Tornier Flex Shoulder System DWF510Recalls & reported problems
Tornier · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Tornier Flex shoulder replacement system has 106 reports, no deaths and 106 injuries. Many reports describe the joint dislocating or feeling unstable, and some describe infection that required revision surgery, sometimes with the implant loosening. Some reports describe pain or loss of motion. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 108
- Deaths reported
- 0
- Injuries reported
- 108
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 108 adverse event reports that list this exact device (DI 03700434022148), received between Feb 9, 2022 and Sep 21, 2026 — classified as 108 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (33 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem67
- Device Dislodged or Dislocated33
- Unstable10
- Biocompatibility2
- Loss of Osseointegration2
- Osseointegration Problem2
- Break1
- Detachment of Device or Device Component1
Patient problems
- Unspecified Infection42
- Joint Dislocation23
- Joint Laxity12
- Implant Pain11
- Subluxation8
- Loss of Range of Motion7
- Failure of Implant6
- Inadequate Osseointegration6
Recent adverse event reports
The 25 most recent of 108 reports, in the reporter's own words as published by the FDA.
InjurySep 21, 2026Adverse Event Without Identified Device or Use Problem
A revision surgery was performed as a result of a suspected infection after the patient fell.
Patient problems: Fall; Post Operative Wound Infection
MAUDE report 50841162 ↗InjurySep 17, 2026Unstable
The manufacturer received a report from the National Joint Registry that contains unpublished collected data on the usage and the outcomes of the shoulder joint replacement. The report details analysis provided for revision procedures performed until (b)(6) 2026.During the review of the report, it was identified that on (b)(6) 2026 a patient required revision surgery due to instability, which was not previously reported to the manufacturer.
Patient problems: Joint Laxity
MAUDE report 50803922 ↗InjuryAug 5, 2026Adverse Event Without Identified Device or Use Problem
The manufacturer received a report from the UK National Joint Registry that contains unpublished collected data on the usage and the outcomes of the shoulder joint replacement. The report details analysis provided for revision procedures performed until (b)(6)2026.During the review of the report, it was identified that a patient required revision surgery due to infection, which was not previously reported to the manufacturer.
Patient problems: Unspecified Infection
MAUDE report 50142491 ↗InjuryAug 3, 2026Adverse Event Without Identified Device or Use Problem
It was reported that, originally revision surgery cancelled due to unknown reasons. It was rebooked for (b)(6)2026. Surgeon had planned to revise glenoid due to glenoid loosening during a recent fall. During case, surgeon identified that arthroplasty joint had been infected in his opinion. All humeral and glenoid components were removed without difficulty. Multiple tissue samples were taken. Debridement and copious lavage was performed. Antibiotic cement spacer was implanted for a two stage revision. Dr (b)(6) stated that there were no clinical markers indicative of joint infection. Second stage revision planned for 2-3 months. No further x-ray & CT available No pathology report available No other patient details available. Second stage procedure using competitor product.
Patient problems: Unspecified Infection
MAUDE report 50122403 ↗InjuryJul 7, 2026Unstable; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT PREVIOUSLY UNDERWENT A LEFT REVERSE SHOULDER REPLACEMENT FOLLOWING A FRACTURE OF THE LEFT PROXIMAL HUMERAL HEAD. IN RECENT MONTHS, THE PATIENT EXPERIENCED SHOULDER INSTABILITY, RECURRENT DISLOCATIONS, AND PAIN. DURING THE REVISION PROCEDURE, THE PROSTHETIC SHOULDER JOINT WAS FOUND TO BE EASILY DISLOCATED WITH GROSS SOFT TISSUE LAXITY. ALL IMPLANTED COMPONENTS WERE WELL FIXED, AND NO POLYETHYLENE WEAR WAS OBSERVED. THE FLEX REVERSED TRAY, FLEX REVERSED INSERT, AND PERFORM REVERSED GLENOID STANDARD GLENOSPHERE WERE REMOVED AND REPLACED WITH A PERFORM REVERSED 36 MM ECCENTRIC GLENOSPHERE, LOW OFFSET +0 MM REVERSED TRAY, AND 36 MM/6 MM RETENTIVE B ANGLE POLYETHYLENE INSERT. ALL REMAINING IMPLANTS WERE LEFT IN PLACE. FOLLOWING IMPLANTATION OF THE NEW COMPONENTS, THE SHOULDER WAS REDUCED SUCCESSFULLY AND WAS STABLE THROUGHOUT THE FULL RANGE OF MOTION.
Patient problems: Joint Dislocation; Joint Laxity; Pain
MAUDE report 25738517 ↗InjuryJun 22, 2026Adverse Event Without Identified Device or Use Problem
REVISION OF ASCEND FLEX REVERSE STEM WITH PERFORM REVERSED GLENOID DUE TO INFECTION.
Patient problems: Post Operative Wound Infection
MAUDE report 25586482 ↗InjuryMay 20, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES OF MONTHLY REPORT OF (B)(6) 2026 DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2026, A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 25229472 ↗InjuryMay 20, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PROVIDED IN MONTHLY REPORT OF APRIL 2026 DURING THE REVIEW OF THE REPORT, A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 25229462 ↗InjuryMay 20, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION FOR THE MONTHLY REPORT OF APRIL 2026 DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2026, A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 25229276 ↗InjuryApr 30, 2026Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem
A REVISION SURGERY WAS PERFORMED DUE TO HUMERAL STEM LOOSENING CAUSED BY A SUSPECTED CHRONIC INFECTION AS WELL AS DISLOCATION.
Patient problems: Inadequate Osseointegration; Joint Dislocation; Post Operative Wound Infection
MAUDE report 25049781 ↗InjuryMar 10, 2026Device Dislodged or Dislocated
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED BETWEEN 2021 - 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2025, A PATIENT REQUIRED REVISION SURGERY DUE TO COMPONENT DISSOCIATION AND UNEXPLAINED PAIN, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Pain
MAUDE report 24560492 ↗InjuryFeb 18, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 24381033 ↗InjuryFeb 16, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL AUG. 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 24342508 ↗InjuryFeb 16, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 24342164 ↗InjuryFeb 12, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED IN 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2025, A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION AND PERIPROSTHETIC FRACTURE, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Bone Fracture(s); Unspecified Infection
MAUDE report 24322965 ↗InjuryJan 28, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024, A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 24197589 ↗InjuryJan 28, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024, A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 24197581 ↗InjuryJan 6, 2026Device Dislodged or Dislocated
IT WAS REPORTED THAT, THE PATIENT HAD A REVISION FOR A BIO-RSA GLENOID PATIENT WHICH HAD APPEARED TO HAVE A RE-ABSORBED. THE SURGEON REMOVED THE GLENOSPHERE AND BASEPLATE WITH EASE FROM THE GLENOID SOCKET AND PACKED IT FULL WITH FEMORAL HEAD ALLOGRAFT. THE TRAY AND POLY WERE REMOVED FROM THE FLEX STEM AND SURGEON PLACED A 50X16 COCR FLEX HUMERAL HEAD ONTO THE STEM TO TURN IT TO A HEMI. IT WAS REPORTED IN A FINAL REPORT FROM SURGEON, THE RIGHT SHOULDER PROSTHESIS WAS DISLOCATED WITH THE HUMERAL END LYING INFERIOR TO THE GLENOID PROSTHETIC MARGIN. EXTENSIVE OSTEOLYSIS SURROUNDING THE GLENOID COMPONENT.
Patient problems: Subluxation
MAUDE report 23976122 ↗InjuryJan 5, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2025 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 23958140 ↗InjuryJan 5, 2026Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED UNTIL JUNE 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER. REVISION PROCEDURE TYPE: STAGE 1 OF 2 STAGE REVISION.
Patient problems: Unspecified Infection
MAUDE report 23957278 ↗InjuryDec 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, REVISION SURGERY WAS PERFORMED. 2 STAGE PROCEDURE AS LIKELY TO BE INFECTED. PRIMARY SURGERY WAS PERFORMED ON (B)(6) 2022 WITH ASCEND FLEX AND REVERSE II IMPLANTS. ALL IMPLANTS WERE REMOVED AND PICTURED. A CEMENT SPACER WAS PLACED IN. IT WAS FURTHER REPORTED, THE PATIENT HAD PAIN AND SURGEON ORDERED X-RAYS (THESE WERE ATTACHED IN MY SUBMISSION) AND SURGEON NOTICED BONE REABSORPTION. DID NOT EXPECT INFECTION AS NO PATHOLOGY TESTS INDICATED IT. HOWEVER IN REVISION SURGEON THERE WAS STRONG EVIDENCE OF INFECTION WITH PUS AND INFECTED TISSUE.
Patient problems: Implant Pain; Osteolysis; Unspecified Infection
MAUDE report 23870835 ↗InjuryOct 28, 2025Loss of Osseointegration; Loss of Osseointegration
IT WAS REPORTED THAT, THE REVISION SURGERY WAS PERFORMED AND THE ASCEND FLEX/REVERSED II IMPLANTS WERE REVISED TO THE PERFORM REVERSE GLENOID IMPLANTS AS THE GLENOID IMPLANTS HAD MIGRATED SUPERIORLY. REVISION SURGERY WAS SUCCESSFULLY COMPLETED. IT WAS FURTHER REPORTED THAT, ALL GLENOID IMPLANTS WERE REMOVED (SUSPECTS), THE HUMERAL TRAY AND LINER WAS REMOVED (CONCOMITANTS). HUMERAL STEM REMAINED STABLE AND SO SURGEON LEFT IN SITU. NEW TRAY AND LINER AND GLENOID SYSTEM INSERTED SUCCESSFULLY.
Patient problems: Inadequate Osseointegration
MAUDE report 23405231 ↗InjuryOct 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT A PATIENT REQUIRED REVISION SURGERY DUE TO IMPINGEMENT, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Joint Impingement
MAUDE report 23335087 ↗InjurySep 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, A PATIENT WAS BOOKING IN FOR A REVISION PROCEDURE DUE TO ONGOING PAIN AND STIFFNESS. DURING THE DISSECTION LOTS OF FLUID/ PUSS WAS SEEN AND THE DECISION WAS MADE TO DO A 2 STAGE PROCEDURE TO CLEAR THE INFECTION. ANOTHER REVISION PROCEDURE WILL BE NEEDED AFTER THE INFECTION HAS CLEARED. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Loss of Range of Motion; Pain; Unspecified Infection
MAUDE report 23083919 ↗InjurySep 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE SHOULDER JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2025. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2025 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER.
Patient problems: Unspecified Infection
MAUDE report 23001789 ↗
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page