Skip to content

Blueprint Mixed Reality MRUE205Recalls & reported problems

Tornier · BP MxR Glenoid Box Lid · Class 1

All models of Blueprint Mixed Reality (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Mar 5, 2025Class IIOpen, Classified

    Stryker identified an elevated complaint rate related to difficulties using the Blueprint Mixed Reality Guidance Instrumentation system is multifactorial, but in each instance prohibits the system from meeting its intended use. The most commonly reported complaints were related to impaired software functioning, impaired instrumentation functioning, and general difficulties using the product.

    Z-1539-2025 · Root cause: Device Design · Recall details · FDA record ↗

Other TRAY, SURGICAL, INSTRUMENT devices

All TRAY, SURGICAL, INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page