Skip to content

SOPHYSA PSO-VTRecalls & reported problems

Sophysa · PRESSIO ICP MONITORING KIT, VENTRICULAR TUNNELING · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 03760124131482), received between Jan 19, 2017 and Feb 10, 2026 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Device Sensing Problem” (7 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K062584); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 9, 2024Open, Classified

    An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient. The cause of the leaks is indicated to be manufacturing error involving perforation of the inner channel of the catheter during stylet insertion.

    Z-1825-2024 · Root cause: Process change control · FDA record ↗

Most reported problems

Device problems

  • Device Sensing Problem7
  • Electrical /Electronic Property Problem3
  • Arcing1
  • Incorrect Measurement1
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 10, 2026Device Sensing Problem

    A PSO-VT WAS INDICATED TO MONITOR ICP IN A CASE OF TRAUMATIC BRAIN INJURY. THE MONITOR SHOWED E001 AFTER THE IMPLANTATION. THIS INCIDENT LEADS TO ADDITIONAL SURGERY FOR THE PATIENT. THE DEVICE WAS EXPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24304498 ↗
  • MalfunctionAug 5, 2025Device Sensing Problem; Device Sensing Problem

    A PSO-VT WAS INDICATED TO MONITOR ICP IN A CASE OF HYDROCEPHALUS. DURING FOLLOW-UP, AN ERROR E001 WAS OBSERVED. NO CLINICALS SIGNS OR SYMPTOMS WERE OBSERVED IN THE PATIENT. THE CATHETER FUNCTIONED FOR 12 HOURS. THIS INCIDENT LED TO THE PATIENT'S HOSPITALISATION BEING EXTENDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22715092 ↗
  • MalfunctionJul 2, 2025Device Sensing Problem; Device Sensing Problem

    A PSO-VT WAS INDICATED TO MONITOR ICP IN A CASE OF TRAUMATIC BRAIN INJURY. DURING FOLLOW-UP, AN ERROR E001 HAS OCCURED ON THE MONITOR PSO-3000. THERE IS NO CLINICAL SIGNS OR SYMPTOMS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22382109 ↗
  • MalfunctionJan 15, 2024Arcing; Device Sensing Problem; Arcing; Device Sensing Problem

    FOUR DAYS AFTER THE CATHETER WAS IMPLANTED, THE MACHINE SHOWED E001.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18510925 ↗
  • MalfunctionNov 27, 2023Device Sensing Problem; Device Sensing Problem

    THE MACHINE SHOWED E001 AFTER THE CATHETER WAS IMPLANTED FOR 24 HOURS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18208260 ↗
  • MalfunctionSep 26, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CATHETER WAS IMPLANTED ON (B)(6), THE MACHINE SHOWED E001 ON (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17817504 ↗
  • MalfunctionMay 29, 2023Obstruction of Flow; Device Sensing Problem; Obstruction of Flow; Device Sensing Problem

    ERROR IN PRESSURE MEASUREMENT AT ANY TIME FOR NO APPARENT REASON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17017175 ↗
  • MalfunctionMar 29, 2022Device Sensing Problem; Device Sensing Problem

    DEVICE IMPLANTED ON (B)(6) 2022. THE SAME DAY, FOUR HOURS AFTER INITIAL SURGERY, THE MONITOR DISPLAYED ERROR E001. THE MONITORING COULDN'T BE MAINTAINED SO THE DEVICE WAS EXPLANTED. THERE WERE NO HEALTH COMPLICATIONS FOR THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13939220 ↗
  • MalfunctionSep 1, 2020Incorrect Measurement; Incorrect Measurement

    THE MONITOR DISPLAYED AN ERROR MESSAGE E001 AFTER IMPLANTATION OF THE PROBE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10475695 ↗
  • MalfunctionJan 24, 2017Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    PROBLEM WITH ICP MONITORING KIT : THE CATHETER SHOWED E001 ERROR.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6272447 ↗
  • MalfunctionJan 19, 2017Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    PROBLEM WITH THE ICP MONITORING KIT : THE CATHETER SHOWED E002 ERROR.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6263048 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K062584. One submission often covers many models, so check the product and lot details in each record.

  • Z-1826-2024 · Apr 9, 2024 · Open, Classified
    The Sophysa Pressio Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure and temperature with tunneling (PSO-VTT

Other DEVICE, MONITORING, INTRACRANIAL PRESSURE devices

All DEVICE, MONITORING, INTRACRANIAL PRESSURE devices →

Sources and updates