A.L.N. · The device is an Optional Vena Cava Filter - Jugular Kit. It consists of a nine legs 316L stainless steel filter supplied with a delivery kit including a puncture needle, a J guidewire, a 7F introduction sheath, a dilatator and a pusher catheter. The filter is pre-packaged in a filter holder designed to facilitate the insertion of the filter in the introducer sheath. · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 03760017680011), received between Oct 2, 2020 and Jul 15, 2024 — classified as 5 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Entrapment of Device” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Entrapment of Device3
  • Improper or Incorrect Procedure or Method2
  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Hematoma2
  • Hemorrhage/Blood Loss/Bleeding1
  • Pain1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 15, 2024Entrapment of Device; Entrapment of Device

    THE PATIENT UNDERWENT THE IVC FILTER PLACEMENT ABOUT 1.5 YEARS AGO (APPROXIMATE IMPLANTATION DATE: (B)(6) 2022). THE ATTENDING PHYSICIAN THIS TIME OBSERVED UNDER COMPUTED TOMOGRAPHY AT THE PATIENT'S 1.5-YEAR FOLLOW-UP VISIT THAT THE INFERIOR VENA CAVA WAS PERFORATED. TO BE MORE SPECIFIC, THE PHYSICIAN OBSERVED THE PERFORATION BY AND DEVIATION OF 6 LEGS OF THE FILTER. DAMAGES TO THE AORTA WERE ALSO DEEPLY CONCERNED. THE PHYSICIAN REMOVED THE FILTER SURGICALLY. THE REMOVED FILTER WAS DISPOSED OF BY THE MEDICAL INSTITUTION. PATIENT'S CONDITION AFTER THE OPERATION: FAVORABLE. THIS INFORMATION WAS OBTAINED THROUGH A POSTER PRESENTATION AT THE JAPANESE SOCIETY FOR VASCULAR SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19738299 ↗
  • MalfunctionJun 2, 2022Entrapment of Device; Entrapment of Device

    ON (B)(6) 2022, THE PATIENT UNDERWENT AN IVC FILTER IMPLANTATION PROCEDURE. THE PATIENT ANATOMY WAS TORTUOUS. THE TWO LONGEST STRUTS PERFORATED THE VESSEL A LITTLE BIT, BUT WERE NOT CONSIDERED SERIOUS. ON (B)(6) 2022, THE CATHETERIST/PHYSICIAN REVIEWED THE IMAGES AND NOTED THAT TWO STRUTS HAD VESSEL PERFORATED APPROXIMATELY 2 CM. THERE WAS A SMALL AMOUNT OF HEMORRHAGE FROM THE RIGHT VESSEL WALL WHERE ONE OF THE STRUTS PERFORATED. ADDITIONAL PROCEDURE HAVE NOT BEEN TAKEN. PATIENT'S CONDITION AFTER THE OPERATION: AS OF TODAY: NO ABDOMINAL PAIN, NO DECREASE IN VITALS, NO SERIOUS ADVERSE EVENT. THE DEVICE WILL BE IMPLANTED PERMANENTLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14572571 ↗
  • InjuryMay 5, 2022Entrapment of Device; Entrapment of Device

    ON (B)(6) 2022, THE PATIENT UNDERWENT AN IVC FILTER IMPLANTATION PROCEDURE. THE IMPLANTED FILTER TILTED AND NOT PERFORATED THE VESSEL WALL. ON AN UNKNOWN DATE OF (B)(6) 2022, THE PATIENT UNDERWENT CT SCAN FOR FILTER EXTRACTION PROCEDURE, WHICH SHOWED TWO OF THE ANCHORAGE STRUTS PERFORATED THE VESSEL WALL. ON (B)(6) 2022, THE PATIENT UNDERWENT A FILTER EXTRACTION PROCEDURE. THE FILTER WAS EXTRACTED. POSTOPERATIVE ANGIOGRAPHY DID NOT SHOW CONTRAST AGENT LEAKED FROM THE BLOOD VESSELS. PATIENT'S CONDITION FAVORABLE AFTER THE OPERATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14286470 ↗
  • InjuryFeb 24, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT UNDERWENT AN IVC FILTER IMPLANTATION PROCEDURE FOR CENTRAL DEEP VEIN THROMBOSIS AS A PREVENTIVE TREATMENT OF OOPHORECTOMY. ON THE MORNING OF (B)(6) 2021, THE PATIENT UNDERWENT AN OOPHORECTOMY PROCEDURE. A POSTOPERATIVE X-RAY SHOWED THE FILTER MIGRATED TO THE SUPERIOR VENA CAVA. ON THE AFTERNOON OF (B)(6) 2021, THE PATIENT UNDERWENT A FILTER EXTRACTION PROCEDURE. THE FILTER WAS RETRIEVED SUCCESSFULLY. NO ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13590066 ↗
  • InjuryMar 10, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    EVENT: VASCULAR INJURY. ON (B)(6) 2021, THE PATIENT UNDERWENT AN ALN FILTER IMPLANTATION IN THE INFRARENAL INFERIOR VENA CAVA TO PREVENT PULMONARY THROMBOEMBOLISM. ON (B)(6) 2021, THE PATIENT COMPLAINED OF BACK PAIN. CT WAS PERFORMED AND SHOWED THAT THE LUMBAR ARTERY NEAR THE IMPLANTED FILTER WAS DAMAGED AND BLEEDING. IT BECAME A LARGE HEMATOMA COMPRESSING THE VENA CAVA. THE PATIENT UNDERWENT AN ACUTE PROCEDURE. A CUFF OF AFX STENT GRAFT SYSTEM ((B)(6) LIFELINE) WAS PLACED IN THE AORTA TO STOP THE BLEEDING. THE PROCEDURE WAS FISHED. PATIENT'S CONDITION AFTER THE OPERATION: THE PATIENT IS NO MORE IN PAIN. THE HEMATOMA HOWEVER STILL COMPRESSES THE FILTER AND THE VENA CAVA, WHICH BECAME NARROWED AND DEFORMED. PHYSICIAN'S COMMENT: I IMPLANTED THE FILTER SUCCESSFULLY AND APPROPRIATELY IN ORDER NOT TO COMPRESS THE VESSELS, ALTHOUGH THE TARGET VEIN MIGHT HAVE BEEN SMALLER IN DIAMETER THAN I THOUGHT. IMAGES TAKEN AFTER THE IMPLANTATION SUGGESTED NO PROBLEM. I SUSPECTED THE IMPLANTED FILTER CAUSED THE PAIN BASED ON THE AREA THE PATIENT INDICATED, SO I TOOK A CT SCAN RIGHT AWAY. THE PATIENT MIGHT HAVE DIED IF I HAD NOT FOUND THE BLEEDING THEN.

    Patient problems: Hematoma; Hemorrhage/Blood Loss/Bleeding; Pain

    MAUDE report 11451141 ↗
  • InjuryDec 10, 2020Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    ON (B)(6) 2020, THE PATIENT UNDERWENT A FILTER IMPLANTATION PROCEDURE. ON (B)(6) 2020, A CT SCAN FOUND NO ABNORMALITY IN THE PATIENT. ON (B)(6) 2020, THE PATIENT COMPLAINED OF BACK PAIN. A CT SCAN REVEALED A RETROPERITONEAL HEMATOMA. WHETHER PERFORATION OCCURRED OR NOT WAS UNKNOWN BECAUSE AN ANGIOGRAPHY WAS NOT PERFORMED. THERE ARE NO EXTERNAL FACTORS FOUND WHICH MIGHT CAUSE PERFORATION, AND THE INFLUENCE OF THE TREATED TUMOR WAS UNLIKELY. ON (B)(6) 2020, IT WAS PLANNED TO TRANSFER THE PATIENT TO THE NATIONAL CEREBRAL AND CARDIOVASCULAR CENTER. BECAUSE DR. (B)(6) OF THE INITIAL HOSPITAL COULD NOT PERFORM THE FILTER RETRIEVAL PROCEDURE DUE TO MISSING THE SURGICAL BACKUP SECTION. PATIENT'S CONDITION AFTER THE OPERATION: THE PATIENT IS TREATED WITH OBSERVATION.

    Patient problems: Hematoma

    MAUDE report 10981730 ↗
  • InjuryOct 2, 2020Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    IN (B)(6) 2020, THE PATIENT UNDERWENT AN IVC FILTER IMPLANTATION PROCEDURE FOR DVT. ON (B)(6) 2020, ENDOSCOPIC IMAGES REVEALED PERFORATION IN THE DUODENUM. PATIENT ASYMPTOMATIC. ON (B)(6) 2020, THE PHYSICIAN REMOVED THE FILTER WITH THE FILTER EXTRACTION KIT. A CHECK ANGIOGRAPHY REVEALED THE EXTRAVASATION OCCURRED WHERE THE FILTER PERFORATED. THE PHYSICIAN THINKS THAT IT IS NOT SERIOUS.

    Patient problems: No Clinical Signs, Symptoms or Conditions