ALN Optional Vena Cava Filter - Femoral Approach FF.010995Recalls & reported problems
A.L.N. · The device is an Optional Vena Cava Filter - Femoral Kit. It consists of a nine legs 316L stainless steel filter supplied with a delivery kit including a puncture needle, a J guidewire, a 7F introduction sheath, a dilatator and a pusher catheter. The filter is pre-packaged in a filter holder designed to facilitate the insertion of the filter in the introducer sheath. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 03760017680004), received between Apr 30, 2021 and Jun 17, 2022 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Migration” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Migration2
- Entrapment of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 17, 2022Migration; Migration
ON (B)(6) 2022, IN THE INTERVENTIONAL RADIOLOGY DEPARTMENT, THE PLACEMENT OF AN OPTIONAL VENA CAVA FILTER BY FEMORAL APPROACH IN A PATIENT WITH BILATERAL PULMONARY EMBOLISM WITH DEEP VEIN THROMBOSIS OF THE LOWER LIMBS FAILED. INDEED, SEVERAL ATTEMPTS TO RELEASE THE VENA CAVA WITHIN THE SUBRENAL VENA CAVA WERE MADE, RESULTING IN A RECOIL OF THE MATERIAL CAUDALLY TOWARDS THE ILIAC BIFURCATION. THE CRITICAL STEP WAS THE FIXED POINT-REMOVAL USING THE SUPPLIED PUSHER WHICH FAILED TO RELEASE THE FILTER. THE UNEXPANDED FILTER WAS FINALLY RELEASED THANKS TO THE DILATOR. A RADIOGRAPH CONFIRMED THE POOR DEPLOYMENT OF THE MATERIAL AND ITS MIGRATION IN THE SUPRARENAL SITUATION. THE MEDICAL TEAM THEREFORE ATTEMPTED TO RECOVER THE MATERIAL VIA THE SUPERIOR ROUTE (VIA THE JUGULAR APPROACH), BUT THIS FAILED BECAUSE THE ANGULATION OF THE INFERIOR VENA CAVA ABOVE THE KIDNEYS DID NOT ALLOW THE DEVICE TO BE REMOVED. ON (B)(6) 2022, A NEW ATTEMPT TO REMOVE THE FILTER WAS MADE, THE LATTER BEING LOCALIZED RETRO-HEPATICALLY. IN VIEW OF THE TWO SUCCESSIVE FAILURES TO REMOVE THIS FILTER, IT WAS DECIDED TO LEAVE IT IN PLACE FOR LIFE. THIS IS AN ISOLATED INCIDENT. THE CONSEQUENCES FOR THE PATIENT ARE MAINLY T
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14732761 ↗MalfunctionFeb 24, 2022Migration; Migration
WHEN INSERTING A VENA CAVA FILTER VIA THE FEMORAL ROUTE, THE DEVICE DID NOT DEPLOY CORRECTLY AND MIGRATED TO THE HEART. SIGNIFICANTLY LENGTHENED PROCEDURE. DEVICE RECOVERED BY LASSO IN JUGULAR AFTER TWO HOURS. NO CAVOGRAPHY IN THIS PATIENT WITH RENAL FAILURE. THE INDICATION FOR PLACEMENT IS MACROSCOPIC HEMATURIA, CONTRAINDICATING ANTICOAGULATION INITIALLY PLANNED FOR A PULMONARY EMBOLISM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13590068 ↗InjuryApr 30, 2021Entrapment of Device; Entrapment of Device
EVENT: DUODENUM PERFORATION. ON (B)(6) 2020, THE PATIENT UNDERWENT AN IVC FILTER IMPLANTATION PROCEDURE. THE PATIENT HAD REGULARLY GOTTEN CHECKED UP OTHER UNDERLYING DISEASES AND UNDERWENT A CT SCAN. IT SHOWED PERFORATION BY THE FILTER. THE PATIENT UNDERWENT ENDOSCOPY TO CHECK MORE DETAILS ON (B)(6) 2021. IT WAS OBSERVED THE FILTER DEVICE PERFORATED THE DUODENUM. PATIENT'S CONDITION AFTER THE OPERATION: STABLE. PATIENT'S ANATOMY AFTER THE IMPLANTATION: THERE IS NO TORTUOSITY IN THE IMPLANTATION LESION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11752482 ↗
Other Filter, Intravascular, Cardiovascular devices
- Denali® Vena Cava Filter DL900FBard Peripheral Vascular, Inc. · FILTER, INTRAVASCULAR, CARDIOVASCULAR22 deaths529 reports
- DENALI® Vena Cava Filter DL900JBard Peripheral Vascular, Inc. · FILTER, INTRAVASCULAR, CARDIOVASCULAR15 deaths492 reports
- Option Elite Retrievable Vena Cava Filter System 352506070EArgon Medical Devices, Inc. · Filter, Intravascular, Cardiovascular2 recalls3 deaths342 reports
- Gunther Tulip G33016Cook Incorporated · FILTER, INTRAVASCULAR, CARDIOVASCULAR3 deaths348 reports
- Gunther Tulip G33017Cook Incorporated · FILTER, INTRAVASCULAR, CARDIOVASCULAR3 deaths286 reports
- Cook Celect G34309William Cook Europe ApS · FILTER, INTRAVASCULAR, CARDIOVASCULAR1 recall1 death230 reports
- Gunther Tulip G52916William Cook Europe ApS · FILTER, INTRAVASCULAR, CARDIOVASCULAR1 recall2 deaths189 reports
- Cook Celect G34502William Cook Europe ApS · FILTER, INTRAVASCULAR, CARDIOVASCULAR1 recall2 deaths165 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page