Aequalis Reversed II MWB107Recalls & reported problems
Tornier · Drill Bit · Class 1
All models of AEQUALIS REVERSED II (46)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03700386968709), received between Dec 4, 2018 and Jun 19, 2026 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture3
- Break1
- Material Fragmentation1
Patient problems
- Device Embedded In Tissue or Plaque2
- Foreign Body In Patient1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 19, 2026Fracture
DURING THE STANDARD INTRAOPERATIVE PROCEDURE, IT WAS DISCOVERED THAT THE TIP OF THE DRILL BIT WAS MISSING. SUSPECTING REMAINING THE BROKEN TIP IN THE PATIENT'S BODY, THE BASE PLATE, WHICH WAS ALREADY IMPLANTED IN THE PATIENT, WAS REMOVED FOR INSPECTION. AFTER THE SURGERY, THE DISTRIBUTOR CHECKED THE PHOTO, WHICH WAS TAKEN BETWEEN CLEANING COMPLETION AND STARTING STERILIZATION BEFORE SURGERY, AND FOUND THAT THE TIP OF THE DRILL BIT WAS ALREADY GONE AT THAT TIME.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 25566194 ↗InjuryApr 25, 2025Break; Fracture; Break; Fracture
WHILE DRILLING FOR THE BASEPLATE SCREWS, THE DRILL BIT BROKE WITHIN THE PATIENTS SCAPULA. THE DRILL WAS TOO FAR IN THE EXTRACT SO SURGEON PLACED SHORTER SCREW IN THE SAME HOLE. THE DRILL STILL IN PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 21912357 ↗MalfunctionFeb 25, 2022Fracture; Fracture
THIS PATIENT HAD A SURGERY OF AEQUALIS REVERSED. WHEN THE ''DIA 3MM DRILL BIT (220MM)''WAS USED, THE DRILL TIP PART GOT STUCK AND COULDN'T PULL OUT. IT WAS MANAGED TO PULL OUT FORCEDLY, BUT THE DRILL TIP PART GOT BROKED. THIS DRILL TIP PART COULDN'T BE PULLED OUT , SO IT HAD TO REMAIN INSIDE THE BODY. ACCORDING TO THE SURGEON,THIS CASE WAS CAUSED BY A TECHNICAL ERROR. HE COULD NOT PULL OUT BECAUSE THE DRILL TIP PART WAS CLOGGED IN THE CORACOID PROCESS OF SCAPULA. THE DRILL IS USUALLY USED WITH CAUTION, BUT THIS TIME IT TOOK TOO LONG HANDLING THE DRILL. THE REVISION SURGERY WAS CARRIED OUT ON JUNE 6,AND THE BROKEN DRILL TIP WAS REMOVED WITHOUT ANY PROBLEM.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 13606871 ↗MalfunctionDec 4, 2018Material Fragmentation; Material Fragmentation
DURING DRILLING THE INFERIOR PERIPHERAL HOLE FOR THE BASEPLATE FIXATION: THE DRILL BIT BROKE OFF AND HAS BEEN LEFT INSIDE THE PATIENT.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 8128333 ↗
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,164 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT987 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT632 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT521 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT433 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT397 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page