Skip to content

AEQUALIS REVERSED II DWE730Recalls & reported problems

Tornier · Class 2 · Implant

All models of AEQUALIS REVERSED II (46)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 03700386952661), received between Sep 6, 2018 and Dec 31, 2024 — classified as 9 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated3
  • Adverse Event Without Identified Device or Use Problem2
  • Break2
  • Loosening of Implant Not Related to Bone-Ingrowth2
  • Insufficient Device Problem Information1
  • Loss of Osseointegration1
  • Unintended Ejection1
  • Unstable1

Patient problems

  • Failure of Implant4
  • Implant Pain3
  • Inadequate Osseointegration2
  • Joint Dislocation2
  • Bone Fracture(s)1
  • Fall1
  • Joint Laxity1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 31, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED: "SUBJECT (B)(6) SHOUT - TORNIER SHOULDER OUTCOMES STUDY PAIN IN THE SHOULDER DUE TO A FALL."

    Patient problems: Fall; Implant Pain

    MAUDE report 21045853 ↗
  • InjuryDec 31, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED: "SUBJECT 008-00078 SHOUT - TORNIER SHOULDER OUTCOMES STUDY PATIENT REPORTS PAIN AND SWELLING."

    Patient problems: Implant Pain; Swelling/ Edema

    MAUDE report 21045849 ↗
  • InjuryJun 12, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT A PATIENT REQUIRED REVISION SURGERY AFTER DISLODGMENT AND POSSIBLE INFECTION.

    Patient problems: Joint Dislocation; Unspecified Infection

    MAUDE report 19521280 ↗
  • InjuryJun 5, 2024Unstable; Device Dislodged or Dislocated; Unstable; Device Dislodged or Dislocated

    AS REPORTED: "INSTABILITY/DISLOCATION. REVISION SURGERY (LEFT SHOULDER). HUMERAL SIDE COMPONENTS RETAINED FROM FIRST SURGERY". SERIES OF EVENTS BELOW: (B)(6) 2016: AEQUALIS ASCEND FLEX HUMERAL COMPONENTS / AEQUALIS REVERSED II GLENOID COMPONENTS (WITH A THREADED POST BASEPLATE) IMPLANTED (REVERSED CONFIGURATION). (B)(6) 2018: PATIENT FALL FROM BED AT HOME ("NO FRACTURE OR PROTHESIS DISLOCATION") *. (B)(6) 2018: INDICATION: MEDICAL EVALUATION OF PAIN FOLLOWING TRAUMA EVENT. DIFFUSE BONE DEMINERALIZATION. A DISLOCATION BACKWARDS AND DOWNWARDS OF THE SHOULDER PROSTHESIS. ABSENCE OF TRAUMATIC BONE LESION CLEARLY DEMONSTRATED (NUMEROUS ARTI-FACTS LINKED TO THE PROSTHESIS. (B)(6) 2018: INDICATION: MEDICAL EVALUATION AFTER SURGERY. DISLOCATION OF THE TOTAL LEFT REVERSED SHOULDER WITH A SMALL FRACTURE OF THE GLENOID INFERIOR SCREW. (B)(6) 2018: REVISION SURGERY WITH A REVERSED SHOULDER ARTHROPLASTY CONFIGURATION. -EXPLANTED: GLENOID COMPONENTS -RETAINED: HUMERAL COMPONENTS. -IMPLANTED: NEW BASEPLATE (WITH LONG POST BASEPLATE) AND SPHERE WITH GRAFT. -SLING, IMMOBILIZATION.

    Patient problems: Implant Pain; Joint Dislocation; Joint Laxity

    MAUDE report 19467895 ↗
  • InjuryJun 6, 2022Loss of Osseointegration; Loss of Osseointegration

    IT WAS REPORTED THAT THE PATIENT HAD A LEFT REVERSE SHOULDER IMPLANTED IN APPROXIMATELY 2019. THE GLENOID BASEPLATE COMPONENT OF THE SYSTEM BECAME LOOSE OVER TIME. THE GLENOID COMPONENT OF THE SYSTEM WAS REMOVED WITHOUT ISSUE AND THE IMPLANT WAS CONVERTED TO A HEMI SHOULDER WITH THE POLY REMOVED ON THE HUMERAL SIDE AND REPLACED WITH A CONVERTOR AND HUMERAL HEAD.

    Patient problems: Inadequate Osseointegration; No Clinical Signs, Symptoms or Conditions

    MAUDE report 14606044 ↗
  • MalfunctionFeb 25, 2022Unintended Ejection; Unintended Ejection

    THE OLIVE BROKE FROM BASEPLATE.

    Patient problems: Failure of Implant

    MAUDE report 13606700 ↗
  • InjuryJun 3, 2021Device Dislodged or Dislocated; Insufficient Device Problem Information; Device Dislodged or Dislocated; Insufficient Device Problem Information

    CASE BOOKED FOR A REVISION. REVERSED THREADED POST BASEPLATE PUT IN, IN (B)(6) 2019 NEEDS TO BE REVISED, PROBABLY USING A PERFORM REVERSE, ANOTHER REVERSED THREADED POST OR CONVERT TO A HEMI. BOOKING FROM SURGEONS ROOMS, AFTER IMPLANT TO PATIENT (2019) UPDATE (B)(6) 2021 : REVISION DUE TO SCREW SNAPPING AND BASEPLATE LOOSENING.

    Patient problems: Failure of Implant

    MAUDE report 11933829 ↗
  • InjuryAug 28, 2020Break; Loosening of Implant Not Related to Bone-Ingrowth; Break; Loosening of Implant Not Related to Bone-Ingrowth

    VISIT DATE ON (B)(6) 2018 (ONSET DATE OF ADVERSE EVENT: (B)(6) 2017). POST-OPERATIVE COMPLICATION: HUMERAL FRACTURE / INJURY (TRAUMATIC). TRAUMA ON LEFT SHOULDER : 1 SCREW BROKEN AND LOOSENING OF GLENOÏD. NO SURGICAL PROCEDURE AT THAT TIME. (B)(6) 2018: REVISION SURGERY FOR IMPLANT BREAKAGE ON THE GLENOID SIDE.

    Patient problems: Bone Fracture(s); Failure of Implant; Inadequate Osseointegration

    MAUDE report 10464814 ↗
  • InjuryApr 9, 2019Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    "IT WAS REPORTED THAT A PATIENT UNDERWENT A REVISION SURGERY DUE TO LOOSENING/LYSIS 3 YEARS AFTER PRIMARY SURGERY. SUBJECT ID: (B)(6)"

    Patient problems: Patient Problem/Medical Problem

    MAUDE report 8495325 ↗
  • InjurySep 6, 2018Break; Break

    "IT WAS REPORTED THAT A PATIENT UNDERWENT A REVISION SURGERY FOR IMPLANT BREAKAGE (GLENOID) 2 YEARS POST-OPERATIVE. STEM AND INSERT REMAINED IMPLANTED FROM PREVIOUS SURGERY. ONLY THE BASEPLATE AND GLENOID SPHERE WERE CHANGED. PATIENT ID : (B)(6)".

    Patient problems: Failure of Implant

    MAUDE report 7852641 ↗

Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page