AEQUALIS REVERSED II DWD180Recalls & reported problems

Tornier · Class 2 · Implant

All models of AEQUALIS REVERSED II (46)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Tornier Aequalis Reversed II shoulder prosthesis has 108 adverse event reports, with no deaths and 94 injuries. Many reports describe the joint dislocating or partly slipping out of place, the glenoid part failing or loosening, and some describe infection requiring revision surgery. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
108
Deaths reported
0
Injuries reported
94
Recalls naming it
0

Summary

The FDA's MAUDE database contains 108 adverse event reports that list this exact device (DI 03700386935824), received between Nov 22, 2016 and Sep 18, 2026 — classified as 94 injuries, 14 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (30 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 5 reportsJul 2024: 1 reportsAug 2024: 3 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 2 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 4 reportsFeb 2026: 6 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem45
  • Device Dislodged or Dislocated30
  • Device Difficult to Setup or Prepare9
  • Mechanical Jam7
  • Loosening of Implant Not Related to Bone-Ingrowth5
  • Unstable5
  • Inadequacy of Device Shape and/or Size4
  • Migration4

Patient problems

  • Unspecified Infection28
  • Joint Dislocation17
  • Failure of Implant16
  • Patient Problem/Medical Problem13
  • Subluxation11
  • Implant Pain8
  • Joint Laxity8
  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 25 most recent of 108 reports, in the reporter's own words as published by the FDA.

  • InjurySep 18, 2026Device Dislodged or Dislocated

    The manufacturer received a report from the National Joint Registry that contains unpublished collected data on the usage and the outcomes of the shoulder joint replacement. The report details analysis provided for revision procedures performed until August 2026.During the review of the report, it was identified that on (b)(6) 2025 a patient required revision surgery due to DISLOCATION/SUBLUXATION, which was not previously reported to the manufacturer.

    Patient problems: Subluxation

    MAUDE report 50815721 ↗
  • InjuryJun 8, 2026Device Dislodged or Dislocated

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2026. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2023 A PATIENT REQUIRED REVISION SURGERY DUE TO DISLOCATION/SUBLUXATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.

    Patient problems: Subluxation

    MAUDE report 25426222 ↗
  • InjuryMar 3, 2026Unstable

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (b)(6) 2024.DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (b)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO INSTABILITY WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.

    Patient problems: Joint Laxity

    MAUDE report 24494103 ↗
  • InjuryFeb 16, 2026Device Dislodged or Dislocated

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO DISLOCATION/SUBLUXATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.

    Patient problems: Subluxation

    MAUDE report 24342758 ↗
  • InjuryFeb 16, 2026Adverse Event Without Identified Device or Use Problem

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.

    Patient problems: Unspecified Infection

    MAUDE report 24342504 ↗
  • InjuryFeb 16, 2026Adverse Event Without Identified Device or Use Problem

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL AUG. 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.

    Patient problems: Unspecified Infection

    MAUDE report 24342417 ↗
  • InjuryFeb 13, 2026Device Dislodged or Dislocated

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED BETWEEN MARCH 2023 - SEPTEMBER 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO CUFF INSUFFICIENCY, OTHER, DISLOCATION, SUBLUXATION, TRAUMATIC DISLOCATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.

    Patient problems: Subluxation

    MAUDE report 24331299 ↗
  • InjuryFeb 12, 2026Device Dislodged or Dislocated

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED BETWEEN MARCH 2023 - SEPTEMBER 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO COMPONENT DISSOCIATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.

    Patient problems: Subluxation

    MAUDE report 24323706 ↗
  • InjuryFeb 12, 2026Adverse Event Without Identified Device or Use Problem

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED IN 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2025 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION AND PERIPROSTHETIC FRACTURE, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.

    Patient problems: Bone Fracture(s); Unspecified Infection

    MAUDE report 24323133 ↗
  • InjuryJan 28, 2026Adverse Event Without Identified Device or Use Problem

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2023 A PATIENT REQUIRED REVISION SURGERY DUE TO HETEROTOPIC OSSIFICATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.

    Patient problems: Extraskeletal ossification

    MAUDE report 24198187 ↗
  • InjuryJan 28, 2026Unstable

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024, A PATIENT REQUIRED REVISION SURGERY DUE TO INSTABILITY, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.

    Patient problems: Joint Laxity

    MAUDE report 24197891 ↗
  • InjuryJan 6, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT, THE PATIENT HAD A REVISION FOR A BIO-RSA GLENOID PATIENT WHICH HAD APPEARED TO HAVE A RE-ABSORBED. THE SURGEON REMOVED THE GLENOSPHERE AND BASEPLATE WITH EASE FROM THE GLENOID SOCKET AND PACKED IT FULL WITH FEMORAL HEAD ALLOGRAFT. THE TRAY AND POLY WERE REMOVED FROM THE FLEX STEM AND SURGEON PLACED A 50X16 COCR FLEX HUMERAL HEAD ONTO THE STEM TO TURN IT TO A HEMI. IT WAS REPORTED IN A FINAL REPORT FROM SURGEON, THE RIGHT SHOULDER PROSTHESIS WAS DISLOCATED WITH THE HUMERAL END LYING INFERIOR TO THE GLENOID PROSTHETIC MARGIN. EXTENSIVE OSTEOLYSIS SURROUNDING THE GLENOID COMPONENT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Subluxation

    MAUDE report 23976686 ↗
  • InjuryJan 5, 2026Unstable

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED UNTIL JUNE 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO INSTABILITY, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER. REVISION PROCEDURE TYPE: SINGLE STAGE REVISION (INCLUDES DAIR WITH MODULAR EXCHANGE AND MODULAR EXCHANGE FOR INDICATIONS OTHER THAN INFECTION)

    Patient problems: Joint Laxity

    MAUDE report 23957562 ↗
  • InjuryDec 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT PRIMARY LEFT SHOULDER REVERSE ARTHROPLASTY ON (B)(6) 2022. HE EXPERIENCED PAIN AND DISCOMFORT, AND DECREASED RANGE OF MOTION OVER LAST 6 MONTHS. DURING SURGERY, SURGEON FOUND DARK STAINED TISSUE AROUND PROSTHETIC JOINTS. DEBRIDED, NO TISSUE SAMPLES TAKEN. HUMERAL STEMS ARE FOUND TO BE WELL FIXED. POLY WAS ALSO REMOVED (CONCOMITANT). GLENOSPHERE WAS FOUND TO BE LOOSE, BUT NO DAMAGE TO LOCKING SCREW (NOT BROKEN). BASEPLATE WAS FOUND TO BE WELL FIXED, WITH NO BONE LOSS. 36/25MM CENTERED GLENOSPHERE REPLACED, AS WITH A 36 / 6MM STANDARD POLY. JOINT REDUCED, FOUND TO BE STABLE, ACCEPTABLE RANGE OF MOTION. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Discomfort; Loss of Range of Motion; Metal Related Pathology; Pain

    MAUDE report 23783173 ↗
  • InjuryNov 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2025 A PATIENT REQUIRED REVISION SURGERY DUE TO CUFF INSUFFICIENCY, LYSIS - HUMERUS & LYSIS - GLENOID, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.

    Patient problems: Osteolysis

    MAUDE report 23671487 ↗
  • InjuryOct 8, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO DISLOCATION/SUBLUXATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.

    Patient problems: Subluxation

    MAUDE report 23244528 ↗
  • InjuryAug 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD INFECTION AFTER SHOULDER REPLACEMENT SURGERY. PAIN WAS ALSO REPORTED. NO FURTHER INFORMATION WAS AVAILABLE.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 22795313 ↗
  • InjuryJul 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED: "THE INITIAL SURGERY WAS PERFORMED ON (B)(6) 2025, AND ON (B)(6) DURING A FOLLOW-UP OUTPATIENT VISIT, THE DOCTOR CONFIRMED THAT THE PATIENT HAD A DISLOCATION. SINCE THE DISLOCATION WAS SELF-REDUCED, THE PATIENT WAS ADVISED TO WAIT AND SEE, BUT SINCE THE PATIENT REQUESTED A REOPERATION, A REOPERATION WAS SCHEDULED FOR (B)(6)."

    Patient problems: Joint Dislocation

    MAUDE report 22440559 ↗
  • InjuryJul 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, A SHOULDER REVISION WAS PLANNED ON (B)(6) 2025 FOR INFECTION. NO FURTHER INFORMATION WAS AVAILABLE.

    Patient problems: Unspecified Infection

    MAUDE report 22374116 ↗
  • InjuryMay 26, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2025 A PATIENT REQUIRED REVISION SURGERY DUE TO DISLOCATION/SUBLUXATION AND TRAUMATIC DISLOCATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.

    Patient problems: Joint Dislocation; Subluxation

    MAUDE report 22083320 ↗
  • InjuryMay 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN (B)(6) 2013 ¿ (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2023, A PATIENT REQUIRED REVISION SURGERY DUE TO HETEROTOPIC OSSIFICATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.

    Patient problems: Extraskeletal ossification

    MAUDE report 22056348 ↗
  • InjuryApr 28, 2025Osseointegration Problem; Osseointegration Problem

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN (B)(6) 2013 ¿ (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2023 A PATIENT REQUIRED REVISION SURGERY DUE TO ASEPTIC LOOSENING GLENOID, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.

    Patient problems: Inadequate Osseointegration

    MAUDE report 21919086 ↗
  • InjuryApr 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION OF REV II HUMERAL & GLENOID. HARD TO GET BOTH STEM AND GLENOSPHERE OUT. GT FRACTURE AND GLENOID BONE LOSS. REASON FOR REVISION: PAIN, SUSPECTED POTENTIAL INFECTION.

    Patient problems: Medical device site infection; Pain

    MAUDE report 21803706 ↗
  • InjuryMar 19, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN DECEMBER 2013 ¿ FEBRUARY 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024, A PATIENT REQUIRED REVISION SURGERY DUE TO DISLOCATION/SUBLUXATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.

    Patient problems: Subluxation

    MAUDE report 21644295 ↗
  • InjuryMar 18, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    THE AEQUALIS DEVICES WERE IMPLANTED TO THE PATIENT'S RIGHT SHOULDER. THE DEVICES WERE REMOVED BECAUSE THE PATIENT GOT AN INFECTION (MAYBE FROM KIDNEY) WHILE THE SURGEON THOUGHT THAT THE INFECTION WAS NOT CAUSED BY THE IMPLANT. BECAUSE THE PATIENT WAS QUITE ELDERLY, REVISION WAS NOT CONSIDERED, AND THE SURGERY WAS COMPLETED WITH A CEMENT MOLD CONTAINING ANTIBIOTICS.

    Patient problems: Unspecified Infection

    MAUDE report 21631720 ↗

Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices →