Tornier Flex Shoulder System DWF602BRecalls & reported problems
Tornier · Class 2 · Implant
All models of Tornier Flex Shoulder System (91)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Tornier Flex shoulder implant is used for shoulder replacement. The 26 reports describe patients needing repeat surgery for infections, fractures around the implant, and loosening or bone loss, with no deaths and 26 injuries mentioned. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 28
- Deaths reported
- 0
- Injuries reported
- 28
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 03700434020748), received between Feb 9, 2022 and Sep 18, 2026 — classified as 28 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem24
- Device Dislodged or Dislocated4
- Detachment of Device or Device Component1
- Loss of Osseointegration1
- Unstable1
Patient problems
- Unspecified Infection12
- Bone Fracture(s)5
- Implant Pain3
- Limb Fracture3
- Osteolysis3
- Failure of Implant2
- Joint Dislocation2
- Joint Laxity2
Recent adverse event reports
The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.
InjurySep 18, 2026Loss of Osseointegration
The manufacturer received a report from the National Joint Registry that contains unpublished collected data on the usage and the outcomes of the shoulder joint replacement. The report details analysis provided for revision procedures performed until (b)(6)2026.During the review of the report, it was identified that on (b)(6)-2026 a patient required revision surgery due to ASEPTIC LOOSENING HUMERUS, which was not previously reported to the manufacturer.
Patient problems: Osteolysis
MAUDE report 50815435 ↗InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem
The manufacturer received a report from the National Joint Registry that contains unpublished collected data on the usage and the outcomes of the shoulder joint replacement. The report details analysis provided for revision procedures performed until (b)(6) 2026.During the review of the report, it was identified that on (b)(6) 2025 a patient required revision surgery due to unexplained pain, which was not previously reported to the manufacturer.
Patient problems: Pain
MAUDE report 50803912 ↗InjuryJun 5, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED BETWEEN 2024-2026. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2026 A PATIENT REQUIRED REVISION SURGERY DUE TO PERIPROSTHETIC FRACTURE HUMERAL SIDE, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Bone Fracture(s)
MAUDE report 25404296 ↗InjuryMay 20, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PROVIDED IN MONTHLY REPORT OF APRIL 2026 DURING THE REVIEW OF THE REPORT, A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 25229283 ↗InjuryMay 20, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION FOR THE MONTHLY REPORT OF APRIL 2026 DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2026, A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 25229274 ↗InjuryMar 10, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL ON (B)(6) 2026. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2025 A PATIENT REQUIRED REVISION SURGERY DUE TO LYSIS-HUMERUS, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Osteolysis
MAUDE report 24558913 ↗InjuryFeb 18, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 24380778 ↗InjuryJan 28, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL JULY 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024, A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, PERIPROSTHETIC FRACTURE & LYSIS- HUMERUS, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Bone Fracture(s); Osteolysis; Unspecified Infection
MAUDE report 24197595 ↗InjuryNov 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, THE REVISION OF ASCEND FLEX STEM AND GLENO (CONCOMITANT) WAS DONE DUE TO FRACTURE OF THE GREATER TUBEROSITY. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Bone Fracture(s)
MAUDE report 23477855 ↗InjuryOct 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6)2025 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 23334479 ↗InjuryAug 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 22747533 ↗InjuryJul 9, 2025Unstable; Unstable
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL APRIL 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION & INSTABILITY, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER. STAGE 1 OF 2 STAGE REVISION.
Patient problems: Joint Laxity; Unspecified Infection
MAUDE report 22456094 ↗InjuryMay 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2025 A PATIENT REQUIRED REVISION SURGERY DUE TO STIFFNESS AND NATIVE GLENOID SURFACE EROSION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Erosion; Loss of Range of Motion
MAUDE report 22083347 ↗InjuryMay 20, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN (B)(6) 2013 ¿ (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2023, A PATIENT REQUIRED REVISION SURGERY DUE TO COMPONENT DISSOCIATION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Failure of Implant
MAUDE report 22056216 ↗InjuryMar 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN (B)(6) 2013 ¿ (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024, A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 21533754 ↗InjurySep 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN APRIL 2012 ¿ MAY 2022. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2022 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER. REVISION PROCEDURE TYPE: DEBRIDEMENT AND IMPLANT RETENTION (DAIR).
Patient problems: Unspecified Infection
MAUDE report 20180546 ↗InjuryJul 3, 2024Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem; Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THERE WAS A THIRD REVISION SHOULDER REPLACEMENT SURGERY PERFORMED ON A PATIENT WITH RECURRENT DISLOCATIONS. THIS WAS THE FOURTH SURGERY ON THE SAME SHOULDER. ALL HUMERAL COMPONENTS (STEM, TRAY, POLY) WERE REMOVED AND REPLACED: A 2B STANDARD STEM WAS REPLACED WITH A 3C STANDARD STEM, A HIGH OFFSET +6 TRAY WAS REPLACED WITH A HIGH OFFSET +12 TRAY, AND A FLEX REVERSED INSERT B +9 WAS REPLACED WITH A FLEX REVERSED INSERT B +9 RETENTIVE. THE GLENOID IMPLANTS WERE LEFT INTACT.
Patient problems: Joint Dislocation
MAUDE report 19670887 ↗InjuryFeb 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT MAY HAVE AN ALLERGY TO THE IMPLANT. THE PATIENT IS ALSO EXPERIENCING PAIN. IT WAS LATER FOUND THAT THE PATIENT HAD AN INFECTION AND WOULD NEED A REVISION SURGERY.
Patient problems: Implant Pain; Unspecified Infection
MAUDE report 18636297 ↗InjuryDec 19, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN APRIL 2012 ¿ MAY 2022. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT A PATIENT REQUIRED REVISION SURGERY DUE TO PERIPROSTHETIC FRACTURE, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Limb Fracture
MAUDE report 18360532 ↗InjuryDec 18, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN (B)(6)2012 ¿ (B)(6) 2022. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT A PATIENT REQUIRED REVISION SURGERY DUE TO PERIPROSTHETIC FRACTURE, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Limb Fracture
MAUDE report 18352436 ↗InjuryNov 8, 2023Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem; Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem
A PATIENT PRESENTED WITH SHOULDER PAIN EXACERBATED BY LYING ON THE LEFT SIDE. INITIAL X-RAYS WERE INCONCLUSIVE BUT INDICATED A SUSPECTED STRESS REACTION DUE TO DELTOID TENSION. A BONE SCAN ON A LATER DATE, CONFIRMED AN ACROMIAL STRESS REACTION. TREATMENT INVOLVED SIX WEEKS OF SLING USE FOLLOWED BY PHYSICAL THERAPY (PT), WHICH DID NOT ALLEVIATE SYMPTOMS. ANOTHER BONE SCAN ON THE SAME DAY CONFIRMED AN ACROMIAL STRESS FRACTURE. PT CONTINUED, SUPPLEMENTED BY A PLATELET-RICH PLASMA (PRP) INJECTION TO THE DELTOID. THE ADVERSE EVENT RESOLVED WITHOUT FURTHER ISSUES.
Patient problems: Bone Fracture(s); Implant Pain; Joint Laxity; Subluxation
MAUDE report 18096919 ↗InjuryJul 17, 2023Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem; Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT WAS A FROM AN ANATOMIC TOTAL SHOULDER. PATIENT KEPT DISLOCATING.
Patient problems: Joint Dislocation
MAUDE report 17337249 ↗InjuryMay 24, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THERE WAS A REVISION SURGERY CAUSED BY AN INFECTION. ADDITIONALLY, ALL TEN IMPLANTS WERE REMOVED. THE NURSE STILL HAD THE CENTRAL SCREW WITHIN THE BASEPLATE AND THE REVERSED INSERT WAS UNATTACHED FROM THE REVERSED TRAY.
Patient problems: Unspecified Infection
MAUDE report 16996592 ↗InjuryNov 8, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THERE WAS LUXATION OF THE IMPLANT AFTER REHABILITATION AND THE PATIENT UNDERWENT A REVISION SURGERY.
Patient problems: Failure of Implant
MAUDE report 15756585 ↗InjuryOct 25, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD A PERIPROSTHETIC FRACTURE OF THE REVERSED SHOULDER PROSTHESIS WHICH REQUIRED REOPERATION WITH A OSTEOSYNTHESIS PLATE.
Patient problems: Fall; Limb Fracture
MAUDE report 15668109 ↗
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page