO-RSA GLENOID PATIENT WHICH HAD APPEARED TO HAVE A RE-ABSORBED. THE SURGEON REMOVED THE GLENOSPHERE AND BASEPLATE WITH EASE FROM THE GLENOID SOCKET AND PACKED IT FULL WITH FEMORAL HEAD ALLOGRAFT. THE TRAY AND POLY WERE REMOVED FROM THE FLEX STEM AND SURGEON PLACED A 50X16 COCR FLEX HUMERAL HEAD ONTO THE STEM TO TURN IT TO A HEMI. IT WAS REPORTED IN A FINAL REPORT FROM SURGEON, THE RIGHT SHOULDER PROSTHESIS WAS DISLOCATED WITH THE HUMERAL END LYING INFERIOR TO THE GLENOID PROSTHETIC MARGIN. EXTENSIVE OSTEOLYSIS SURROUNDING THE GLENOID COMPONENT. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Subluxation
MAUDE report 23976686 ↗InjuryJan 5, 2026Unstable
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED UNTIL JUNE 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO INSTABILITY, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER. REVISION PROCEDURE TYPE: SINGLE STAGE REVISION (INCLUDES DAIR WITH MODULAR EXCHANGE AND MODULAR EXCHANGE FOR INDICATIONS OTHER THAN INFECTION)
Patient problems: Joint Laxity
MAUDE report 23957562 ↗InjuryDec 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT PRIMARY LEFT SHOULDER REVERSE ARTHROPLASTY ON (B)(6) 2022. HE EXPERIENCED PAIN AND DISCOMFORT, AND DECREASED RANGE OF MOTION OVER LAST 6 MONTHS. DURING SURGERY, SURGEON FOUND DARK STAINED TISSUE AROUND PROSTHETIC JOINTS. DEBRIDED, NO TISSUE SAMPLES TAKEN. HUMERAL STEMS ARE FOUND TO BE WELL FIXED. POLY WAS ALSO REMOVED (CONCOMITANT). GLENOSPHERE WAS FOUND TO BE LOOSE, BUT NO DAMAGE TO LOCKING SCREW (NOT BROKEN). BASEPLATE WAS FOUND TO BE WELL FIXED, WITH NO BONE LOSS. 36/25MM CENTERED GLENOSPHERE REPLACED, AS WITH A 36 / 6MM STANDARD POLY. JOINT REDUCED, FOUND TO BE STABLE, ACCEPTABLE RANGE OF MOTION. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Discomfort; Loss of Range of Motion; Metal Related Pathology; Pain
MAUDE report 23783173 ↗InjuryNov 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2025 A PATIENT REQUIRED REVISION SURGERY DUE TO CUFF INSUFFICIENCY, LYSIS - HUMERUS & LYSIS - GLENOID, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Osteolysis
MAUDE report 23671487 ↗InjuryOct 8, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO DISLOCATION/SUBLUXATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Subluxation
MAUDE report 23244528 ↗InjuryAug 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD INFECTION AFTER SHOULDER REPLACEMENT SURGERY. PAIN WAS ALSO REPORTED. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: Pain; Unspecified Infection
MAUDE report 22795313 ↗InjuryJul 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED: "THE INITIAL SURGERY WAS PERFORMED ON (B)(6) 2025, AND ON (B)(6) DURING A FOLLOW-UP OUTPATIENT VISIT, THE DOCTOR CONFIRMED THAT THE PATIENT HAD A DISLOCATION. SINCE THE DISLOCATION WAS SELF-REDUCED, THE PATIENT WAS ADVISED TO WAIT AND SEE, BUT SINCE THE PATIENT REQUESTED A REOPERATION, A REOPERATION WAS SCHEDULED FOR (B)(6)."
Patient problems: Joint Dislocation
MAUDE report 22440559 ↗InjuryJul 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, A SHOULDER REVISION WAS PLANNED ON (B)(6) 2025 FOR INFECTION. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: Unspecified Infection
MAUDE report 22374116 ↗InjuryMay 26, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2025 A PATIENT REQUIRED REVISION SURGERY DUE TO DISLOCATION/SUBLUXATION AND TRAUMATIC DISLOCATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Joint Dislocation; Subluxation
MAUDE report 22083320 ↗InjuryMay 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN (B)(6) 2013 ¿ (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2023, A PATIENT REQUIRED REVISION SURGERY DUE TO HETEROTOPIC OSSIFICATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Extraskeletal ossification
MAUDE report 22056348 ↗InjuryApr 28, 2025Osseointegration Problem; Osseointegration Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN (B)(6) 2013 ¿ (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2023 A PATIENT REQUIRED REVISION SURGERY DUE TO ASEPTIC LOOSENING GLENOID, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Inadequate Osseointegration
MAUDE report 21919086 ↗InjuryApr 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REVISION OF REV II HUMERAL & GLENOID. HARD TO GET BOTH STEM AND GLENOSPHERE OUT. GT FRACTURE AND GLENOID BONE LOSS. REASON FOR REVISION: PAIN, SUSPECTED POTENTIAL INFECTION.
Patient problems: Medical device site infection; Pain
MAUDE report 21803706 ↗InjuryMar 19, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN DECEMBER 2013 ¿ FEBRUARY 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024, A PATIENT REQUIRED REVISION SURGERY DUE TO DISLOCATION/SUBLUXATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Subluxation
MAUDE report 21644295 ↗InjuryMar 18, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
THE AEQUALIS DEVICES WERE IMPLANTED TO THE PATIENT'S RIGHT SHOULDER. THE DEVICES WERE REMOVED BECAUSE THE PATIENT GOT AN INFECTION (MAYBE FROM KIDNEY) WHILE THE SURGEON THOUGHT THAT THE INFECTION WAS NOT CAUSED BY THE IMPLANT. BECAUSE THE PATIENT WAS QUITE ELDERLY, REVISION WAS NOT CONSIDERED, AND THE SURGERY WAS COMPLETED WITH A CEMENT MOLD CONTAINING ANTIBIOTICS.
Patient problems: Unspecified Infection
MAUDE report 21631720 ↗
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page