AEQUALIS REVERSED II DWD190Recalls & reported problems
Tornier · Class 2 · Implant
All models of AEQUALIS REVERSED II (46)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Tornier Aequalis Reversed II shoulder implant is a cemented metal-and-polymer partial shoulder replacement. The 39 reports mostly describe the implant loosening or coming out of place, along with infection, bone fracture around the implant, and joint dislocation or partial dislocation, often leading to revision surgery. No deaths were reported and 36 injuries were. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 40
- Deaths reported
- 0
- Injuries reported
- 37
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 03700386935886), received between Aug 8, 2016 and Sep 21, 2026 — classified as 37 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem20
- Device Dislodged or Dislocated12
- Loosening of Implant Not Related to Bone-Ingrowth2
- Naturally Worn2
- Break1
- Detachment of Device or Device Component1
- Device Difficult to Setup or Prepare1
- Disassembly1
Patient problems
- Patient Problem/Medical Problem11
- Unspecified Infection8
- Subluxation5
- Bone Fracture(s)4
- Joint Dislocation4
- Failure of Implant3
- Implant Pain3
- Pain3
Recent adverse event reports
The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.
InjurySep 21, 2026Adverse Event Without Identified Device or Use Problem
A revision surgery was performed as a result of a suspected infection after the patient fell.
Patient problems: Fall; Post Operative Wound Infection
MAUDE report 50841161 ↗InjuryApr 14, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2026. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2026 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 24873783 ↗InjuryJan 5, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2025 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 23958155 ↗InjuryJan 5, 2026Device Dislodged or Dislocated
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2025 A PATIENT REQUIRED REVISION SURGERY DUE TO PERIPROSTHETIC FRACTURE, DISLOCATION & SUBLUXATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Bone Fracture(s); Subluxation
MAUDE report 23957840 ↗InjuryDec 2, 2025Unstable; Unstable
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED BETWEEN (B)(6) 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2025 A PATIENT REQUIRED REVISION SURGERY DUE TO INSTABILITY, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Joint Laxity
MAUDE report 23691659 ↗InjuryOct 9, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2021 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, ASEPTIC GLENOID LOOSENING, GLENOID LYSIS AND DISLOCATION/SUBLUXATION WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Inadequate Osseointegration; Medical device site infection; Osteolysis; Subluxation
MAUDE report 23255493 ↗InjuryAug 13, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT A PATIENT REQUIRED REVISION SURGERY DUE TO DISLOCATION/ SUBLUXATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Subluxation
MAUDE report 22792322 ↗InjuryAug 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 22747635 ↗InjuryApr 22, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2025 A PATIENT REQUIRED REVISION SURGERY DUE TO DISLOCATION/SUBLUXATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Joint Dislocation; Subluxation
MAUDE report 21882509 ↗InjuryApr 4, 2025Device Dislodged or Dislocated; Naturally Worn; Device Dislodged or Dislocated; Naturally Worn
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL FEBRUARY 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO PERIPROSTHETIC FRACTURE, GLENOID IMPLANT WEAR AND DISLOCATION/SUBLUXATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Bone Fracture(s); Subluxation
MAUDE report 21772198 ↗InjuryDec 31, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "SUBJECT (B)(6) SHOUT - TORNIER SHOULDER OUTCOMES STUDY PATIENT REPORTS PAIN AND SWELLING."
Patient problems: Implant Pain; Swelling/ Edema
MAUDE report 21045851 ↗InjuryDec 30, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
AS REPORTED: "SUBJECT: (B)(6) SHOUT - TORNIER SHOULDER OUTCOMES STUDY. SUBJECT NOTICED A SUDDEN INCREASE IN PAIN ON (B)(6) 2021, YET DID NOT SUFFER ANY ACCIDENT, INCIDENT, OR FALL. FOLLOW UP X-RAYS ON 18 AUG 2021 REVEALED AN ANTERIOR SUPERIOR REVERSE SHOULDER DISLOCATION. THE HUMERAL COMPONENTS WERE STILL STABLE AND THERE WAS A HINT OF BONE BRIDGING FROM THE NATIVE UPPER HUMERUS TO THE NATIVE LOWER HUMERUS ACROSS THE ALLOGRAFT. REVISION SURGERY ON (B)(6) 2021."
Patient problems: Implant Pain; Joint Dislocation
MAUDE report 21036006 ↗InjuryAug 6, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN APRIL 2012 ¿ MAY 2022. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2021 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER. REVISION PROCEDURE TYPE: SINGLE STAGE REVISION.
Patient problems: Unspecified Infection
MAUDE report 19918701 ↗InjuryJul 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN (B)(6)2012 ¿ (B)(6) 2022. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2022 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER. REVISION PROCEDURE TYPE: SINGLE STAGE REVISION.
Patient problems: Unspecified Infection
MAUDE report 19702259 ↗InjuryMar 27, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN (B)(6) 2012 ON (B)(6) 2022. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2018 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 18987274 ↗InjuryMar 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IN THE TORNIER SHOULDER OUTCOMES STUDY, A TRAUMATIC ACROMIAL FRACTURE OCCURRED POSTOPERATIVELY. THE ADVERSE EVENT RESOLVED WITHOUT SEQUELAE. IT WAS ASSESSED AS NOT RELATED TO THE STUDY DEVICE OR THE SURGICAL PROCEDURE. AN UPDATE NOTED THAT NO ACTION WAS TAKEN, AND NO TREATMENT WAS ADMINISTERED. DURING THE VISIT, GLENOID NOTCHING WAS OBSERVED, WITH THE NOTCH EXTENDING TO THE LOWER SCREW AND GLENOID LUCENCIES INVOLVING THE BASEPLATE (SUPERIOR OR INFERIOR).
Patient problems: Bone Fracture(s)
MAUDE report 18969300 ↗InjuryOct 10, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN (B)(6) 2012 ¿ (B)(6) 2022. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2018 A PATIENT REQUIRED REVISION SURGERY DUE UNEXPLAINED PAIN, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER. REVISION PROCEDURE TYPE : SINGLE STAGE REVISION.
Patient problems: Implant Pain
MAUDE report 17909367 ↗InjuryAug 22, 2023Naturally Worn; Naturally Worn
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN APRIL 2012 ¿ MAY 2022. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2019 A PATIENT REQUIRED REVISION SURGERY DUE STIFFNESS AND GLENOID IMPLANT WEAR, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Loss of Range of Motion
MAUDE report 17599372 ↗InjuryApr 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT (B)(6) UNDERWENT A REVISION SURGERY OF THE COMPLETE HUMERAL AND GLENOID SIDES DUE TO CALCIFIC PSEUDOTUMOR. NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: Seroma
MAUDE report 16778832 ↗MalfunctionMar 14, 2023Mechanical Jam; Mechanical Jam
THE GRUB SCREW OF THE GLENOSPHERE COULD NOT BE SCREWED; IT WAS TIGHT OR JAMMED AND THEREFORE COULD NOT BE SCREWED/SECURED INTO THE BASE PLATE. AS AN ALTERNATIVE, THE ECCENTRIC GLENOSPHERE WAS USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16542253 ↗InjuryFeb 25, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PAIN IS TRAVELING AND IS SO SEVERE AT TIMES. ALL INTO NECK AND DOWN ARM. SURGERY IN 2016 GOT BETTER BUT 6 MONTHS AGO, MY ARM IS MUCH MORE PAINFUL. I HAVE 105 + 2 LOT OF MOBILITY IN IT. CAN'T AFFORD TO GO TO DR AND CHECKED. I AM ON. WAS DOING REALLY GOOD ON HEALING. 6 MONTHS AGO STARTED GETTING WEAKER. THE PAIN IS ALL THE WAY INTO NECK AND MY HAND IS NUMB AND TINGLING. ELBOW PAIN. CAN'T AFFORD TO GO TO DOCTOR.
Patient problems: Pain
MAUDE report 13606874 ↗MalfunctionFeb 25, 2022Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare
THE SCREW WOULD NOT ENGAGE THE BASEPLATE (WHEN HE WENT TO SCREW, AFTER IMPACTING). THE CENTER SCREW KEPT SPINNING. THE GLENOID SPERE WAS LEFT IN PATIENT. DR. FELT THE SPHERE WAS STILL ATTACHED TO THE BASEPLATE IN A SOUND & SECURE FASHION.
Patient problems: Failure of Implant
MAUDE report 13606746 ↗InjuryMay 21, 2021Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem; Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED ON (B)(6) 2021 DUE TO INSTABILITY POST REVERSE TOTAL SHOULDER REPLACEMENT (TORNIER THREADED POST BASEPLATE, SCREWS, GLENOSPHERE, TRAY AND INSERT REMOVED BY THE SURGEON IN DURANGO CO). TORNIER FLEX PTC 6B LONG STEM WAS REPORTED WELL FIXED BY THE SURGEON, SO RETAINED IN PATIENT. THE PATIENT WAS COMPLIANT TO MOBILIZATION INSTRUCTIONS OF THE SURGEON BUT NOT THE LONG TERM WEIGHT RESTRICTIONS. THE SURGEON DID NOT WANT TO PROVIDE THE CT SCANS. THE PATIENT WAS PULLING HOSES 9 MONTHS POST OP AND FELT A "POP" THAT WAS PRESUMABLY A DISLOCATION HE STATES, MULTIPLE SELF REPORTED DISLOCATIONS, NONE DOCUMENTED UNTIL LATE WINTER EARLY OF 2021. THE SURGEON WANTS TO MAKE SURE IT¿S CLEARLY UNDERSTOOD THAT HE HAS NO COMPLAINT AGAINST OUR PRODUCTS, JUST REPORTING AS ASKED OF A REVISION TOTAL SHOULDER. REVERSED INSERT WERE IMPLANTED.
Patient problems: Failure of Implant; Joint Laxity
MAUDE report 11866404 ↗InjurySep 9, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
ON (B)(6) 2020: THIS PATIENT BECAME PSEUDO-ARTICULATED AFTER A PROXIMAL HUMERUS FRACTURE AND HAD AN AEQUALIS REVERSE SURGERY. ON (B)(6) 2020: THE SURGEON FOUND A DISLOCATION BETWEEN THE INSERT AND THE GLENOID SPHERE. ON (B)(6) 2020: REVISION SURGERY WAS PERFORMED FOR THE REPLACEMENT OF THE INSERT AND THE GLENOID SPHERE. TO ADJUST THE TENSION, DWD989 AND DWD194, WERE NEWLY USED RESPECTIVELY. THE REMAINING IMPLANTS WERE TO CONTINUE TO BE USED AS THEY WERE.
Patient problems: Joint Dislocation
MAUDE report 10506681 ↗InjuryJul 7, 2020Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
A PATIENT WHO HAD AN INITIAL SURGERY OF THE AEQUALIS REVERSE ON (B)(6) 2015, VISITED THE OUTPATIENT DEPARTMENT ON (B)(6) 2020, AND COMPLAINED OF PAIN IN THE AFFECTED AREA. UPON X-RAY EXAMINATION AND PUNCTURE EXAMINATION OF THE AFFECTED AREA, LOOSENESS OF GLENOID IMPLANT AND INFECTION WERE CONFIRMED. SURGERY WAS PERFORMED ON (B)(6) 2020. SINCE THE BASE PLATE ON THE GLENOID SIDE WAS UNSTABLE, ALL IMPLANTS ON THE GLENOID SIDE WERE REMOVED EXCEPT THE HUMERAL STEM IMPLANT AND THE METAPHYSIS IMPLANT, AND THE SURGERY WAS COMPLETED. THE SURGEON COMMENTED:REGARDING THE IMPLANT ON THE HUMERAL SIDE WAS CONFIRMED TO BE STABLE, SO IT WAS LEFT AS IT WAS.
Patient problems: Joint Dislocation; Pain; Unspecified Infection
MAUDE report 10238773 ↗
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page