Skip to content

TRIWAY N02 00004Recalls & reported problems

In2bones · FIXATION SCREW TRIWAY ARTHRODESIS NAIL · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03760225717882), received between Mar 30, 2026 and Mar 30, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Mechanical Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem1
  • Use of Device Problem1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 30, 2026Mechanical Problem; Use of Device Problem

    DURING THE ASSEMBLY OF THE JIG FOR THE TRIWAY NAIL, THE CONNECTOR EQUIPPED WITH THE FIXATION SCREW BECAME STUCK IN THE COMPRESSION WHEEL AND THE POSITIONING GUIDE. DUE TO THE FLAT SIDE OF THE SCREW, IT TENDS TO SIT AT AN ANGLE IN THE THREADING, CAUSING IT TO JAM. THE SCREW THEN BECOMES VERY DIFFICULT TO LOOSEN OR REMOVE. THE CUSTOMER EVENTUALLY MANAGED, AFTER SEVERAL ATTEMPTS, TO FREE THE SCREW AND CONTINUE THE PROCEDURE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24711933 ↗

Other Guide, Surgical, Instrument devices

All Guide, Surgical, Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page