AEQUALIS REVERSED II DWE730Recalls & reported problems
Tornier · Class 2 · Implant
All models of AEQUALIS REVERSED II (46)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 03700386952661), received between Sep 6, 2018 and Dec 31, 2024 — classified as 9 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated3
- Adverse Event Without Identified Device or Use Problem2
- Break2
- Loosening of Implant Not Related to Bone-Ingrowth2
- Insufficient Device Problem Information1
- Loss of Osseointegration1
- Unintended Ejection1
- Unstable1
Patient problems
- Failure of Implant4
- Implant Pain3
- Inadequate Osseointegration2
- Joint Dislocation2
- Bone Fracture(s)1
- Fall1
- Joint Laxity1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryDec 31, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "SUBJECT (B)(6) SHOUT - TORNIER SHOULDER OUTCOMES STUDY PAIN IN THE SHOULDER DUE TO A FALL."
Patient problems: Fall; Implant Pain
MAUDE report 21045853 ↗InjuryDec 31, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "SUBJECT 008-00078 SHOUT - TORNIER SHOULDER OUTCOMES STUDY PATIENT REPORTS PAIN AND SWELLING."
Patient problems: Implant Pain; Swelling/ Edema
MAUDE report 21045849 ↗InjuryJun 12, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT A PATIENT REQUIRED REVISION SURGERY AFTER DISLODGMENT AND POSSIBLE INFECTION.
Patient problems: Joint Dislocation; Unspecified Infection
MAUDE report 19521280 ↗InjuryJun 5, 2024Unstable; Device Dislodged or Dislocated; Unstable; Device Dislodged or Dislocated
AS REPORTED: "INSTABILITY/DISLOCATION. REVISION SURGERY (LEFT SHOULDER). HUMERAL SIDE COMPONENTS RETAINED FROM FIRST SURGERY". SERIES OF EVENTS BELOW: (B)(6) 2016: AEQUALIS ASCEND FLEX HUMERAL COMPONENTS / AEQUALIS REVERSED II GLENOID COMPONENTS (WITH A THREADED POST BASEPLATE) IMPLANTED (REVERSED CONFIGURATION). (B)(6) 2018: PATIENT FALL FROM BED AT HOME ("NO FRACTURE OR PROTHESIS DISLOCATION") *. (B)(6) 2018: INDICATION: MEDICAL EVALUATION OF PAIN FOLLOWING TRAUMA EVENT. DIFFUSE BONE DEMINERALIZATION. A DISLOCATION BACKWARDS AND DOWNWARDS OF THE SHOULDER PROSTHESIS. ABSENCE OF TRAUMATIC BONE LESION CLEARLY DEMONSTRATED (NUMEROUS ARTI-FACTS LINKED TO THE PROSTHESIS. (B)(6) 2018: INDICATION: MEDICAL EVALUATION AFTER SURGERY. DISLOCATION OF THE TOTAL LEFT REVERSED SHOULDER WITH A SMALL FRACTURE OF THE GLENOID INFERIOR SCREW. (B)(6) 2018: REVISION SURGERY WITH A REVERSED SHOULDER ARTHROPLASTY CONFIGURATION. -EXPLANTED: GLENOID COMPONENTS -RETAINED: HUMERAL COMPONENTS. -IMPLANTED: NEW BASEPLATE (WITH LONG POST BASEPLATE) AND SPHERE WITH GRAFT. -SLING, IMMOBILIZATION.
Patient problems: Implant Pain; Joint Dislocation; Joint Laxity
MAUDE report 19467895 ↗InjuryJun 6, 2022Loss of Osseointegration; Loss of Osseointegration
IT WAS REPORTED THAT THE PATIENT HAD A LEFT REVERSE SHOULDER IMPLANTED IN APPROXIMATELY 2019. THE GLENOID BASEPLATE COMPONENT OF THE SYSTEM BECAME LOOSE OVER TIME. THE GLENOID COMPONENT OF THE SYSTEM WAS REMOVED WITHOUT ISSUE AND THE IMPLANT WAS CONVERTED TO A HEMI SHOULDER WITH THE POLY REMOVED ON THE HUMERAL SIDE AND REPLACED WITH A CONVERTOR AND HUMERAL HEAD.
Patient problems: Inadequate Osseointegration; No Clinical Signs, Symptoms or Conditions
MAUDE report 14606044 ↗MalfunctionFeb 25, 2022Unintended Ejection; Unintended Ejection
InjuryJun 3, 2021Device Dislodged or Dislocated; Insufficient Device Problem Information; Device Dislodged or Dislocated; Insufficient Device Problem Information
CASE BOOKED FOR A REVISION. REVERSED THREADED POST BASEPLATE PUT IN, IN (B)(6) 2019 NEEDS TO BE REVISED, PROBABLY USING A PERFORM REVERSE, ANOTHER REVERSED THREADED POST OR CONVERT TO A HEMI. BOOKING FROM SURGEONS ROOMS, AFTER IMPLANT TO PATIENT (2019) UPDATE (B)(6) 2021 : REVISION DUE TO SCREW SNAPPING AND BASEPLATE LOOSENING.
Patient problems: Failure of Implant
MAUDE report 11933829 ↗InjuryAug 28, 2020Break; Loosening of Implant Not Related to Bone-Ingrowth; Break; Loosening of Implant Not Related to Bone-Ingrowth
VISIT DATE ON (B)(6) 2018 (ONSET DATE OF ADVERSE EVENT: (B)(6) 2017). POST-OPERATIVE COMPLICATION: HUMERAL FRACTURE / INJURY (TRAUMATIC). TRAUMA ON LEFT SHOULDER : 1 SCREW BROKEN AND LOOSENING OF GLENOÏD. NO SURGICAL PROCEDURE AT THAT TIME. (B)(6) 2018: REVISION SURGERY FOR IMPLANT BREAKAGE ON THE GLENOID SIDE.
Patient problems: Bone Fracture(s); Failure of Implant; Inadequate Osseointegration
MAUDE report 10464814 ↗InjuryApr 9, 2019Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
"IT WAS REPORTED THAT A PATIENT UNDERWENT A REVISION SURGERY DUE TO LOOSENING/LYSIS 3 YEARS AFTER PRIMARY SURGERY. SUBJECT ID: (B)(6)"
Patient problems: Patient Problem/Medical Problem
MAUDE report 8495325 ↗InjurySep 6, 2018Break; Break
"IT WAS REPORTED THAT A PATIENT UNDERWENT A REVISION SURGERY FOR IMPLANT BREAKAGE (GLENOID) 2 YEARS POST-OPERATIVE. STEM AND INSERT REMAINED IMPLANTED FROM PREVIOUS SURGERY. ONLY THE BASEPLATE AND GLENOID SPHERE WERE CHANGED. PATIENT ID : (B)(6)".
Patient problems: Failure of Implant
MAUDE report 7852641 ↗
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page