Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03700386968709), received between Dec 4, 2018 and Jun 19, 2026 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3
  • Break1
  • Material Fragmentation1

Patient problems

  • Device Embedded In Tissue or Plaque2
  • Foreign Body In Patient1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 19, 2026Fracture

    DURING THE STANDARD INTRAOPERATIVE PROCEDURE, IT WAS DISCOVERED THAT THE TIP OF THE DRILL BIT WAS MISSING. SUSPECTING REMAINING THE BROKEN TIP IN THE PATIENT'S BODY, THE BASE PLATE, WHICH WAS ALREADY IMPLANTED IN THE PATIENT, WAS REMOVED FOR INSPECTION. AFTER THE SURGERY, THE DISTRIBUTOR CHECKED THE PHOTO, WHICH WAS TAKEN BETWEEN CLEANING COMPLETION AND STARTING STERILIZATION BEFORE SURGERY, AND FOUND THAT THE TIP OF THE DRILL BIT WAS ALREADY GONE AT THAT TIME.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 25566194 ↗
  • InjuryApr 25, 2025Break; Fracture; Break; Fracture

    WHILE DRILLING FOR THE BASEPLATE SCREWS, THE DRILL BIT BROKE WITHIN THE PATIENTS SCAPULA. THE DRILL WAS TOO FAR IN THE EXTRACT SO SURGEON PLACED SHORTER SCREW IN THE SAME HOLE. THE DRILL STILL IN PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 21912357 ↗
  • MalfunctionFeb 25, 2022Fracture; Fracture

    THIS PATIENT HAD A SURGERY OF AEQUALIS REVERSED. WHEN THE ''DIA 3MM DRILL BIT (220MM)''WAS USED, THE DRILL TIP PART GOT STUCK AND COULDN'T PULL OUT. IT WAS MANAGED TO PULL OUT FORCEDLY, BUT THE DRILL TIP PART GOT BROKED. THIS DRILL TIP PART COULDN'T BE PULLED OUT , SO IT HAD TO REMAIN INSIDE THE BODY. ACCORDING TO THE SURGEON,THIS CASE WAS CAUSED BY A TECHNICAL ERROR. HE COULD NOT PULL OUT BECAUSE THE DRILL TIP PART WAS CLOGGED IN THE CORACOID PROCESS OF SCAPULA. THE DRILL IS USUALLY USED WITH CAUTION, BUT THIS TIME IT TOOK TOO LONG HANDLING THE DRILL. THE REVISION SURGERY WAS CARRIED OUT ON JUNE 6,AND THE BROKEN DRILL TIP WAS REMOVED WITHOUT ANY PROBLEM.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 13606871 ↗
  • MalfunctionDec 4, 2018Material Fragmentation; Material Fragmentation

    DURING DRILLING THE INFERIOR PERIPHERAL HOLE FOR THE BASEPLATE FIXATION: THE DRILL BIT BROKE OFF AND HAS BEEN LEFT INSIDE THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8128333 ↗

Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices

All INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page