ALN Optional Vena Cava Filter - Jugular Approach FJ.120096Recalls & reported problems
A.L.N. · The device is an Optional Vena Cava Filter - Jugular Kit. It consists of a nine legs 316L stainless steel filter supplied with a delivery kit including a puncture needle, a J guidewire, a 7F introduction sheath, a dilatator and a pusher catheter. The filter is pre-packaged in a filter holder designed to facilitate the insertion of the filter in the introducer sheath. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 03760017680011), received between Oct 2, 2020 and Jul 15, 2024 — classified as 5 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Entrapment of Device” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Entrapment of Device3
- Improper or Incorrect Procedure or Method2
- Adverse Event Without Identified Device or Use Problem1
- Insufficient Device Problem Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Hematoma2
- Hemorrhage/Blood Loss/Bleeding1
- Pain1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 15, 2024Entrapment of Device; Entrapment of Device
THE PATIENT UNDERWENT THE IVC FILTER PLACEMENT ABOUT 1.5 YEARS AGO (APPROXIMATE IMPLANTATION DATE: (B)(6) 2022). THE ATTENDING PHYSICIAN THIS TIME OBSERVED UNDER COMPUTED TOMOGRAPHY AT THE PATIENT'S 1.5-YEAR FOLLOW-UP VISIT THAT THE INFERIOR VENA CAVA WAS PERFORATED. TO BE MORE SPECIFIC, THE PHYSICIAN OBSERVED THE PERFORATION BY AND DEVIATION OF 6 LEGS OF THE FILTER. DAMAGES TO THE AORTA WERE ALSO DEEPLY CONCERNED. THE PHYSICIAN REMOVED THE FILTER SURGICALLY. THE REMOVED FILTER WAS DISPOSED OF BY THE MEDICAL INSTITUTION. PATIENT'S CONDITION AFTER THE OPERATION: FAVORABLE. THIS INFORMATION WAS OBTAINED THROUGH A POSTER PRESENTATION AT THE JAPANESE SOCIETY FOR VASCULAR SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19738299 ↗MalfunctionJun 2, 2022Entrapment of Device; Entrapment of Device
ON (B)(6) 2022, THE PATIENT UNDERWENT AN IVC FILTER IMPLANTATION PROCEDURE. THE PATIENT ANATOMY WAS TORTUOUS. THE TWO LONGEST STRUTS PERFORATED THE VESSEL A LITTLE BIT, BUT WERE NOT CONSIDERED SERIOUS. ON (B)(6) 2022, THE CATHETERIST/PHYSICIAN REVIEWED THE IMAGES AND NOTED THAT TWO STRUTS HAD VESSEL PERFORATED APPROXIMATELY 2 CM. THERE WAS A SMALL AMOUNT OF HEMORRHAGE FROM THE RIGHT VESSEL WALL WHERE ONE OF THE STRUTS PERFORATED. ADDITIONAL PROCEDURE HAVE NOT BEEN TAKEN. PATIENT'S CONDITION AFTER THE OPERATION: AS OF TODAY: NO ABDOMINAL PAIN, NO DECREASE IN VITALS, NO SERIOUS ADVERSE EVENT. THE DEVICE WILL BE IMPLANTED PERMANENTLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14572571 ↗InjuryMay 5, 2022Entrapment of Device; Entrapment of Device
ON (B)(6) 2022, THE PATIENT UNDERWENT AN IVC FILTER IMPLANTATION PROCEDURE. THE IMPLANTED FILTER TILTED AND NOT PERFORATED THE VESSEL WALL. ON AN UNKNOWN DATE OF (B)(6) 2022, THE PATIENT UNDERWENT CT SCAN FOR FILTER EXTRACTION PROCEDURE, WHICH SHOWED TWO OF THE ANCHORAGE STRUTS PERFORATED THE VESSEL WALL. ON (B)(6) 2022, THE PATIENT UNDERWENT A FILTER EXTRACTION PROCEDURE. THE FILTER WAS EXTRACTED. POSTOPERATIVE ANGIOGRAPHY DID NOT SHOW CONTRAST AGENT LEAKED FROM THE BLOOD VESSELS. PATIENT'S CONDITION FAVORABLE AFTER THE OPERATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14286470 ↗InjuryFeb 24, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT UNDERWENT AN IVC FILTER IMPLANTATION PROCEDURE FOR CENTRAL DEEP VEIN THROMBOSIS AS A PREVENTIVE TREATMENT OF OOPHORECTOMY. ON THE MORNING OF (B)(6) 2021, THE PATIENT UNDERWENT AN OOPHORECTOMY PROCEDURE. A POSTOPERATIVE X-RAY SHOWED THE FILTER MIGRATED TO THE SUPERIOR VENA CAVA. ON THE AFTERNOON OF (B)(6) 2021, THE PATIENT UNDERWENT A FILTER EXTRACTION PROCEDURE. THE FILTER WAS RETRIEVED SUCCESSFULLY. NO ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13590066 ↗InjuryMar 10, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
EVENT: VASCULAR INJURY. ON (B)(6) 2021, THE PATIENT UNDERWENT AN ALN FILTER IMPLANTATION IN THE INFRARENAL INFERIOR VENA CAVA TO PREVENT PULMONARY THROMBOEMBOLISM. ON (B)(6) 2021, THE PATIENT COMPLAINED OF BACK PAIN. CT WAS PERFORMED AND SHOWED THAT THE LUMBAR ARTERY NEAR THE IMPLANTED FILTER WAS DAMAGED AND BLEEDING. IT BECAME A LARGE HEMATOMA COMPRESSING THE VENA CAVA. THE PATIENT UNDERWENT AN ACUTE PROCEDURE. A CUFF OF AFX STENT GRAFT SYSTEM ((B)(6) LIFELINE) WAS PLACED IN THE AORTA TO STOP THE BLEEDING. THE PROCEDURE WAS FISHED. PATIENT'S CONDITION AFTER THE OPERATION: THE PATIENT IS NO MORE IN PAIN. THE HEMATOMA HOWEVER STILL COMPRESSES THE FILTER AND THE VENA CAVA, WHICH BECAME NARROWED AND DEFORMED. PHYSICIAN'S COMMENT: I IMPLANTED THE FILTER SUCCESSFULLY AND APPROPRIATELY IN ORDER NOT TO COMPRESS THE VESSELS, ALTHOUGH THE TARGET VEIN MIGHT HAVE BEEN SMALLER IN DIAMETER THAN I THOUGHT. IMAGES TAKEN AFTER THE IMPLANTATION SUGGESTED NO PROBLEM. I SUSPECTED THE IMPLANTED FILTER CAUSED THE PAIN BASED ON THE AREA THE PATIENT INDICATED, SO I TOOK A CT SCAN RIGHT AWAY. THE PATIENT MIGHT HAVE DIED IF I HAD NOT FOUND THE BLEEDING THEN.
Patient problems: Hematoma; Hemorrhage/Blood Loss/Bleeding; Pain
MAUDE report 11451141 ↗InjuryDec 10, 2020Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
ON (B)(6) 2020, THE PATIENT UNDERWENT A FILTER IMPLANTATION PROCEDURE. ON (B)(6) 2020, A CT SCAN FOUND NO ABNORMALITY IN THE PATIENT. ON (B)(6) 2020, THE PATIENT COMPLAINED OF BACK PAIN. A CT SCAN REVEALED A RETROPERITONEAL HEMATOMA. WHETHER PERFORATION OCCURRED OR NOT WAS UNKNOWN BECAUSE AN ANGIOGRAPHY WAS NOT PERFORMED. THERE ARE NO EXTERNAL FACTORS FOUND WHICH MIGHT CAUSE PERFORATION, AND THE INFLUENCE OF THE TREATED TUMOR WAS UNLIKELY. ON (B)(6) 2020, IT WAS PLANNED TO TRANSFER THE PATIENT TO THE NATIONAL CEREBRAL AND CARDIOVASCULAR CENTER. BECAUSE DR. (B)(6) OF THE INITIAL HOSPITAL COULD NOT PERFORM THE FILTER RETRIEVAL PROCEDURE DUE TO MISSING THE SURGICAL BACKUP SECTION. PATIENT'S CONDITION AFTER THE OPERATION: THE PATIENT IS TREATED WITH OBSERVATION.
Patient problems: Hematoma
MAUDE report 10981730 ↗InjuryOct 2, 2020Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IN (B)(6) 2020, THE PATIENT UNDERWENT AN IVC FILTER IMPLANTATION PROCEDURE FOR DVT. ON (B)(6) 2020, ENDOSCOPIC IMAGES REVEALED PERFORATION IN THE DUODENUM. PATIENT ASYMPTOMATIC. ON (B)(6) 2020, THE PHYSICIAN REMOVED THE FILTER WITH THE FILTER EXTRACTION KIT. A CHECK ANGIOGRAPHY REVEALED THE EXTRAVASATION OCCURRED WHERE THE FILTER PERFORATED. THE PHYSICIAN THINKS THAT IT IS NOT SERIOUS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10618675 ↗
Other Filter, Intravascular, Cardiovascular devices
- Denali® Vena Cava Filter DL900FBard Peripheral Vascular, Inc. · FILTER, INTRAVASCULAR, CARDIOVASCULAR22 deaths529 reports
- DENALI® Vena Cava Filter DL900JBard Peripheral Vascular, Inc. · FILTER, INTRAVASCULAR, CARDIOVASCULAR15 deaths492 reports
- Option Elite Retrievable Vena Cava Filter System 352506070EArgon Medical Devices, Inc. · Filter, Intravascular, Cardiovascular2 recalls3 deaths342 reports
- Gunther Tulip G33016Cook Incorporated · FILTER, INTRAVASCULAR, CARDIOVASCULAR3 deaths348 reports
- Gunther Tulip G33017Cook Incorporated · FILTER, INTRAVASCULAR, CARDIOVASCULAR3 deaths286 reports
- Cook Celect G34309William Cook Europe ApS · FILTER, INTRAVASCULAR, CARDIOVASCULAR1 recall1 death230 reports
- Gunther Tulip G52916William Cook Europe ApS · FILTER, INTRAVASCULAR, CARDIOVASCULAR1 recall2 deaths189 reports
- Cook Celect G34502William Cook Europe ApS · FILTER, INTRAVASCULAR, CARDIOVASCULAR1 recall2 deaths165 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page