Skip to content

ALN Optional Vena Cava Filter - Femoral Approach with Hook FF.HOOKRecalls & reported problems

A.L.N. · The device is an Optional Vena Cava Filter - Femoral Kit with Hook. It consists of a nine legs 316L stainless steel filter with a hook at the apex supplied with a delivery kit including a puncture needle, a J guidewire, a 7F introduction sheath, a dilatator and a pusher catheter. The filter is pre-packaged in a filter holder designed to facilitate the insertion of the filter in the introducer sheath. · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03760017680042), received between Feb 7, 2019 and Jul 17, 2025 — classified as 1 death, 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Device Slipped” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Slipped1
  • Entrapment of Device1
  • Fracture1
  • Improper or Incorrect Procedure or Method1
  • Migration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Exsanguination1
  • Failure of Implant1
  • Hematoma1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 17, 2025Fracture; Fracture

    THE FILTER WAS PLACED ON THE (B)(6) 2024 AND REMOVED ON (B)(6) 2025.LASSO TECHNIQUE WAS USED FOR THE FILTER REMOVAL (COOK CLOVERSNARE®).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22529990 ↗
  • MalfunctionFeb 24, 2023Migration; Migration

    UPON DEPLOYMENT OF IVC FILTER, MIGRATED TO THE LEVEL OF THE PULMONARY ARTERY. THE PHYSICIAN WAS ABLE TO SNARE AND REMOVE THE FILTER IN ITS ENTIRETY.

    Patient problems: Failure of Implant

    MAUDE report 16434056 ↗
  • InjuryDec 9, 2020Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    IMPLANTATION BY RIGHT FEMORAL APPROACH FOLLOWING FLOATING VENOUS THROMBUS IN FEMORAL VEIN IN LEFT LIMB. INCOMPLETE OPENING OF THE FILTER LEADING TO MIGRATION TO THE LEFT PULMONARY ARTERY. FILTER REMOVED FROM THE PULMONARY ARTERY WITH A LASSO BUT ANCHORED STRONGLY IN RIGHT ILIAC VEIN IN A TRANSVERSE POSITION. DECISION TO LEAVE THE FILTER IN THAT POSITION AFTER MULTIPLE ATTEMPTS OF EXTRACTION USING A SNARE AND THE FILTER EXTRACTION KIT BOTH VIA FEMORAL AND JUGULAR ROUTES. SUCCESSFUL IMPLANTATION OF A SECOND FILTER BY JUGULAR APPROACH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10975548 ↗
  • DeathFeb 7, 2019Entrapment of Device; Device Slipped; Entrapment of Device; Device Slipped

    A (B)(6) FEMALE WITH PAST MEDICAL HISTORY OF RECURRENT DEEP VEIN THROMBOSIS AND PULMONARY THROMBOEMBOLI, HYPERTENSION, AND COPD PRESENTED ON (B)(6) 2019 WITH COMPLAINT OF RIGHT UPPER QUADRANT ABDOMINAL PAIN AND WAS DIAGNOSED WITH ACUTE CHOLECYSTITIS. SHE UNDERWENT LAPAROSCOPIC CHOLECYSTECTOMY AND PLACEMENT OF INFERIOR VENA CAVE FILTER. PT WAS DISCHARGED ON (B)(6) 2019. AFTER DISCHARGE, PT WAS FOUND DECEASED. AUTOPSY WAS REQUESTED BY PATIENT'S FAMILY. ANATOMIC PRELIMINARY DIAGNOSIS FOUND RETROPERITONEAL HEMORRHAGE (RIGHT RETROPERITONEAL HEMATOMA, SIX STRUTS OF AN INFERIOR VENA CAVA FILTER PERFORATING THE WALL OF THE INFERIOR VENA CAVA). THE AUTOPSY FINDINGS NOTED THAT THE CHOLECYSTECTOMY SURGICAL BED WAS FREE OF UNDUE HEMORRHAGE OR EVIDENCE OF INFECTION. THE AUTOPSY FINDINGS ARE PRELIMINARY. HISTOLOGIC EXAMINATION IS PENDING.

    Patient problems: Exsanguination; Hematoma

    MAUDE report 8358178 ↗

Other Filter, Intravascular, Cardiovascular devices

All Filter, Intravascular, Cardiovascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page