Tornier Flex Shoulder System DWF422BRecalls & reported problems
Tornier · Class 2 · Implant
All models of Tornier Flex Shoulder System (91)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 03700434019001), received between Feb 7, 2022 and Aug 18, 2026 — classified as 10 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem8
- Device Dislodged or Dislocated3
Patient problems
- Unspecified Infection4
- Bacterial Infection2
- Joint Dislocation2
- Failure of Implant1
- Implant Pain1
- Post Operative Wound Infection1
- Subluxation1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryAug 18, 2026Adverse Event Without Identified Device or Use Problem
A revision was performed due to infection. An antibiotic spacer was placed.
Patient problems: Post Operative Wound Infection
MAUDE report 50302417 ↗InjuryJun 15, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2026. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2025 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 25504189 ↗InjuryJan 28, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2021, A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 24197594 ↗InjuryMay 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2020, A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 22078583 ↗InjuryAug 22, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN APRIL 2012 ¿ MAY 2022. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2019 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER. REVISION PROCEDURE TYPE: DEBRIDEMENT AND IMPLANT RETENTION (DAIR).
Patient problems: Unspecified Infection
MAUDE report 20039388 ↗InjuryJul 3, 2024Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem; Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THERE WAS A THIRD REVISION SHOULDER REPLACEMENT SURGERY PERFORMED ON A PATIENT WITH RECURRENT DISLOCATIONS. THIS WAS THE FOURTH SURGERY ON THE SAME SHOULDER. ALL HUMERAL COMPONENTS (STEM, TRAY, POLY) WERE REMOVED AND REPLACED: A 2B STANDARD STEM WAS REPLACED WITH A 3C STANDARD STEM, A HIGH OFFSET +6 TRAY WAS REPLACED WITH A HIGH OFFSET +12 TRAY, AND A FLEX REVERSED INSERT B +9 WAS REPLACED WITH A FLEX REVERSED INSERT B +9 RETENTIVE. THE GLENOID IMPLANTS WERE LEFT INTACT.
Patient problems: Joint Dislocation
MAUDE report 19671045 ↗InjuryMar 4, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN (B)(6) 2012 ¿ (B)(6) 2022. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2021 A PATIENT REQUIRED REVISION SURGERY DUE TO DISLOCATION/SUBLUXATION , WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER. REVISION PROCEDURE TYPE: SINGLE STAGE REVISION.
Patient problems: Joint Dislocation; Subluxation
MAUDE report 18826352 ↗InjuryNov 7, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THERE WAS LUXATION OF THE IMPLANT AFTER REHABILITATION AND THE PATIENT UNDERWENT A REVISION SURGERY.
Patient problems: Failure of Implant
MAUDE report 15749088 ↗InjuryJun 7, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT SUFFERED AN INFECTION FOLLOWING A SHOULDER REPLACEMENT SURGERY. THE PATIENT UNDERWENT A REVISION SURGERY.
Patient problems: Bacterial Infection
MAUDE report 14621985 ↗InjuryFeb 7, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT (B)(6) HAD A POST-OPERATIVE COMPLICATION: DEEP INFECTION. "PATIENT COMPLAINING OF ONGOING PAIN- CT ASPIRATION WAS POSITIVE FOR C ACNES" PATIENT WILL SCHEDULE REVISION SURGERY. (THE ADVERSE EVENT WAS IDENTIFIED AS NOT RELATED TO THE STUDY DEVICE AND POSSIBLY RELATED TO THE SURGICAL PROCEDURE)
Patient problems: Bacterial Infection; Implant Pain
MAUDE report 13471129 ↗
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page